Background: Mini-implants are crucial in orthodontics for absolute anchorage, allowing precise dental movements without affecting adjacent structures. Exploring alternative materials with proven biocompatibility and potential for improved clinical performance is essential. This study compares the mechanical resistance of polyetheretherketone (PEEK) and stainless steel mini-implants to evaluate PEEK's feasibility as an alternative.
Objectives: To compare PEEK and stainless steel for orthodontic mini-implants, focusing on maximum load resistance and suitability for temporary anchorage.
Results: Stainless steel mini-implants showed a mean maximum load of 255.77 N whereas PEEK averaged 57.14 N after outlier adjustment. Welch's t -test confirmed a significant difference between groups ( P < 0.001) with a large effect size (Hedges' g = 2.31). Statistically significant differences were observed between groups, with a mean difference of 198.6 N (95% CI 126.3-271.0).
Conclusion: Although stainless steel exhibited greater load resistance, PEEK mini-implants exceeded clinically relevant orthodontic forces (50-400 g), supporting their potential as temporary anchorage devices. Further in vivo and clinical studies are needed to confirm long-term performance.
期刊介绍:
Indian Journal of Dental Research (IJDR) is the official publication of the Indian Society for Dental Research (ISDR), India section of the International Association for Dental Research (IADR), published quarterly. IJDR publishes scientific papers on well designed and controlled original research involving orodental sciences. Papers may also include reports on unusual and interesting case presentations and invited review papers on significant topics.