Beyond CTDIvol: Patient-specific SSDE and organ dose assessment in routine adult CT practice

IF 3.5 3区 物理与天体物理 Q3 CHEMISTRY, PHYSICAL
Radiation Physics and Chemistry Pub Date : 2026-06-01 Epub Date: 2026-01-21 DOI:10.1016/j.radphyschem.2026.113658
Omar Bentiane, Omaima Khettabi, Omar Berradi, El Mehdi Sadiki, Fatimaezzahra Bouzzit, Rodouan Touti
{"title":"Beyond CTDIvol: Patient-specific SSDE and organ dose assessment in routine adult CT practice","authors":"Omar Bentiane,&nbsp;Omaima Khettabi,&nbsp;Omar Berradi,&nbsp;El Mehdi Sadiki,&nbsp;Fatimaezzahra Bouzzit,&nbsp;Rodouan Touti","doi":"10.1016/j.radphyschem.2026.113658","DOIUrl":null,"url":null,"abstract":"<div><div>Computed Tomography (CT) is a key diagnostic tool in medical imaging, but concerns persist regarding radiation exposure, particularly in repeat or high-dose examinations. The CTDI<sub>vol</sub> quantifies scanner output but does not accurately represent the patient dose. The Size-Specific Dose Estimate (SSDE) provides a more clinically relevant patient-specific dose estimate by accounting for body size and composition. This study aims to quantify the differences between CTDI<sub>vol</sub> and SSDE in adult thoraco-abdominopelvic and abdominopelvic CT scans, compare manual and automatic methods of patient size estimation, and evaluate organ doses. A retrospective review was conducted on 52 adult CT scans (thoraco-abdominopelvic and abdominopelvic). Four patient-size metrics were collected: the manually measured effective diameter (D<sub>eff-M</sub>), the automatically averaged z-axis effective diameter (D<sub>eff-AZ</sub>), and two water-equivalent diameters (D<sub>w-E</sub> and D<sub>w-AZ</sub>). The SSDE values were calculated according to AAPM TG-204 and TG-220. The organ doses were estimated with <em>IndoseCT</em> based on Monte Carlo-derived correlations. The mean CTDI<sub>vol</sub> was 13.0 ± 3.6 mGy, with SSDE values 31–37 % higher across all methods. D<sub>eff-AZ</sub> was 4.7 % greater than D<sub>eff-M</sub>, while D<sub>w-E</sub> and D<sub>w-AZ</sub> showed near-perfect agreement. The liver, kidneys, and bladder received the highest doses (&gt;15 mGy) as they are directly irradiated, whereas radiosensitive organs outside the primary scan field, such as the thyroid and eyes, received measurable scatter doses. SSDE provides a more accurate representation of patient radiation dose than CTDI<sub>vol</sub> and should be integrated into routine CT protocols. Among the evaluated size metrics, attenuation-based D<sub>w</sub> proved to be the most robust and reproducible. Incorporating organ-dose estimations and SSDE into clinical practice can enhance patient safety, optimize imaging protocols, and support compliance with radiation protection regulations.</div></div>","PeriodicalId":20861,"journal":{"name":"Radiation Physics and Chemistry","volume":"243 ","pages":"Article 113658"},"PeriodicalIF":3.5000,"publicationDate":"2026-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiation Physics and Chemistry","FirstCategoryId":"92","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0969806X2600068X","RegionNum":3,"RegionCategory":"物理与天体物理","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/21 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CHEMISTRY, PHYSICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Computed Tomography (CT) is a key diagnostic tool in medical imaging, but concerns persist regarding radiation exposure, particularly in repeat or high-dose examinations. The