Workflow, Alarm Notification and Technology Acceptance of Continuous, Wearable-Based Vital Sign Monitoring on A Neurosurgical Ward-A Pilot Feasibility Study.
Joan Alsolivany, Lina Mosch, Lars Wessels, Laura Hallek, Tarik Alp Sargut, Claudia Spies, Peter Vajkoczy, Felix Balzer, Akira Poncette, Nils Hecht
{"title":"Workflow, Alarm Notification and Technology Acceptance of Continuous, Wearable-Based Vital Sign Monitoring on A Neurosurgical Ward-A Pilot Feasibility Study.","authors":"Joan Alsolivany, Lina Mosch, Lars Wessels, Laura Hallek, Tarik Alp Sargut, Claudia Spies, Peter Vajkoczy, Felix Balzer, Akira Poncette, Nils Hecht","doi":"10.1097/ANA.0000000000001094","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Continuous, wearable-based vital sign monitoring can enhance patient safety and reduce intensive care demands but is not routinely used on standard neurosurgical wards. This study assessed alarm events and technology acceptance during the implementation of such a system for postoperative patients in a university hospital's neurosurgical ward.</p><p><strong>Methods: </strong>In this pilot feasibility study, selected patients were continuously monitored for 12 to 24 hours after elective neurosurgery on a 44-bed ward. Phase 1 (June to December 2019) used 12 mobile units with manual alarm documentation; Phase 2 (August 2023 to January 2024) used 44 fixed bedside units with automatic documentation. Alarm patterns and nurse-rated technology acceptance (4-point Likert scale) were analyzed.</p><p><strong>Results: </strong>Vital signs were monitored in 214 patients (median age: 58, IQR: 42 to 70) in Phase 1 and 290 patients (median age: 60, IQR: 43 to 77) in Phase 2. Procedures included cranial, spinal, and peripheral nerve surgeries. Phase 1 recorded 30 nonclinical alarms and no clinical alarms. In contrast, Phase 2 registered 14,500 clinical and 185,744 nonclinical alarms, mostly from detached sensors (177,989/185,744; 96%). Clinical alarms were mainly due to hypoxia (8305/14,500; 57%) and tachycardia (3487/14,500; 30%). Most alarms were acknowledged within 30 seconds (clinical: 12,969/14,500; 90% and nonclinical: 181,447/185,744; 98%), with delayed responses mostly for nonclinical events (3625/4189; 87%). Nurses reported improved convenience, usability, and system connection in Phase 2.</p><p><strong>Conclusions: </strong>Wearable-based vital sign monitoring is feasible on neurosurgical wards, but high nonclinical alarm rates highlight the need to refine alarm management strategies for effective clinical integration.</p>","PeriodicalId":16550,"journal":{"name":"Journal of neurosurgical anesthesiology","volume":" ","pages":"166-171"},"PeriodicalIF":2.7000,"publicationDate":"2026-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12952484/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of neurosurgical anesthesiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/ANA.0000000000001094","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2026/1/13 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ANESTHESIOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Continuous, wearable-based vital sign monitoring can enhance patient safety and reduce intensive care demands but is not routinely used on standard neurosurgical wards. This study assessed alarm events and technology acceptance during the implementation of such a system for postoperative patients in a university hospital's neurosurgical ward.
Methods: In this pilot feasibility study, selected patients were continuously monitored for 12 to 24 hours after elective neurosurgery on a 44-bed ward. Phase 1 (June to December 2019) used 12 mobile units with manual alarm documentation; Phase 2 (August 2023 to January 2024) used 44 fixed bedside units with automatic documentation. Alarm patterns and nurse-rated technology acceptance (4-point Likert scale) were analyzed.
Results: Vital signs were monitored in 214 patients (median age: 58, IQR: 42 to 70) in Phase 1 and 290 patients (median age: 60, IQR: 43 to 77) in Phase 2. Procedures included cranial, spinal, and peripheral nerve surgeries. Phase 1 recorded 30 nonclinical alarms and no clinical alarms. In contrast, Phase 2 registered 14,500 clinical and 185,744 nonclinical alarms, mostly from detached sensors (177,989/185,744; 96%). Clinical alarms were mainly due to hypoxia (8305/14,500; 57%) and tachycardia (3487/14,500; 30%). Most alarms were acknowledged within 30 seconds (clinical: 12,969/14,500; 90% and nonclinical: 181,447/185,744; 98%), with delayed responses mostly for nonclinical events (3625/4189; 87%). Nurses reported improved convenience, usability, and system connection in Phase 2.
Conclusions: Wearable-based vital sign monitoring is feasible on neurosurgical wards, but high nonclinical alarm rates highlight the need to refine alarm management strategies for effective clinical integration.
期刊介绍:
The Journal of Neurosurgical Anesthesiology (JNA) is a peer-reviewed publication directed to an audience of neuroanesthesiologists, neurosurgeons, neurosurgical monitoring specialists, neurosurgical support staff, and Neurosurgical Intensive Care Unit personnel. The journal publishes original peer-reviewed studies in the form of Clinical Investigations, Laboratory Investigations, Clinical Reports, Review Articles, Journal Club synopses of current literature from related journals, presentation of Points of View on controversial issues, Book Reviews, Correspondence, and Abstracts from affiliated neuroanesthesiology societies.
JNA is the Official Journal of the Society for Neuroscience in Anesthesiology and Critical Care, the Neuroanaesthesia and Critical Care Society of Great Britain and Ireland, the Association de Neuro-Anesthésiologie Réanimation de langue Française, the Wissenschaftlicher Arbeitskreis Neuroanästhesie der Deutschen Gesellschaft fur Anästhesiologie und Intensivmedizen, the Arbeitsgemeinschaft Deutschsprachiger Neuroanästhesisten und Neuro-Intensivmediziner, the Korean Society of Neuroanesthesia, the Japanese Society of Neuroanesthesia and Critical Care, the Neuroanesthesiology Chapter of the Colegio Mexicano de Anesthesiología, the Indian Society of Neuroanesthesiology and Critical Care, and the Thai Society for Neuroanesthesia.