Enhanced flexibility and dexterity in robotic endoscopy via a 6-DOF parallel mechanism and eye-gaze-assisted field-of-view control

IF 6 2区 计算机科学 Q1 AUTOMATION & CONTROL SYSTEMS
Robotics and Autonomous Systems Pub Date : 2026-04-01 Epub Date: 2025-12-27 DOI:10.1016/j.robot.2025.105322
Mengtang Li , Shen Zhao , Shuai Wang , Fanmao Liu
{"title":"Enhanced flexibility and dexterity in robotic endoscopy via a 6-DOF parallel mechanism and eye-gaze-assisted field-of-view control","authors":"Mengtang Li ,&nbsp;Shen Zhao ,&nbsp;Shuai Wang ,&nbsp;Fanmao Liu","doi":"10.1016/j.robot.2025.105322","DOIUrl":null,"url":null,"abstract":"<div><div>In conventional minimally invasive surgery, an assistant manually steers the endoscope based on the surgeon’s verbal commands, but fatigue and tremor can degrade field-of-view (FOV) stability and efficiency. Robotic endoscopes address this limitation through automated FOV adjustment via image-based visual servoing, ensuring smooth and stable visualization. However, most robotic implementations mount rigid straight-rod endoscopes on external serial arms, limiting dexterity and complicating remote-center-of-motion (RCM) control. Moreover, many automated FOV methods track surgical-tool tips without representing the surgeon’s intention. This work therefore presents a compact 6-DOF parallel endoscopic mechanism that improves flexibility and dexterity while simplifying RCM constraint satisfaction, together with an eye-gaze-assisted multi-tool tracking controller that dynamically weights tools according to surgeon attention. Simulations and experiments across diverse scenarios demonstrate FOV stabilization within 2 s, mean image-space tracking error <span><math><mo>&lt;</mo></math></span> 20 pixels, eye hand error <span><math><mo>&lt;</mo></math></span> 3°, and at least a 30% reduction in unnecessary FOV adjustments. Supplementary video is available.</div></div>","PeriodicalId":49592,"journal":{"name":"Robotics and Autonomous Systems","volume":"198 ","pages":"Article 105322"},"PeriodicalIF":6.0000,"publicationDate":"2026-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Robotics and Autonomous Systems","FirstCategoryId":"94","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0921889025004191","RegionNum":2,"RegionCategory":"计算机科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/12/27 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"AUTOMATION & CONTROL SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

In conventional minimally invasive surgery, an assistant manually steers the endoscope based on the surgeon’s verbal commands, but fatigue and tremor can degrade field-of-view (FOV) stability and efficiency. Robotic endoscopes address this limitation through automated FOV adjustment via image-based visual servoing, ensuring smooth and stable visualization. However, most robotic implementations mount rigid straight-rod endoscopes on external serial arms, limiting dexterity and complicating remote-center-of-motion (RCM) control. Moreover, many automated FOV methods track surgical-tool tips without representing the surgeon’s intention. This work therefore presents a compact 6-DOF parallel endoscopic mechanism that improves flexibility and dexterity while simplifying RCM constraint satisfaction, together with an eye-gaze-assisted multi-tool tracking controller that dynamically weights tools according to surgeon attention. Simulations and experiments across diverse scenarios demonstrate FOV stabilization within 2 s, mean image-space tracking error < 20 pixels, eye hand error < 3°, and at least a 30% reduction in unnecessary FOV adjustments. Supplementary video is available.
通过六自由度并联机构和眼球辅助视场控制,提高了机器人内窥镜的灵活性和灵活性
在传统的微创手术中,助手根据外科医生的口头指令手动操纵内窥镜,但疲劳和震颤会降低视野(FOV)的稳定性和效率。机器人内窥镜通过基于图像的视觉伺服自动调整视场来解决这一限制,确保平滑和稳定的可视化。然而,大多数机器人实现在外部串行臂上安装刚性直杆内窥镜,限制了灵活性并使远程运动中心(RCM)控制复杂化。此外,许多自动化的FOV方法跟踪手术工具提示,而不代表外科医生的意图。因此,这项工作提出了一个紧凑的6自由度平行内窥镜机构,提高了灵活性和灵活性,同时简化了RCM约束的满足,以及一个眼睛辅助的多工具跟踪控制器,根据外科医生的注意力动态加权工具。在不同场景下的模拟和实验表明,视场稳定在2秒内,平均图像空间跟踪误差为20像素,眼手误差为3°,并且至少减少了30%的不必要的视场调整。补充视频是可用的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Robotics and Autonomous Systems
Robotics and Autonomous Systems 工程技术-机器人学
CiteScore
9.00
自引率
7.00%
发文量
164
审稿时长
4.5 months
期刊介绍: Robotics and Autonomous Systems will carry articles describing fundamental developments in the field of robotics, with special emphasis on autonomous systems. An important goal of this journal is to extend the state of the art in both symbolic and sensory based robot control and learning in the context of autonomous systems. Robotics and Autonomous Systems will carry articles on the theoretical, computational and experimental aspects of autonomous systems, or modules of such systems.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书