CTDIvol quantifies scanner output but does not accurately represent the patient dose. The Size-Specific Dose Estimate (SSDE) provides a more clinically relevant patient-specific dose estimate by accounting for body size and composition. This study aims to quantify the differences between CTDIvol and SSDE in adult thoraco-abdominopelvic and abdominopelvic CT scans, compare manual and automatic methods of patient size estimation, and evaluate organ doses. A retrospective review was conducted on 52 adult CT scans (thoraco-abdominopelvic and abdominopelvic). Four patient-size metrics were collected: the manually measured effective diameter (Deff-M), the automatically averaged z-axis effective diameter (Deff-AZ), and two water-equivalent diameters (Dw-E and Dw-AZ). The SSDE values were calculated according to AAPM TG-204 and TG-220. The organ doses were estimated with IndoseCT based on Monte Carlo-derived correlations. The mean CTDIvol was 13.0 ± 3.6 mGy, with SSDE values 31–37 % higher across all methods. Deff-AZ was 4.7 % greater than Deff-M, while Dw-E and Dw-AZ showed near-perfect agreement. The liver, kidneys, and bladder received the highest doses (>15 mGy) as they are directly irradiated, whereas radiosensitive organs outside the primary scan field, such as the thyroid and eyes, received measurable scatter doses. SSDE provides a more accurate representation of patient radiation dose than CTDIvol and should be integrated into routine CT protocols. Among the evaluated size metrics, attenuation-based Dw proved to be the most robust and reproducible. Incorporating organ-dose estimations and SSDE into clinical practice can enhance patient safety, optimize imaging protocols, and support compliance with radiation protection regulations.
超越CTDIvol:常规成人CT实践中患者特异性SSDE和器官剂量评估
计算机断层扫描(CT)是医学成像中的关键诊断工具,但对辐射暴露的担忧持续存在,特别是在重复或高剂量检查中。CTDIvol量化扫描仪输出,但不能准确代表患者剂量。尺寸特异性剂量估计(SSDE)通过考虑身体尺寸和组成提供了更具临床相关性的患者特异性剂量估计。本研究旨在量化CTDIvol和SSDE在成人胸腹骨盆和腹部骨盆CT扫描中的差异,比较人工和自动估计患者大小的方法,并评估器官剂量。回顾性分析了52例成人CT扫描(胸腹骨盆和腹部骨盆)。收集4个患者尺寸指标:人工测量的有效直径(def - m)、自动平均z轴有效直径(def - az)和两个水当量直径(Dw-E和Dw-AZ)。根据AAPM TG-204和TG-220计算SSDE值。使用IndoseCT根据蒙特卡罗导出的相关性估计器官剂量。CTDIvol平均值为13.0±3.6 mGy, SSDE值比所有方法高31 - 37%。def - az比def - m高4.7%,而Dw-E和Dw-AZ表现出近乎完美的一致性。肝脏、肾脏和膀胱受到的辐射剂量最高(15毫戈瑞),因为它们直接受到辐射,而初级扫描场外的辐射敏感器官,如甲状腺和眼睛,则受到可测量的散射剂量。SSDE提供比CTDIvol更准确的患者辐射剂量表示,应纳入常规CT方案。在评估的尺寸指标中,基于衰减的Dw被证明是最稳健和可重复性的。将器官剂量评估和SSDE纳入临床实践可以提高患者安全,优化成像方案,并支持遵守辐射防护法规。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Radiation Physics and Chemistry
Radiation Physics and Chemistry 化学-核科学技术
CiteScore
5.60
自引率
17.20%
发文量
574
审稿时长
12 weeks
期刊介绍: Radiation Physics and Chemistry is a multidisciplinary journal that provides a medium for publication of substantial and original papers, reviews, and short communications which focus on research and developments involving ionizing radiation in radiation physics, radiation chemistry and radiation processing. The journal aims to publish papers with significance to an international audience, containing substantial novelty and scientific impact. The Editors reserve the rights to reject, with or without external review, papers that do not meet these criteria. This could include papers that are very similar to previous publications, only with changed target substrates, employed materials, analyzed sites and experimental methods, report results without presenting new insights and/or hypothesis testing, or do not focus on the radiation effects.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书