High-Z Contrast Media for Coronary Photon-counting Detector CT Angiography: Improved Quantification of Calcified Stenoses.

IF 7.9 1区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Tristan T Demmert, Konstantin Klambauer, Bernhard Schmidt, Victor Mergen, Lukas J Moser, Philipp N Maintz, Thomas Allmendinger, Thomas Flohr, Hubertus Pietsch, Matthias Eberhard, Hatem Alkadhi
{"title":"High-Z Contrast Media for Coronary Photon-counting Detector CT Angiography: Improved Quantification of Calcified Stenoses.","authors":"Tristan T Demmert, Konstantin Klambauer, Bernhard Schmidt, Victor Mergen, Lukas J Moser, Philipp N Maintz, Thomas Allmendinger, Thomas Flohr, Hubertus Pietsch, Matthias Eberhard, Hatem Alkadhi","doi":"10.1097/RLI.0000000000001262","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Blooming artifacts from calcified plaques can obscure the vessel lumen, leading to overestimation of stenosis severity. Spectral coronary angiography with photon-counting detector CT (PCD-CT) provides virtual monoenergetic images (VMIs) for coronary artery disease assessment. While VMIs at high VMI energy levels reduce calcium blooming, iodine contrast is diminished, limiting diagnostic value. This study evaluated whether contrast media with an atomic number higher than iodine (high-Z) preserve vascular contrast using high VMI energy levels, thereby improving the accuracy of stenosis quantification.</p><p><strong>Methods: </strong>A phantom with 4 and 6 mm diameter rods to mimic small diameter vessels containing eccentric calcified plaques causing 25%, 50%, and 75% diameter stenoses was scanned with a dual-source PCD-CT system. Five different contrast media, including iodine, tungsten, holmium, hafnium, and bismuth, were tested. VMIs were reconstructed from 40 to 190 keV in 1-keV steps. Vessel attenuation, contrast-to-noise ratio (CNR), and stenoses were measured. Qualitative assessment of image quality was performed.</p><p><strong>Results: </strong>Iodine attenuation was high at lower VMI energy levels and dropped below 250 HU at >100 keV. Tungsten, holmium, hafnium, and bismuth maintained >250 HU attenuation throughout the entire energy range. Vessel CNR of iodine was high at lower and decreased at higher VMI energy levels, similar to the CNR of holmium and bismuth, though to a lesser extent. In distinction, CNRs of tungsten and hafnium were lower at lower VMI energy levels and increased to a relatively constant level at higher keV. Tungsten CNR increased with energy, approaching ~40 at high keV. Across all contrast media and stenosis degrees, stenoses were overestimated on low VMI energy levels (24% to 32.5% at 40 keV), while the degree of overestimation decreased at higher VMI energy levels (0% to 13.5% at 190 keV). At 190 keV, tungsten, hafnium, and bismuth showed ≤2.5% stenosis overestimation, compared with iodine (10% to 13.5%). Image quality varied between contrast media and energy levels: new very high-Z contrast media achieved higher scores, while iodine peaked at lower keV (55 to 70 keV) and, due to loss of contrast at higher energies, received the lowest overall scores.</p><p><strong>Conclusions: </strong>As compared with iodine, very high-Z contrast media enable superior lumen definition and more accurate stenosis assessment, also at high VMI energy levels, which minimize calcium blooming.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":"568-574"},"PeriodicalIF":7.9000,"publicationDate":"2026-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13317894/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Investigative Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/RLI.0000000000001262","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Blooming artifacts from calcified plaques can obscure the vessel lumen, leading to overestimation of stenosis severity. Spectral coronary angiography with photon-counting detector CT (PCD-CT) provides virtual monoenergetic images (VMIs) for coronary artery disease assessment. While VMIs at high VMI energy levels reduce calcium blooming, iodine contrast is diminished, limiting diagnostic value. This study evaluated whether contrast media with an atomic number higher than iodine (high-Z) preserve vascular contrast using high VMI energy levels, thereby improving the accuracy of stenosis quantification.

Methods: A phantom with 4 and 6 mm diameter rods to mimic small diameter vessels containing eccentric calcified plaques causing 25%, 50%, and 75% diameter stenoses was scanned with a dual-source PCD-CT system. Five different contrast media, including iodine, tungsten, holmium, hafnium, and bismuth, were tested. VMIs were reconstructed from 40 to 190 keV in 1-keV steps. Vessel attenuation, contrast-to-noise ratio (CNR), and stenoses were measured. Qualitative assessment of image quality was performed.

