Rio Nakamine, Ryoga Matsumoto, Misaki Matsumoto, Yoshiyuki Yoshikawa, Noriaki Maeshige
{"title":"Palpation versus hand‑held Doppler for identifying lower‑limb peripheral arteries: Time‑efficiency and reliability in healthy adults","authors":"Rio Nakamine, Ryoga Matsumoto, Misaki Matsumoto, Yoshiyuki Yoshikawa, Noriaki Maeshige","doi":"10.1111/cpf.70039","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Bedside lower‑limb vascular assessment commonly relies on pedal pulse palpation or hand‑held Doppler, yet their relative time‑efficiency and the reliability of Doppler measurements in novices are unclear.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Twenty healthy adults were examined by three undergraduate raters without prior Doppler experience. For dorsalis pedis, anterior tibial, and posterior tibial arteries, each rater performed three trials with palpation and with a hand‑held continuous‑wave Doppler. Primary outcomes were time‑to‑identification and identification success; for Doppler, peak systolic velocity (PSV) reliability was evaluated (intra‑ and inter‑rater intraclass correlation coefficients [ICC] and Bland–Altman).</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Differences were defined as Doppler—palpation (positive values indicate longer time for Doppler). Palpation was faster than Doppler for dorsalis pedis (Hodges–Lehmann +6.47 s, 95% confidence interval [CI]: 2.40–8.00; <i>p</i> = 0.004) and posterior tibial arteries (+6.38 s, 1.10–13.61; <i>p</i> = 0.012). The anterior tibial artery was never identified by palpation but was identified by Doppler in all participants (exact McNemar <i>p</i> = 1.91 × 10⁻⁶). Doppler reliability for dorsalis pedis PSV was excellent within raters (ICC [3.1]: 0.926–0.971) and between raters (ICC [2.1]: 0.966), with small mean biases and narrow limits of agreement.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>In healthy adults, palpation enables faster identification of dorsalis pedis and posterior tibial pulses, whereas the deeper anterior tibial artery is consistently detected only with Doppler. Doppler‑derived PSV shows excellent reproducibility even among novice examiners. These findings provide methods-level evidence to inform protocol design and training; validation in patient populations is required before any workflow recommendations.</p>\n </section>\n </div>","PeriodicalId":10504,"journal":{"name":"Clinical Physiology and Functional Imaging","volume":"46 1","pages":""},"PeriodicalIF":1.7000,"publicationDate":"2025-12-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical Physiology and Functional Imaging","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/cpf.70039","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"PHYSIOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Bedside lower‑limb vascular assessment commonly relies on pedal pulse palpation or hand‑held Doppler, yet their relative time‑efficiency and the reliability of Doppler measurements in novices are unclear.
Methods
Twenty healthy adults were examined by three undergraduate raters without prior Doppler experience. For dorsalis pedis, anterior tibial, and posterior tibial arteries, each rater performed three trials with palpation and with a hand‑held continuous‑wave Doppler. Primary outcomes were time‑to‑identification and identification success; for Doppler, peak systolic velocity (PSV) reliability was evaluated (intra‑ and inter‑rater intraclass correlation coefficients [ICC] and Bland–Altman).
Results
Differences were defined as Doppler—palpation (positive values indicate longer time for Doppler). Palpation was faster than Doppler for dorsalis pedis (Hodges–Lehmann +6.47 s, 95% confidence interval [CI]: 2.40–8.00; p = 0.004) and posterior tibial arteries (+6.38 s, 1.10–13.61; p = 0.012). The anterior tibial artery was never identified by palpation but was identified by Doppler in all participants (exact McNemar p = 1.91 × 10⁻⁶). Doppler reliability for dorsalis pedis PSV was excellent within raters (ICC [3.1]: 0.926–0.971) and between raters (ICC [2.1]: 0.966), with small mean biases and narrow limits of agreement.
Conclusions
In healthy adults, palpation enables faster identification of dorsalis pedis and posterior tibial pulses, whereas the deeper anterior tibial artery is consistently detected only with Doppler. Doppler‑derived PSV shows excellent reproducibility even among novice examiners. These findings provide methods-level evidence to inform protocol design and training; validation in patient populations is required before any workflow recommendations.
期刊介绍:
Clinical Physiology and Functional Imaging publishes reports on clinical and experimental research pertinent to human physiology in health and disease. The scope of the Journal is very broad, covering all aspects of the regulatory system in the cardiovascular, renal and pulmonary systems with special emphasis on methodological aspects. The focus for the journal is, however, work that has potential clinical relevance. The Journal also features review articles on recent front-line research within these fields of interest.