Ricardo C Deveza, Mina Saweris, Desiree Robson, Pankaj Jain, Sumita Barua, Christian Said, Theodore Abart, Gregor Widhalm, Kavitha Muthiah, Thomas Schlöglhofer, Christopher Hayward
{"title":"Noninvasive Outflow Graft Flow Waveform Assessment Using Echocardiography and the HeartMate 3 Snoopy: A Mock Loop Study.","authors":"Ricardo C Deveza, Mina Saweris, Desiree Robson, Pankaj Jain, Sumita Barua, Christian Said, Theodore Abart, Gregor Widhalm, Kavitha Muthiah, Thomas Schlöglhofer, Christopher Hayward","doi":"10.1097/MAT.0000000000002610","DOIUrl":null,"url":null,"abstract":"<p><p>The instantaneous left ventricular assist device (LVAD) flow waveform of the HeartWare ventricular assist device (HVAD) device was previously used to assess hemodynamic parameters, which is not possible with the HeartMate 3 (HM3). The purpose of this study was to assess the ability of pulsed wave Doppler interrogation of the outflow graft (PWOG) and of a novel noninvasive pump data acquisition system (HM3 Snoopy) to obtain noninvasive flow waveforms (NIFW) and to determine whether these predict hemodynamic changes. The study was conducted using a fluid-filled, biventricular failure mock loop model with a flow probe (FP) placed around the distal outflow. Five different pump speeds, two distinct hematocrits, and preload and afterload changes were used to create 56 unique hemodynamic states. Noninvasive flow waveform parameters were assessed for correlation against FP-derived parameters. Subsequently, NIFW parameters were assessed for their predictive capabilities for preload and afterload changes. There was moderate correlation between NIFW and FP-derived systolic waveform parameters ( r range: 0.43-0.81), and strong correlation for diastolic parameters ( r range: 0.87-0.99). Flow waveform amplitude as obtained by echocardiography was the best preload predictor ( r2 = 0.67). A multiple linear regression model of NIFW parameters provided adequate prediction of afterload ( r2 = 0.85). HeartMate 3 Snoopy and PWOG are promising tools for generating flow waveform surrogates and detecting hemodynamic changes.</p>","PeriodicalId":8844,"journal":{"name":"ASAIO Journal","volume":" ","pages":"812-818"},"PeriodicalIF":2.2000,"publicationDate":"2026-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC13608275/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"ASAIO Journal","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1097/MAT.0000000000002610","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/12/22 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0
Abstract
The instantaneous left ventricular assist device (LVAD) flow waveform of the HeartWare ventricular assist device (HVAD) device was previously used to assess hemodynamic parameters, which is not possible with the HeartMate 3 (HM3). The purpose of this study was to assess the ability of pulsed wave Doppler interrogation of the outflow graft (PWOG) and of a novel noninvasive pump data acquisition system (HM3 Snoopy) to obtain noninvasive flow waveforms (NIFW) and to determine whether these predict hemodynamic changes. The study was conducted using a fluid-filled, biventricular failure mock loop model with a flow probe (FP) placed around the distal outflow. Five different pump speeds, two distinct hematocrits, and preload and afterload changes were used to create 56 unique hemodynamic states. Noninvasive flow waveform parameters were assessed for correlation against FP-derived parameters. Subsequently, NIFW parameters were assessed for their predictive capabilities for preload and afterload changes. There was moderate correlation between NIFW and FP-derived systolic waveform parameters ( r range: 0.43-0.81), and strong correlation for diastolic parameters ( r range: 0.87-0.99). Flow waveform amplitude as obtained by echocardiography was the best preload predictor ( r2 = 0.67). A multiple linear regression model of NIFW parameters provided adequate prediction of afterload ( r2 = 0.85). HeartMate 3 Snoopy and PWOG are promising tools for generating flow waveform surrogates and detecting hemodynamic changes.
期刊介绍:
ASAIO Journal is in the forefront of artificial organ research and development. On the cutting edge of innovative technology, it features peer-reviewed articles of the highest quality that describe research, development, the most recent advances in the design of artificial organ devices and findings from initial testing. Bimonthly, the ASAIO Journal features state-of-the-art investigations, laboratory and clinical trials, and discussions and opinions from experts around the world.
The official publication of the American Society for Artificial Internal Organs.