Results: Iodine attenuation was high at lower VMI energy levels and dropped below 250 HU at >100 keV. Tungsten, holmium, hafnium, and bismuth maintained >250 HU attenuation throughout the entire energy range. Vessel CNR of iodine was high at lower and decreased at higher VMI energy levels, similar to the CNR of holmium and bismuth, though to a lesser extent. In distinction, CNRs of tungsten and hafnium were lower at lower VMI energy levels and increased to a relatively constant level at higher keV. Tungsten CNR increased with energy, approaching ~40 at high keV. Across all contrast media and stenosis degrees, stenoses were overestimated on low VMI energy levels (24% to 32.5% at 40 keV), while the degree of overestimation decreased at higher VMI energy levels (0% to 13.5% at 190 keV). At 190 keV, tungsten, hafnium, and bismuth showed ≤2.5% stenosis overestimation, compared with iodine (10% to 13.5%). Image quality varied between contrast media and energy levels: new very high-Z contrast media achieved higher scores, while iodine peaked at lower keV (55 to 70 keV) and, due to loss of contrast at higher energies, received the lowest overall scores.

Conclusions: As compared with iodine, very high-Z contrast media enable superior lumen definition and more accurate stenosis assessment, also at high VMI energy levels, which minimize calcium blooming.

冠状动脉光子计数检测器CT血管造影的高z造影剂:改进钙化狭窄的量化。
背景:钙化斑块形成的虚影会模糊血管腔,导致对狭窄严重程度的高估。光谱冠状动脉造影与光子计数检测器CT (PCD-CT)提供虚拟单能图像(vmi)的冠状动脉疾病的评估。高VMI能量水平的VMI降低钙盛开,碘造影剂降低,限制了诊断价值。本研究评估了原子序数高于碘(高z)的造影剂是否能使用高VMI能级保持血管造影剂,从而提高狭窄定量的准确性。方法:采用双源PCD-CT系统扫描直径为4和6mm的假体,模拟小直径血管中含有导致25%、50%和75%直径狭窄的偏心钙化斑块。测试了五种不同的造影剂,包括碘、钨、钬、铪和铋。vmi以1 keV为步长,从40 keV重构至190 keV。测量血管衰减、噪声对比比(CNR)和狭窄程度。对图像质量进行定性评价。结果:碘在较低的VMI能级下衰减较大,在> - 100 keV时降至250 HU以下。钨、钬、铪和铋在整个能量范围内保持bbb250 HU衰减。碘的血管CNR在较低的VMI能级时高,在较高的VMI能级时降低,与钬和铋的CNR相似,但程度较轻。相比之下,钨和铪的cnr在较低的VMI能级下较低,在较高的keV能级下增加到一个相对恒定的水平。钨的CNR随能量的增加而增加,在高keV时接近~40。在所有造影剂和狭窄程度中,低VMI能量水平的狭窄被高估(40 keV时为24%至32.5%),而高VMI能量水平的狭窄被高估的程度下降(190 keV时为0%至13.5%)。在190kev时,钨、铪和铋的狭窄度高估≤2.5%,而碘的狭窄度高估为10% ~ 13.5%。造影剂和能量水平之间的图像质量有所不同:新的高z造影剂获得了更高的分数,而碘在较低的keV(55至70 keV)达到峰值,并且由于在较高能量下对比度的损失,获得了最低的总分。结论:与碘相比,非常高的z造影剂在高VMI能量水平下能够更好地定义管腔,更准确地评估狭窄,从而最大限度地减少钙盛开。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Investigative Radiology
Investigative Radiology 医学-核医学
CiteScore
15.10
自引率
16.40%
发文量
188
审稿时长
4-8 weeks
期刊介绍: Investigative Radiology publishes original, peer-reviewed reports on clinical and laboratory investigations in diagnostic imaging, the diagnostic use of radioactive isotopes, computed tomography, positron emission tomography, magnetic resonance imaging, ultrasound, digital subtraction angiography, and related modalities. Emphasis is on early and timely publication. Primarily research-oriented, the journal also includes a wide variety of features of interest to clinical radiologists.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书