Development of L1 Vertebral Anthropomorphic Model for Densitometric Phantom Improvement.

Sovremennye tekhnologii v meditsine Pub Date : 2025-01-01 Epub Date: 2025-08-29 DOI:10.17691/stm2025.17.4.05
A V Petraikin, A M Mikhailova, N D Kudryavtsev, M V Cherkasskaya, V O Yastrebova, O V Omelyanskaya, Y A Vasilev
{"title":"Development of L1 Vertebral Anthropomorphic Model for Densitometric Phantom Improvement.","authors":"A V Petraikin, A M Mikhailova, N D Kudryavtsev, M V Cherkasskaya, V O Yastrebova, O V Omelyanskaya, Y A Vasilev","doi":"10.17691/stm2025.17.4.05","DOIUrl":null,"url":null,"abstract":"<p><p><b>The aim of the study</b> was to develop L1 vertebral anthropomorphic model to improve the previously developed PHK FK2 phantom.</p><p><strong>Materials and methods: </strong>The vertebra was made using 3D printing by a digital model obtained from DICOM files of the abdominal CT examination. The phantom construction consists of three layers of different X-ray density. In the base made of photopolymer resin there is a cylindrical recess filled with a plastic mixture to imitate the normal state of the spongy substance (high density), osteopenia (moderate density decrease) and osteoporosis (significant decrease in density). Mineral density is regulated by changing β-tricalcium phosphate concentration. The cortical layer is modelled by applying metal foil on the base surface.</p><p><strong>Results: </strong>In X-ray tube voltage of 120 kV, the mean square deviations of the measured X-ray density values of the vertebral body, spongy substance and the cortical layer were 12.40, 3.96, and 57.23 HU, respectively. The mineral density assessment of the spongy substance for three X-ray tube voltages (100, 120, 140 kV) showed the mean absolute error to be 7.4 mg/ml, and the mean relative error - 7.3% (variation coefficient). The correction coefficient equal 7 mg/ml was used to correct the values, and after using the coefficient the mean absolute error decreased up to 0.4 mg/ml, and the mean relative error - up to 0.4% (variation coefficient). The relative measurement errors of the ventral, medial and dorsal vertebral body dimensions were 3.6, 2.7, and 2.9%, respectively.</p><p><strong>Conclusion: </strong>The methods used in developing a vertebral model can be applied in modeling the entire range of the mineral density of bone spongy substance: from osteoporosis to norm. The developed model demonstrates high stability of X-ray characteristics and anatomical accuracy; therefore, it can be used for equipment calibration, quality control of diagnostic systems, and in the training process to demonstrate the bone structure changes.</p>","PeriodicalId":520289,"journal":{"name":"Sovremennye tekhnologii v meditsine","volume":"17 4","pages":"52-57"},"PeriodicalIF":0.0000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12715482/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Sovremennye tekhnologii v meditsine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.17691/stm2025.17.4.05","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/8/29 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The aim of the study was to develop L1 vertebral anthropomorphic model to improve the previously developed PHK FK2 phantom.

Materials and methods: The vertebra was made using 3D printing by a digital model obtained from DICOM files of the abdominal CT examination. The phantom construction consists of three layers of different X-ray density. In the base made of photopolymer resin there is a cylindrical recess filled with a plastic mixture to imitate the normal state of the spongy substance (high density), osteopenia (moderate density decrease) and osteoporosis (significant decrease in density). Mineral density is regulated by changing β-tricalcium phosphate concentration. The cortical layer is modelled by applying metal foil on the base surface.

Results: In X-ray tube voltage of 120 kV, the mean square deviations of the measured X-ray density values of the vertebral body, spongy substance and the cortical layer were 12.40, 3.96, and 57.23 HU, respectively. The mineral density assessment of the spongy substance for three X-ray tube voltages (100, 120, 140 kV) showed the mean absolute error to be 7.4 mg/ml, and the mean relative error - 7.3% (variation coefficient). The correction coefficient equal 7 mg/ml was used to correct the values, and after using the coefficient the mean absolute error decreased up to 0.4 mg/ml, and the mean relative error - up to 0.4% (variation coefficient). The relative measurement errors of the ventral, medial and dorsal vertebral body dimensions were 3.6, 2.7, and 2.9%, respectively.

Conclusion: The methods used in developing a vertebral model can be applied in modeling the entire range of the mineral density of bone spongy substance: from osteoporosis to norm. The developed model demonstrates high stability of X-ray characteristics and anatomical accuracy; therefore, it can be used for equipment calibration, quality control of diagnostic systems, and in the training process to demonstrate the bone structure changes.

Abstract Image

Abstract Image

Abstract Image

L1椎体拟人模型的建立及密度幻像改善。
本研究的目的是建立L1椎体拟人化模型,以改善先前开发的PHK FK2幻影。材料和方法:利用腹部CT检查DICOM文件获取的数字模型,采用3D打印技术制作椎体。幻影结构由三层不同的x射线密度组成。在光聚合物树脂制成的底座上有一个填充塑料混合物的圆柱形凹槽,以模仿海绵状物质的正常状态(高密度),骨质减少(中等密度降低)和骨质疏松(密度显著降低)。矿物密度是通过改变β-磷酸三钙浓度来调节的。通过在基面上施加金属箔来模拟皮质层。结果:在120 kV x射线管电压下,椎体、海绵状物质和皮质层x射线密度测量值的均方差分别为12.40、3.96和57.23 HU。对3种x射线管电压(100、120、140 kV)下海绵状物质的矿物密度评估结果显示,平均绝对误差为7.4 mg/ml,平均相对误差为7.3%(变异系数)。校正系数为7 mg/ml,校正后的平均绝对误差可达0.4 mg/ml,平均相对误差可达0.4%(变异系数)。椎体腹侧、内侧和背侧尺寸的相对测量误差分别为3.6、2.7和2.9%。结论:椎体模型的建立方法可用于骨质疏松到正常骨密度的全范围骨海绵物质的建模。该模型具有较高的x线特征稳定性和解剖精度;因此,它可以用于设备校准,诊断系统的质量控制,并在训练过程中展示骨结构的变化。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书