Hongyu Ning , Yilei Fan , Haodong Wang , Huijun Liu , Zhongping Huang , Haixing Wang , Xing Ke , Yu Xu
{"title":"Preparation of ZIF-8 grafted magnetic solid-phase extraction sorbent for sensitive determination of four cannabinoids in urine and oral fluid","authors":"Hongyu Ning , Yilei Fan , Haodong Wang , Huijun Liu , Zhongping Huang , Haixing Wang , Xing Ke , Yu Xu","doi":"10.1016/j.aca.2026.345194","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>The global increase in cannabis use has created a strong demand for reliable and efficient detection methods. Traditional urine-based assays are widely used and standardized, but suffer from complex hydrolysis requirements, long detection windows, and an inability to monitor recent use. In contrast, oral fluid testing enables non-invasive, real-time detection of parent cannabinoids, but is limited by low sample volume, matrix interference, and insufficient sensitivity in common rapid tests. It is clear that a sample preparation and detection approach is needed that is robust, highly sensitive, streamlined, and compatible with high-throughput confirmatory analysis such as LC-MS/MS.</div></div><div><h3>Results</h3><div>In this study, a magnetic solid-phase extraction (MSPE) method was developed using a novel magnetic metal–organic framework composite, Fe<sub>3</sub>O<sub>4</sub>@poly(GMA/DVB-ZIF-8), for the efficient extraction of four cannabinoids from urine and oral fluid. The sorbent uses multiple interactions—including hydrophobic, π–π, electrostatic, and Zn–O coordination—to achieve high analyte recovery (exceeding 85%). Coupled with UHPLC-MS/MS analysis, the method demonstrates wide linearity (0.05–25 ng/mL, r<sup>2</sup> > 0.9945), low detection limits (as low as 0.017 ng/mL), and good repeatability (RSDs <13.8%). Comparative evaluation with conventional liquid–liquid extraction showed minimal deviation (relative percent difference <6.0%), confirming high trueness and reliability for cannabinoid monitoring in biological samples.</div></div><div><h3>Significance</h3><div>This MSPE approach provides a simple, efficient and high-throughput sample preparation method for reliable and rapid forensic analysis of cannabinoids in complex biological samples.</div></div>","PeriodicalId":240,"journal":{"name":"Analytica Chimica Acta","volume":"1394 ","pages":"Article 345194"},"PeriodicalIF":6.1000,"publicationDate":"2026-04-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Analytica Chimica Acta","FirstCategoryId":"92","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0003267026001443","RegionNum":2,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/2/10 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0
Abstract
Background
The global increase in cannabis use has created a strong demand for reliable and efficient detection methods. Traditional urine-based assays are widely used and standardized, but suffer from complex hydrolysis requirements, long detection windows, and an inability to monitor recent use. In contrast, oral fluid testing enables non-invasive, real-time detection of parent cannabinoids, but is limited by low sample volume, matrix interference, and insufficient sensitivity in common rapid tests. It is clear that a sample preparation and detection approach is needed that is robust, highly sensitive, streamlined, and compatible with high-throughput confirmatory analysis such as LC-MS/MS.
Results
In this study, a magnetic solid-phase extraction (MSPE) method was developed using a novel magnetic metal–organic framework composite, Fe3O4@poly(GMA/DVB-ZIF-8), for the efficient extraction of four cannabinoids from urine and oral fluid. The sorbent uses multiple interactions—including hydrophobic, π–π, electrostatic, and Zn–O coordination—to achieve high analyte recovery (exceeding 85%). Coupled with UHPLC-MS/MS analysis, the method demonstrates wide linearity (0.05–25 ng/mL, r2 > 0.9945), low detection limits (as low as 0.017 ng/mL), and good repeatability (RSDs <13.8%). Comparative evaluation with conventional liquid–liquid extraction showed minimal deviation (relative percent difference <6.0%), confirming high trueness and reliability for cannabinoid monitoring in biological samples.
Significance
This MSPE approach provides a simple, efficient and high-throughput sample preparation method for reliable and rapid forensic analysis of cannabinoids in complex biological samples.
期刊介绍:
Analytica Chimica Acta has an open access mirror journal Analytica Chimica Acta: X, sharing the same aims and scope, editorial team, submission system and rigorous peer review.
Analytica Chimica Acta provides a forum for the rapid publication of original research, and critical, comprehensive reviews dealing with all aspects of fundamental and applied modern analytical chemistry. The journal welcomes the submission of research papers which report studies concerning the development of new and significant analytical methodologies. In determining the suitability of submitted articles for publication, particular scrutiny will be placed on the degree of novelty and impact of the research and the extent to which it adds to the existing body of knowledge in analytical chemistry.