Applicability of wide-pore columns in reversed-phase liquid chromatography for human blood lipid profiling

IF 3.6
Journal of chromatography open Pub Date : 2026-05-01 Epub Date: 2025-11-28 DOI:10.1016/j.jcoa.2025.100295
Sergey Girel , Davy Guillarme , Serge Rudaz , Isabel Meister
{"title":"Applicability of wide-pore columns in reversed-phase liquid chromatography for human blood lipid profiling","authors":"Sergey Girel ,&nbsp;Davy Guillarme ,&nbsp;Serge Rudaz ,&nbsp;Isabel Meister","doi":"10.1016/j.jcoa.2025.100295","DOIUrl":null,"url":null,"abstract":"<div><div>For the resolution of isomer-rich lipid extracts, it is common practice during RPLC method development to evaluate column efficiency and mobile phase composition. However, further gains can be made by tuning other parameters, such as the microstructure of the packing material. Today, lipid separations are mostly performed with columns packed with particles having narrow pore sizes (80–120 Å). This is unsurprising, given the expected hydrodynamic radius of lipids up to of 23 Å. However, specific investigations of pore size influence on lipid separations are scarce. Therefore, a systematic comparison of 150 mm columns packed with C<sub>18</sub> particles having pore sizes of 130 and 300 Å was performed with respect to selectivity, peak capacity and chromatographic resolution. An RPLC/HRMS-based lipidomic analysis was developed on 150 mm columns. Similar loadability, selectivity, sensitivity and average peak capacity were demonstrated by 300 Å column compared to its narrow-pore counterpart. Overall, current experimental results indicate that wide-pore columns are equally applicable for RPLC/HRMS-based discovery lipidomics and may be even beneficial for the analysis of lipids with high structural complexity. In addition, these columns appear to be better suitable for high-throughput analysis due to lower lipid retention via smaller surface area.</div></div>","PeriodicalId":93576,"journal":{"name":"Journal of chromatography open","volume":"9 ","pages":"Article 100295"},"PeriodicalIF":3.6000,"publicationDate":"2026-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of chromatography open","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2772391725000945","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/11/28 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

For the resolution of isomer-rich lipid extracts, it is common practice during RPLC method development to evaluate column efficiency and mobile phase composition. However, further gains can be made by tuning other parameters, such as the microstructure of the packing material. Today, lipid separations are mostly performed with columns packed with particles having narrow pore sizes (80–120 Å). This is unsurprising, given the expected hydrodynamic radius of lipids up to of 23 Å. However, specific investigations of pore size influence on lipid separations are scarce. Therefore, a systematic comparison of 150 mm columns packed with C18 particles having pore sizes of 130 and 300 Å was performed with respect to selectivity, peak capacity and chromatographic resolution. An RPLC/HRMS-based lipidomic analysis was developed on 150 mm columns. Similar loadability, selectivity, sensitivity and average peak capacity were demonstrated by 300 Å column compared to its narrow-pore counterpart. Overall, current experimental results indicate that wide-pore columns are equally applicable for RPLC/HRMS-based discovery lipidomics and may be even beneficial for the analysis of lipids with high structural complexity. In addition, these columns appear to be better suitable for high-throughput analysis due to lower lipid retention via smaller surface area.

Abstract Image

宽孔柱在反相液相色谱法中人血脂分析中的适用性
对于富含异构体的脂质提取物,在RPLC方法开发过程中评估柱效率和流动相组成是常见的做法。然而,进一步的增益可以通过调整其他参数,如包装材料的微观结构来实现。今天,脂质分离主要是用柱填充的颗粒具有狭窄的孔径(80-120 Å)。这并不奇怪,因为脂质的流体动力学半径可达23 Å。然而,孔径对脂质分离影响的具体研究很少。因此,对孔径为130和300 Å的C18颗粒填充的150 mm柱进行了选择性、峰容量和色谱分辨率的系统比较。在150 mm柱上建立了基于RPLC/ hrms的脂质组学分析。与窄孔柱相比,300 Å柱具有相似的负载性、选择性、灵敏度和平均峰值容量。总的来说,目前的实验结果表明,宽孔柱同样适用于基于RPLC/ hrms的发现脂质组学,甚至可能有利于分析高结构复杂性的脂质。此外,这些色谱柱似乎更适合于高通量分析,因为通过较小的表面积降低脂质滞留。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of chromatography open
Journal of chromatography open Analytical Chemistry
CiteScore
2.50
自引率
0.00%
发文量
0
审稿时长
50 days
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书