{"title":"Cardiac Safety of Conducted Energy Devices: Electrophysiological mechanisms, risk assessment, and forensic interpretation","authors":"Charles D. Deakin","doi":"10.1016/j.jflm.2026.103082","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Conducted Energy Devices (CEDs) or Conducted Energy Weapons (CEWs), among which the best known are the TASER® brand, are increasingly deployed by law-enforcement agencies. Their established cardiac safety is overshadowed by temporally-associated sudden deaths, which are often directly attributed to CED discharge. Concern remains that in rare cases, high-voltage CED pulses might directly induce fatal cardiac arrhythmias.</div></div><div><h3>Methods</h3><div>This review considers electrophysiological evidence, computational modelling, and forensic case data to understand the mechanisms governing cardiac stimulation during CED exposure. It discusses strength–duration relationships, differential excitement of neuromuscular and myocardial tissues, and compares predicted electrical stimulation thresholds with those determined by finite element modelling. This evidence is compared with clinical evidence of CED-induced malignant arrythmias.</div></div><div><h3>Findings</h3><div>CED pulses of 50–100 μs are both too brief and too weak to depolarise cardiac myocytes under normal anatomical conditions. Finite-element and finite-integration models show typical myocardial current densities of 0.27–0.6 mA cm<sup>−2</sup> which are more than two orders of magnitude below the short-pulse VF threshold of ∼100–150 mA cm<sup>−2</sup> (≈4 mA cm<sup>−2</sup> RMS). Layered thoracic modelling demonstrates that muscle anisotropy and chest-wall thickness diverts almost 90 % of current laterally, producing cardiac field strengths far beneath depolarisation levels. Rare trans-cardiac probe alignments or thin chest walls, however, may approach capture thresholds and therefore cannot be considered completely risk-free.</div></div><div><h3>Conclusions</h3><div>Current evidence indicates that in all but extreme probe geometries or paediatric exposures, CED are at minimal risk of inducing malignant arrhythmias. Forensic assessment of CED-related deaths should consider electrophysiological principles with probe placement, the timing of the collapse, and toxicological findings in order to distinguish coincidental from causally related events.</div></div>","PeriodicalId":16098,"journal":{"name":"Journal of forensic and legal medicine","volume":"118 ","pages":"Article 103082"},"PeriodicalIF":1.7000,"publicationDate":"2026-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of forensic and legal medicine","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1752928X26000144","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/21 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"MEDICINE, LEGAL","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Conducted Energy Devices (CEDs) or Conducted Energy Weapons (CEWs), among which the best known are the TASER® brand, are increasingly deployed by law-enforcement agencies. Their established cardiac safety is overshadowed by temporally-associated sudden deaths, which are often directly attributed to CED discharge. Concern remains that in rare cases, high-voltage CED pulses might directly induce fatal cardiac arrhythmias.
Methods
This review considers electrophysiological evidence, computational modelling, and forensic case data to understand the mechanisms governing cardiac stimulation during CED exposure. It discusses strength–duration relationships, differential excitement of neuromuscular and myocardial tissues, and compares predicted electrical stimulation thresholds with those determined by finite element modelling. This evidence is compared with clinical evidence of CED-induced malignant arrythmias.
Findings
CED pulses of 50–100 μs are both too brief and too weak to depolarise cardiac myocytes under normal anatomical conditions. Finite-element and finite-integration models show typical myocardial current densities of 0.27–0.6 mA cm−2 which are more than two orders of magnitude below the short-pulse VF threshold of ∼100–150 mA cm−2 (≈4 mA cm−2 RMS). Layered thoracic modelling demonstrates that muscle anisotropy and chest-wall thickness diverts almost 90 % of current laterally, producing cardiac field strengths far beneath depolarisation levels. Rare trans-cardiac probe alignments or thin chest walls, however, may approach capture thresholds and therefore cannot be considered completely risk-free.
Conclusions
Current evidence indicates that in all but extreme probe geometries or paediatric exposures, CED are at minimal risk of inducing malignant arrhythmias. Forensic assessment of CED-related deaths should consider electrophysiological principles with probe placement, the timing of the collapse, and toxicological findings in order to distinguish coincidental from causally related events.
背景传导能量装置(ced)或传导能量武器(CEWs),其中最著名的是泰瑟®品牌,越来越多地被执法机构部署。他们建立的心脏安全性被暂时性猝死所掩盖,后者通常直接归因于CED出院。在极少数情况下,高压CED脉冲可能直接诱发致命性心律失常。方法本综述考虑了电生理证据、计算模型和法医案例数据,以了解在CED暴露期间控制心脏刺激的机制。它讨论了强度-持续时间的关系,神经肌肉和心肌组织的不同兴奋,并将预测的电刺激阈值与有限元模型确定的阈值进行了比较。这一证据与ced诱发的恶性心律失常的临床证据进行了比较。发现在正常解剖条件下,50-100 μs的sced脉冲既短又弱,无法使心肌细胞去极化。有限元和有限积分模型显示,典型的心肌电流密度为0.27-0.6 mA cm - 2,比短脉冲VF阈值约100-150 mA cm - 2 (RMS≈4 mA cm - 2)低两个数量级以上。分层胸腔模型表明,肌肉各向异性和胸壁厚度将近90%的电流侧向转移,产生远低于去极化水平的心脏场强度。然而,罕见的经心探头排列或胸壁薄可能接近捕获阈值,因此不能认为完全没有风险。结论目前的证据表明,除了极端探头几何形状或儿科暴露外,CED诱发恶性心律失常的风险很小。对ed相关死亡的法医评估应考虑探针放置的电生理学原理、崩溃的时间和毒理学结果,以区分偶然事件和因果相关事件。
期刊介绍:
The Journal of Forensic and Legal Medicine publishes topical articles on aspects of forensic and legal medicine. Specifically the Journal supports research that explores the medical principles of care and forensic assessment of individuals, whether adult or child, in contact with the judicial system. It is a fully peer-review hybrid journal with a broad international perspective.