87 - Effects of aging on the mechanical properties of GraphyⓇ orthodontic aligners: An in vitro study

IF 3.5 Q1 DENTISTRY, ORAL SURGERY & MEDICINE
Vande Vannet Bart , Ursat Pauline , Ursat Eric , Hocquel Thibaut , Canceill Thibault
{"title":"87 - Effects of aging on the mechanical properties of GraphyⓇ orthodontic aligners: An in vitro study","authors":"Vande Vannet Bart ,&nbsp;Ursat Pauline ,&nbsp;Ursat Eric ,&nbsp;Hocquel Thibaut ,&nbsp;Canceill Thibault","doi":"10.1016/j.ejwf.2025.07.091","DOIUrl":null,"url":null,"abstract":"<div><div><strong>Background:</strong> Direct 3D printing of aligners using Tera Harz TC-85 material (Graphy<sup>Ⓡ</sup>, Seoul, Republic of Korea) eliminates the need for printed dental models, offering a potentially more precise, efficient, and cost-effective alternative. Current guidelines recommend wearing aligners for 22 hours daily and replacing them every 15 days.</div><div><strong>Objectives:</strong> This in vitro study evaluates the mechanical properties of Tera Harz TC-85 aligners at different time intervals (D0, D7, D14) in artificial saliva to assess force delivery and deformation over time.</div><div><strong>Methods:</strong> Eighteen identical Class I aligners (GRAPHY<sup>Ⓡ</sup>, Forestadent Bernhard Foerster GmbH<sup>Ⓡ</sup>, Pforzheim, Germany) were divided into three groups (D0, D7, D14) and subjected to continuous aging in Fusayama-Mayer artificial saliva (Serlabo Technologies, Vedene, France) at 37°C. Following aging, aligners underwent traction testing using a Universal Testing Machine (LRX Plus Dynamometer Lloyd Instruments) to measure maximum tensile and yield strength.</div><div><strong>Results:</strong> The average rupture force was 169.98 N (SD 32.32, CV 19.02%) at the molars and 222.67 N (SD 35.56, CV 15.97%) at the incisors. Deformation at D0 was significantly different from D7 and D14 in both incisors and molars (p &lt; 0.001).</div><div><strong>Conclusions:</strong> The mechanical stability of Tera Harz TC-85 suggests that aligners should be replaced weekly rather than every 15 days. Direct 3D printing does not reduce the total number of aligners needed, as shape memory becomes unstable over time.</div></div>","PeriodicalId":43456,"journal":{"name":"Journal of the World Federation of Orthodontists","volume":"14 6","pages":"Page 428"},"PeriodicalIF":3.5000,"publicationDate":"2025-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of the World Federation of Orthodontists","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2212443825001468","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/12/9 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Direct 3D printing of aligners using Tera Harz TC-85 material (GraphyⓇ, Seoul, Republic of Korea) eliminates the need for printed dental models, offering a potentially more precise, efficient, and cost-effective alternative. Current guidelines recommend wearing aligners for 22 hours daily and replacing them every 15 days.
Objectives: This in vitro study evaluates the mechanical properties of Tera Harz TC-85 aligners at different time intervals (D0, D7, D14) in artificial saliva to assess force delivery and deformation over time.
Methods: Eighteen identical Class I aligners (GRAPHYⓇ, Forestadent Bernhard Foerster GmbHⓇ, Pforzheim, Germany) were divided into three groups (D0, D7, D14) and subjected to continuous aging in Fusayama-Mayer artificial saliva (Serlabo Technologies, Vedene, France) at 37°C. Following aging, aligners underwent traction testing using a Universal Testing Machine (LRX Plus Dynamometer Lloyd Instruments) to measure maximum tensile and yield strength.
Results: The average rupture force was 169.98 N (SD 32.32, CV 19.02%) at the molars and 222.67 N (SD 35.56, CV 15.97%) at the incisors. Deformation at D0 was significantly different from D7 and D14 in both incisors and molars (p < 0.001).
Conclusions: The mechanical stability of Tera Harz TC-85 suggests that aligners should be replaced weekly rather than every 15 days. Direct 3D printing does not reduce the total number of aligners needed, as shape memory becomes unstable over time.
老化对GraphyⓇ正畸矫正器力学性能的影响:一项体外研究
背景:使用Tera Harz TC-85材料直接3D打印牙齿矫正器(GraphyⓇ,Seoul, Republic of Korea)消除了对打印牙齿模型的需求,提供了一种潜在的更精确、更高效、更经济的替代方案。目前的指南建议每天佩戴牙齿矫正器22小时,每15天更换一次。目的:本体外研究评估Tera Harz TC-85对准器在人工唾液中不同时间间隔(D0, D7, D14)的力学性能,以评估力传递和变形随时间的变化。方法:18个相同的I类对准器(GRAPHYⓇ,Forestadent Bernhard Foerster GmbHⓇ,Pforzheim, Germany)分为3组(D0, D7, D14),在37°C的Fusayama-Mayer人工唾液(Serlabo Technologies, Vedene, France)中持续老化。老化后,矫直器使用通用试验机(LRX Plus Dynamometer Lloyd Instruments)进行牵引测试,以测量最大拉伸强度和屈服强度。结果:磨牙的平均破裂力为169.98 N (SD 32.32, CV 19.02%),门牙的平均破裂力为222.67 N (SD 35.56, CV 15.97%)。门牙和磨牙在D0的变形与D7和D14有显著差异(p < 0.001)。结论:Tera Harz TC-85的机械稳定性建议每周更换矫正器,而不是每15天更换一次。直接3D打印不会减少所需对准器的总数,因为形状记忆会随着时间的推移而变得不稳定。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of the World Federation of Orthodontists
Journal of the World Federation of Orthodontists DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
3.80
自引率
4.80%
发文量
34
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书