3D printing enabled biomechanical evaluation of a novel expandable wedge spacer for atlantoaxial reduction.

IF 3.1 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Chih-Chang Chang, Shao-Fu Huang, Rong-Chen Lin, Chun-Li Lin
{"title":"3D printing enabled biomechanical evaluation of a novel expandable wedge spacer for atlantoaxial reduction.","authors":"Chih-Chang Chang, Shao-Fu Huang, Rong-Chen Lin, Chun-Li Lin","doi":"10.1186/s41205-026-00314-0","DOIUrl":null,"url":null,"abstract":"<p><p>Atlantoaxial subluxation (AAS) is a severe instability of the craniovertebral junction that may cause progressive spinal cord compression and neurological impairment. Achieving complete reduction remains challenging when anterior inclination deformity of the atlas limits posterior traction. This study introduces a novel titanium alloy expandable wedge-shaped spacer designed to correct the anterior inclination of the atlantoaxial joint (AAJ) and facilitate controlled distraction and reduction. The implant integrates a two-stage expansion mechanism that enables sequential anterior elevation and fixation stability. To overcome the scarcity of cadaveric specimens, a patient-specific 3D-printed C1–C2 model was reconstructed from clinical CT data to replicate pathological morphology, simulate ligament behavior, and provide a reproducible biomechanical testing platform. Static and fatigue compression tests, performed according to ASTM F2077 and ISO 23089-2, demonstrated superior load-bearing strength (6725 ± 123 N) and long-term durability over 5 million cycles at 1650 N, with an expansion torque of only 52.56 ± 7.54 N·mm. Reduction experiments using the 3D-printed platform showed that the spacer decreased the required traction force by approximately 70% and redirected the reduction trajectory toward the physiological axis. The integration of medical imaging and high-resolution 3D printing enabled precise replication of patient-specific anatomy and standardized mechanical evaluation that is otherwise unachievable with cadaveric models. This study highlights how 3D printing enables patient-specific modeling, quantitative validation, and iterative optimization of implant design—illustrating its critical role in bridging preclinical testing and clinical translation for complex spinal applications.</p>","PeriodicalId":72036,"journal":{"name":"3D printing in medicine","volume":" ","pages":"4"},"PeriodicalIF":3.1000,"publicationDate":"2026-02-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12874940/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"3D printing in medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s41205-026-00314-0","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Atlantoaxial subluxation (AAS) is a severe instability of the craniovertebral junction that may cause progressive spinal cord compression and neurological impairment. Achieving complete reduction remains challenging when anterior inclination deformity of the atlas limits posterior traction. This study introduces a novel titanium alloy expandable wedge-shaped spacer designed to correct the anterior inclination of the atlantoaxial joint (AAJ) and facilitate controlled distraction and reduction. The implant integrates a two-stage expansion mechanism that enables sequential anterior elevation and fixation stability. To overcome the scarcity of cadaveric specimens, a patient-specific 3D-printed C1–C2 model was reconstructed from clinical CT data to replicate pathological morphology, simulate ligament behavior, and provide a reproducible biomechanical testing platform. Static and fatigue compression tests, performed according to ASTM F2077 and ISO 23089-2, demonstrated superior load-bearing strength (6725 ± 123 N) and long-term durability over 5 million cycles at 1650 N, with an expansion torque of only 52.56 ± 7.54 N·mm. Reduction experiments using the 3D-printed platform showed that the spacer decreased the required traction force by approximately 70% and redirected the reduction trajectory toward the physiological axis. The integration of medical imaging and high-resolution 3D printing enabled precise replication of patient-specific anatomy and standardized mechanical evaluation that is otherwise unachievable with cadaveric models. This study highlights how 3D printing enables patient-specific modeling, quantitative validation, and iterative optimization of implant design—illustrating its critical role in bridging preclinical testing and clinical translation for complex spinal applications.

Abstract Image

Abstract Image

Abstract Image

3D打印可用于寰枢复位的新型可膨胀楔形垫片的生物力学评估。
寰枢椎半脱位(AAS)是一种严重的颅椎交界处不稳定,可导致进行性脊髓压迫和神经功能障碍。当寰椎前倾畸形限制了后路牵引时,实现完全复位仍然具有挑战性。本研究介绍了一种新型钛合金可伸缩楔形垫片,用于矫正寰枢关节(AAJ)前倾,促进可控牵张和复位。植入物集成了一个两阶段的扩张机制,可实现连续前路抬高和固定稳定性。为了克服尸体标本的稀缺性,根据临床CT数据重建患者特异性3d打印C1-C2模型,以复制病理形态,模拟韧带行为,并提供可复制的生物力学测试平台。根据ASTM F2077和ISO 23089-2进行的静态和疲劳压缩试验显示,该材料具有优异的承载强度(6725±123 N)和在1650 N下超过500万次的长期耐久性,膨胀扭矩仅为52.56±7.54 N·mm。利用3d打印平台进行的减速实验表明,该垫片将所需的牵引力降低了约70%,并将减速轨迹重新定向到生理轴上。医学成像和高分辨率3D打印的集成使患者特定解剖结构的精确复制和标准化的机械评估成为可能,否则尸体模型无法实现。本研究强调了3D打印如何实现患者特异性建模、定量验证和植入物设计的迭代优化,并说明了其在连接复杂脊柱应用的临床前测试和临床转化方面的关键作用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
审稿时长
5 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书