Endovascular repair of aortic dissection with a bioresorbable patch: Computational and experimental study

IF 3.5 Q2 ENGINEERING, MULTIDISCIPLINARY
Applications in engineering science Pub Date : 2026-03-01 Epub Date: 2025-12-03 DOI:10.1016/j.apples.2025.100277
Marta I. Bracco , Francesc Canalejo-Codina , Agustina Giuliodori , Andrea Montanino , Alejandro Aranda , Jordi Martorell , Eduardo Soudah
{"title":"Endovascular repair of aortic dissection with a bioresorbable patch: Computational and experimental study","authors":"Marta I. Bracco ,&nbsp;Francesc Canalejo-Codina ,&nbsp;Agustina Giuliodori ,&nbsp;Andrea Montanino ,&nbsp;Alejandro Aranda ,&nbsp;Jordi Martorell ,&nbsp;Eduardo Soudah","doi":"10.1016/j.apples.2025.100277","DOIUrl":null,"url":null,"abstract":"<div><div>This study introduces an experimentally-calibrated finite-element framework to predict the endovascular sealing performance of a bioresorbable patch for aortic dissection repair. The patch–aortic wall interaction was modeled using an adhesion-enabled contact formulation, with parameters derived from a custom dye-penetration test to replicate in-vivo tissue adhesion. A parametric analysis assessed the impact of tear size (10–20 mm), tear morphology (round vs. circumferential ellipse), and deployment angle (5<span><math><mo>°</mo></math></span>–20<span><math><mo>°</mo></math></span>) on patch sealing efficiency, wall compliance, and local stress distribution under physiological loading. Tear geometry was identified as the dominant determinant of sealing: large round tears reduced effective apposition, while circumferential elliptical tears promoted full wall coupling at lower deployment forces. Increasing deployment angle raised insertion forces and impaired circumferential contact. Importantly, pulsatile hemodynamic loading demonstrated that the patch preserved native wall compliance without inducing adverse stress concentrations. By integrating experimental calibration with computational modeling, this framework offers a quantitative tool to evaluate anatomical and procedural factors influencing endovascular sealing. These insights may support the design optimization and clinical translation of resorbable patch-based strategies for aortic dissection repair.</div></div>","PeriodicalId":72251,"journal":{"name":"Applications in engineering science","volume":"25 ","pages":"Article 100277"},"PeriodicalIF":3.5000,"publicationDate":"2026-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Applications in engineering science","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2666496825000755","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/12/3 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ENGINEERING, MULTIDISCIPLINARY","Score":null,"Total":0}
引用次数: 0

Abstract

This study introduces an experimentally-calibrated finite-element framework to predict the endovascular sealing performance of a bioresorbable patch for aortic dissection repair. The patch–aortic wall interaction was modeled using an adhesion-enabled contact formulation, with parameters derived from a custom dye-penetration test to replicate in-vivo tissue adhesion. A parametric analysis assessed the impact of tear size (10–20 mm), tear morphology (round vs. circumferential ellipse), and deployment angle (5°–20°) on patch sealing efficiency, wall compliance, and local stress distribution under physiological loading. Tear geometry was identified as the dominant determinant of sealing: large round tears reduced effective apposition, while circumferential elliptical tears promoted full wall coupling at lower deployment forces. Increasing deployment angle raised insertion forces and impaired circumferential contact. Importantly, pulsatile hemodynamic loading demonstrated that the patch preserved native wall compliance without inducing adverse stress concentrations. By integrating experimental calibration with computational modeling, this framework offers a quantitative tool to evaluate anatomical and procedural factors influencing endovascular sealing. These insights may support the design optimization and clinical translation of resorbable patch-based strategies for aortic dissection repair.

Abstract Image

生物可吸收贴片血管内修复主动脉夹层:计算与实验研究
本研究引入了一个实验校准的有限元框架来预测用于主动脉夹层修复的生物可吸收贴片的血管内密封性能。膜片-主动脉壁相互作用的模型采用了一种能够粘附的接触配方,其参数来源于一种定制的染料渗透测试,以复制体内组织粘附。参数分析评估了撕裂尺寸(10-20 mm)、撕裂形态(圆形或圆周椭圆形)和展开角度(5°-20°)对补片密封效率、壁面柔顺性和生理载荷下局部应力分布的影响。撕裂几何形状被认为是密封的主要决定因素:大的圆形撕裂降低了有效的对抗,而圆周椭圆形撕裂在较低的展开力下促进了全壁耦合。增大展开角会增加插入力,破坏周向接触。重要的是,脉动血流动力学加载表明,贴片保留了原生壁面的顺应性,而不会引起不利的应力集中。通过将实验校准与计算建模相结合,该框架为评估影响血管内密封的解剖学和程序性因素提供了定量工具。这些见解可能支持基于可吸收贴片的主动脉夹层修复策略的设计优化和临床转化。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Applications in engineering science
Applications in engineering science Mechanical Engineering
CiteScore
3.60
自引率
0.00%
发文量
0
审稿时长
68 days
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书