REPEATIBILITY AND RELIABILITY OF KNEE JOINT ROTATIONAL LAXITY MEASUREMENTS WITH DOUBLE GONIOMETRY.

IF 0.5 4区 医学 Q3 MEDICINE, GENERAL & INTERNAL
Alan Albanese, Miroljub Jakovljević
{"title":"REPEATIBILITY AND RELIABILITY OF KNEE JOINT ROTATIONAL LAXITY MEASUREMENTS WITH DOUBLE GONIOMETRY.","authors":"Alan Albanese, Miroljub Jakovljević","doi":"10.20471/acc.2025.64.02.11","DOIUrl":null,"url":null,"abstract":"<p><p>The results of existing studies on the range of motion in the knee joint rotation are contradictory. The best method for measuring knee rotation has not yet been found. The purpose of this study was to investigate the repeatability and reliability of knee joint rotational laxity measurement using two magnetic goniometers. Thirty healthy subjects (15 males and 15 females) aged between 20 and 32 years participated in the study. Internal and external rotation of the left and right knee joints were measured with double goniometry in the sitting position with 90° flexion of the knee. There was no statistically significant difference in measurements between first and second session. The intraclass correlation coefficient (ICC) for repeatability ranged from 0.812 to 0.858 and for reliability from 0.655 to 0.837. Within-subject coefficient of variation was less than 10%, standard error of measurement was between 1.4<sup>o</sup> and 2.0<sup>o</sup>, minimal detectable change was between 2.5<sup>o</sup> and 3.8<sup>o</sup>. The measured values of internal and external rotation did not completely agree with those reported by other authors. The ICC values indicated good to very good repeatability and reliability of the double goniometry measurement of rotational knee laxity. Technical improvements will be required before translation into clinical practice.</p>","PeriodicalId":7072,"journal":{"name":"Acta clinica Croatica","volume":"64 2","pages":"271-279"},"PeriodicalIF":0.5000,"publicationDate":"2025-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12863343/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta clinica Croatica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.20471/acc.2025.64.02.11","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0

Abstract

The results of existing studies on the range of motion in the knee joint rotation are contradictory. The best method for measuring knee rotation has not yet been found. The purpose of this study was to investigate the repeatability and reliability of knee joint rotational laxity measurement using two magnetic goniometers. Thirty healthy subjects (15 males and 15 females) aged between 20 and 32 years participated in the study. Internal and external rotation of the left and right knee joints were measured with double goniometry in the sitting position with 90° flexion of the knee. There was no statistically significant difference in measurements between first and second session. The intraclass correlation coefficient (ICC) for repeatability ranged from 0.812 to 0.858 and for reliability from 0.655 to 0.837. Within-subject coefficient of variation was less than 10%, standard error of measurement was between 1.4o and 2.0o, minimal detectable change was between 2.5o and 3.8o. The measured values of internal and external rotation did not completely agree with those reported by other authors. The ICC values indicated good to very good repeatability and reliability of the double goniometry measurement of rotational knee laxity. Technical improvements will be required before translation into clinical practice.

Abstract Image

双角度测量法测量膝关节旋转松弛度的重复性和可靠性。
现有的关于膝关节旋转活动范围的研究结果是矛盾的。目前还没有找到测量膝关节旋转的最佳方法。本研究的目的是探讨使用两个磁测角仪测量膝关节旋转松弛度的可重复性和可靠性。30名年龄在20至32岁之间的健康受试者(男15名,女15名)参加了这项研究。膝关节屈曲90°时,采用双测角法测量左、右膝关节的内外旋。第一阶段和第二阶段的测量结果没有统计学上的显著差异。类内相关系数(ICC)重复性为0.812 ~ 0.858,信度为0.655 ~ 0.837。受试者内变异系数小于10%,测量标准误差在1.40 ~ 2.0之间,最小可检测变化在2.50 ~ 3.80之间。内旋和外旋的测量值与其他作者报道的不完全一致。ICC值表明双角度测量旋转膝关节松弛度的重复性和可靠性良好。在转化为临床实践之前,需要进行技术改进。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Acta clinica Croatica
Acta clinica Croatica 医学-医学:内科
CiteScore
1.10
自引率
16.70%
发文量
38
审稿时长
6-12 weeks
期刊介绍: Acta Clinica Croatica is a peer reviewed general medical journal that publishes original articles that advance and improve medical science and practice and that serve the purpose of transfer of original and valuable information to journal readers. Acta Clinica Croatica is published in English four times a year.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书