Artur Hausch, Maria Deisenhammer, Nenad Mitrovic, Ulrich J Sachs, Thomas Mueller
{"title":"Effects of extended storage of blood samples on the results of platelet function testing: an <i>in vitro</i> stability study.","authors":"Artur Hausch, Maria Deisenhammer, Nenad Mitrovic, Ulrich J Sachs, Thomas Mueller","doi":"10.1080/00365513.2026.2630329","DOIUrl":null,"url":null,"abstract":"<p><p>In the present study, we aimed to explore the effects of extended storage of blood samples from 30 patients on the results of platelet function testing using the PFA-200 device (Collagen/EPI and Collagen/ADP tests), Multiplate instrument (ASPI, ADP, and TRAP tests), and VerifyNow analyzer (Aspirin and PRU tests). We determined platelet function (1) immediately after blood collection (baseline), (2) after storage of blood samples at room temperature for 3-4 h (i.e. the maximum sample storage period permitted by the manufacturers), and (3) after storage of blood samples for 5 h at room temperature (extended storage). Per the concept of acceptable change limits, platelet function was considered stable in time course if the median relative deviation from baseline values at a given time was <19% or <28%, depending on the test used. By applying our analyte stability limits, we found that blood samples could be stored for up to 5 h when performing the Collagen/ADP, ASPI, ADP, TRAP, Aspirin, and PRU tests. However, we observed significant deviations in the Collagen/EPI test results over time. The PFA-200 Collagen/EPI test results showed a median percentage deviation of 35% after 5 h of storage. However, no significant differences were observed between the measurements taken after 4 and 5 h of storage. We assume, that the baseline values of the Collagen/EPI test are measured incorrectly (presumably owing to malfunction of the test cartridges) and the following results are measured correctly.</p>","PeriodicalId":21474,"journal":{"name":"Scandinavian Journal of Clinical & Laboratory Investigation","volume":" ","pages":"117-125"},"PeriodicalIF":1.4000,"publicationDate":"2026-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Scandinavian Journal of Clinical & Laboratory Investigation","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/00365513.2026.2630329","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/2/14 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"MEDICINE, RESEARCH & EXPERIMENTAL","Score":null,"Total":0}
引用次数: 0
Abstract
In the present study, we aimed to explore the effects of extended storage of blood samples from 30 patients on the results of platelet function testing using the PFA-200 device (Collagen/EPI and Collagen/ADP tests), Multiplate instrument (ASPI, ADP, and TRAP tests), and VerifyNow analyzer (Aspirin and PRU tests). We determined platelet function (1) immediately after blood collection (baseline), (2) after storage of blood samples at room temperature for 3-4 h (i.e. the maximum sample storage period permitted by the manufacturers), and (3) after storage of blood samples for 5 h at room temperature (extended storage). Per the concept of acceptable change limits, platelet function was considered stable in time course if the median relative deviation from baseline values at a given time was <19% or <28%, depending on the test used. By applying our analyte stability limits, we found that blood samples could be stored for up to 5 h when performing the Collagen/ADP, ASPI, ADP, TRAP, Aspirin, and PRU tests. However, we observed significant deviations in the Collagen/EPI test results over time. The PFA-200 Collagen/EPI test results showed a median percentage deviation of 35% after 5 h of storage. However, no significant differences were observed between the measurements taken after 4 and 5 h of storage. We assume, that the baseline values of the Collagen/EPI test are measured incorrectly (presumably owing to malfunction of the test cartridges) and the following results are measured correctly.
期刊介绍:
The Scandinavian Journal of Clinical and Laboratory Investigation is an international scientific journal covering clinically oriented biochemical and physiological research. Since the launch of the journal in 1949, it has been a forum for international laboratory medicine, closely related to, and edited by, The Scandinavian Society for Clinical Chemistry.
The journal contains peer-reviewed articles, editorials, invited reviews, and short technical notes, as well as several supplements each year. Supplements consist of monographs, and symposium and congress reports covering subjects within clinical chemistry and clinical physiology.