Deep-learning image reconstruction algorithms for CT: A task-based image quality assessment of four CT systems using a phantom

IF 11.1 2区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Diagnostic and Interventional Imaging Pub Date : 2026-07-01 Epub Date: 2026-01-14 DOI:10.1016/j.diii.2026.01.003
Joël Greffier, Alexa Liogier, Maxime Pastor, Fabien de Oliveira, Quentin Chaine, Skander Sammoud, Jean Paul Beregi, Djamel Dabli
{"title":"Deep-learning image reconstruction algorithms for CT: A task-based image quality assessment of four CT systems using a phantom","authors":"Joël Greffier,&nbsp;Alexa Liogier,&nbsp;Maxime Pastor,&nbsp;Fabien de Oliveira,&nbsp;Quentin Chaine,&nbsp;Skander Sammoud,&nbsp;Jean Paul Beregi,&nbsp;Djamel Dabli","doi":"10.1016/j.diii.2026.01.003","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><div>The purpose of this study was to assess the performance of iterative reconstruction (IR) and deep-learning image reconstruction (DLR) algorithms developed by four CT vendors in terms of image quality.</div></div><div><h3>Materials and methods</h3><div>Acquisitions were performed on an image quality phantom at three dose levels (1.8, 6 and 11 mGy) using four CT systems (further referred to as G-CT, P-CT, U-CT, and C-CT). For each CT, raw data were reconstructed using the commonly used soft tissue kernel and level for IR and DLR algorithms. Noise power spectrum and task-based transfer function were computed to assess noise magnitude, noise texture (f<sub>av</sub>) and spatial resolution, respectively. Detectability indexes (<em>d</em>’) were computed to model the detection of two abdominal lesions.</div></div><div><h3>Results</h3><div>Compared to IR, noise magnitude reduction with DLR was similar for all dose levels for G-CT (-21.1 ± 1.5 [standard deviation (SD)] %) and P-CT (-48.4 ± 0.1 [SD] %) but more pronounced at 1.8 mGy and decreased as the dose level increased for U-CT and C-CT. Noise texture was greater with DLR than IR at all dose levels for all CT systems, except for U-CT, which gave similar f<sub>av</sub> values. For both inserts, spatial resolution was better with DLR than with IR for all CT systems, except for the low-contrast insert with C-CT at 1.8 and 6 mGy and P-CT at 1.8 mGy. For both simulated lesions and all dose levels, <em>d</em>’ values were greater with DLR than with IR by 77.5 ± 8.7 (SD) % for C-CT, 33.7 ± 5.6 (SD) % for G-CT, 112.7 ± 4.7 (SD) % for P-CT and from 158.3 % to 546.6 % on average for U-CT.</div></div><div><h3>Conclusion</h3><div>Compared to IR, DLR algorithms reduce the image noise and improve detectability whilst providing similar or better noise texture and spatial resolution.</div></div>","PeriodicalId":48656,"journal":{"name":"Diagnostic and Interventional Imaging","volume":"107 7","pages":"Pages 256-264"},"PeriodicalIF":11.1000,"publicationDate":"2026-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Diagnostic and Interventional Imaging","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2211568426000033","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/14 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose

The purpose of this study was to assess the performance of iterative reconstruction (IR) and deep-learning image reconstruction (DLR) algorithms developed by four CT vendors in terms of image quality.

Materials and methods

Acquisitions were performed on an image quality phantom at three dose levels (1.8, 6 and 11 mGy) using four CT systems (further referred to as G-CT, P-CT, U-CT, and C-CT). For each CT, raw data were reconstructed using the commonly used soft tissue kernel and level for IR and DLR algorithms. Noise power spectrum and task-based transfer function were computed to assess noise magnitude, noise texture (fav) and spatial resolution, respectively. Detectability indexes (d’) were computed to model the detection of two abdominal lesions.

Results

Compared to IR, noise magnitude reduction with DLR was similar for all dose levels for G-CT (-21.1 ± 1.5 [standard deviation (SD)] %) and P-CT (-48.4 ± 0.1 [SD] %) but more pronounced at 1.8 mGy and decreased as the dose level increased for U-CT and C-CT. Noise texture was greater with DLR than IR at all dose levels for all CT systems, except for U-CT, which gave similar fav values. For both inserts, spatial resolution was better with DLR than with IR for all CT systems, except for the low-contrast insert with C-CT at 1.8 and 6 mGy and P-CT at 1.8 mGy. For both simulated lesions and all dose levels, d’ values were greater with DLR than with IR by 77.5 ± 8.7 (SD) % for C-CT, 33.7 ± 5.6 (SD) % for G-CT, 112.7 ± 4.7 (SD) % for P-CT and from 158.3 % to 546.6 % on average for U-CT.

Conclusion

Compared to IR, DLR algorithms reduce the image noise and improve detectability whilst providing similar or better noise texture and spatial resolution.
CT的深度学习图像重建算法:一个基于任务的图像质量评估的四个CT系统使用一个幻影。
目的:本研究的目的是评估四家CT供应商开发的迭代重建(IR)和深度学习图像重建(DLR)算法在图像质量方面的性能。材料和方法:使用四种CT系统(进一步称为G-CT, P-CT, U-CT和C-CT)在三个剂量水平(1.8,6和11 mGy)的图像质量幻象上进行采集。对于每台CT,使用常用的软组织核和水平进行IR和DLR算法重建原始数据。计算噪声功率谱和基于任务的传递函数,分别评估噪声强度、噪声纹理(fav)和空间分辨率。计算可检测性指数(d')来模拟两种腹部病变的检测。结果:与IR相比,DLR在所有剂量水平下对G-CT(-21.1±1.5[标准差(SD)] %)和P-CT(-48.4±0.1 [SD] %)的降噪幅度相似,但在1.8 mGy时更为明显,U-CT和C-CT随着剂量水平的增加而下降。除U-CT外,DLR在所有CT系统的所有剂量水平下的噪声结构都大于IR, U-CT给出了相似的偏好值。除了C-CT 1.8和6 mGy以及P-CT 1.8 mGy的低对比度插入外,所有CT系统中DLR的空间分辨率都优于IR。对于模拟病变和所有剂量水平,DLR的d'值比IR高,C-CT为77.5±8.7 (SD) %, G-CT为33.7±5.6 (SD) %, P-CT为112.7±4.7 (SD) %, U-CT平均为158.3%至546.6%。结论:与红外相比,DLR算法在提供相似或更好的噪声纹理和空间分辨率的同时,降低了图像噪声,提高了可检测性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Diagnostic and Interventional Imaging
Diagnostic and Interventional Imaging Medicine-Radiology, Nuclear Medicine and Imaging
CiteScore
8.50
自引率
29.10%
发文量
126
审稿时长
11 days
期刊介绍: Diagnostic and Interventional Imaging accepts publications originating from any part of the world based only on their scientific merit. The Journal focuses on illustrated articles with great iconographic topics and aims at aiding sharpening clinical decision-making skills as well as following high research topics. All articles are published in English. Diagnostic and Interventional Imaging publishes editorials, technical notes, letters, original and review articles on abdominal, breast, cancer, cardiac, emergency, forensic medicine, head and neck, musculoskeletal, gastrointestinal, genitourinary, interventional, obstetric, pediatric, thoracic and vascular imaging, neuroradiology, nuclear medicine, as well as contrast material, computer developments, health policies and practice, and medical physics relevant to imaging.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书