Reliability of capturing plantar foot geometry using a handheld 3D scanner for dose-response modeling of custom-made foot orthoses.

IF 1.4 4区 医学 Q4 ORTHOPEDICS
Stijn Rosseel, Giovanni Matricali, Eleonora Ferraris, Hans Hallez, Annelies Bockstael, Kevin Deschamps
{"title":"Reliability of capturing plantar foot geometry using a handheld 3D scanner for dose-response modeling of custom-made foot orthoses.","authors":"Stijn Rosseel, Giovanni Matricali, Eleonora Ferraris, Hans Hallez, Annelies Bockstael, Kevin Deschamps","doi":"10.1097/PXR.0000000000000485","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Efficacy of custom-made foot orthoses largely depends on geometric characterization and outcomes can be suboptimal due to insufficient quantitative insights thereof. Addressing this is crucial for understanding geometric characteristics in dosage-response modeling.</p><p><strong>Objective: </strong>This study evaluates the reliability of digitizing traditional and advanced geometric foot features, using a handheld 3-dimensional (3D) scanning system.</p><p><strong>Methods: </strong>Two examiners used a handheld 3D scanner to capture the foot shape of 30 healthy participants in a non-weight-bearing condition. Intra-tester, test-retest, and inter-tester reliabilities of several linear anthropometric and geometric variables were evaluated. Subsequently, the consistency of capturing the geometry of the plantar foot surface was assessed by quantifying 2 anteroposterior and 2 mediolateral curvilinear plantar surface metrics.</p><p><strong>Results: </strong>Intraclass correlation coefficients for the linear geometric and anthropometric variables' intra- and inter-tester reliability ranged from 0.84 to 1.00. The standard error of measurement for the Z-coordinate of the proximal medial arch point and the forefoot-rearfoot angle was substantially higher compared with the standard error of measurements observed in the other metrics. A pairwise comparison of the 4 curvilinear metrics showed no statistically significant difference on intra-tester and test-retest levels for both examiners. Inter-tester analysis identified significant ( p < 0.05) differences in specific segments of the anteroposterior and mediolateral curvilinear metrics, particularly near the distal and medial ends.</p><p><strong>Conclusions: </strong>These results demonstrated that the digitization of linear anthropometric and geometric measurements of the human foot, obtained using a handheld 3D scanning system in a non-weight-bearing condition, exhibit good to excellent reliability. Additionally, the curvilinear metrics, related to the plantar surface curvature, showed high consistency, supporting its usage to quantify plantar geometry and provide insights into the role of geometric characteristics in dosage-response modeling in custom-made foot orthoses practice.</p>","PeriodicalId":49657,"journal":{"name":"Prosthetics and Orthotics International","volume":" ","pages":"483-490"},"PeriodicalIF":1.4000,"publicationDate":"2026-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Prosthetics and Orthotics International","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/PXR.0000000000000485","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/12/2 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Efficacy of custom-made foot orthoses largely depends on geometric characterization and outcomes can be suboptimal due to insufficient quantitative insights thereof. Addressing this is crucial for understanding geometric characteristics in dosage-response modeling.

Objective: This study evaluates the reliability of digitizing traditional and advanced geometric foot features, using a handheld 3-dimensional (3D) scanning system.

Methods: Two examiners used a handheld 3D scanner to capture the foot shape of 30 healthy participants in a non-weight-bearing condition. Intra-tester, test-retest, and inter-tester reliabilities of several linear anthropometric and geometric variables were evaluated. Subsequently, the consistency of capturing the geometry of the plantar foot surface was assessed by quantifying 2 anteroposterior and 2 mediolateral curvilinear plantar surface metrics.

Results: Intraclass correlation coefficients for the linear geometric and anthropometric variables' intra- and inter-tester reliability ranged from 0.84 to 1.00. The standard error of measurement for the Z-coordinate of the proximal medial arch point and the forefoot-rearfoot angle was substantially higher compared with the standard error of measurements observed in the other metrics. A pairwise comparison of the 4 curvilinear metrics showed no statistically significant difference on intra-tester and test-retest levels for both examiners. Inter-tester analysis identified significant ( p < 0.05) differences in specific segments of the anteroposterior and mediolateral curvilinear metrics, particularly near the distal and medial ends.

Conclusions: These results demonstrated that the digitization of linear anthropometric and geometric measurements of the human foot, obtained using a handheld 3D scanning system in a non-weight-bearing condition, exhibit good to excellent reliability. Additionally, the curvilinear metrics, related to the plantar surface curvature, showed high consistency, supporting its usage to quantify plantar geometry and provide insights into the role of geometric characteristics in dosage-response modeling in custom-made foot orthoses practice.

使用手持式3D扫描仪捕获足底足几何形状的可靠性,用于定制足矫形器的剂量响应建模。
背景:定制足部矫形器的效果在很大程度上取决于几何特征,由于对其定量认识不足,结果可能不理想。解决这个问题对于理解剂量-反应建模中的几何特征至关重要。目的:本研究利用手持式三维(3D)扫描系统评估数字化传统和先进几何足部特征的可靠性。方法:两名检查人员使用手持式3D扫描仪捕捉30名健康参与者在非负重条件下的足形。评估了几个线性人体测量和几何变量的测试者内部、测试-再测试和测试者之间的信度。随后,通过量化2个前后和2个中外侧曲线足底表面指标来评估捕获足底足表面几何形状的一致性。结果:线性几何变量和人体测量变量的组内相关系数在测试者内部和测试者之间的信度范围为0.84 ~ 1.00。近中弓点z坐标和前后足角的测量标准误差明显高于其他指标的测量标准误差。4个曲线指标的两两比较显示,两名考官在测试者内部和测试复试水平上无统计学差异。测试者间分析发现,在特定的前后和中外侧曲线指标中,特别是在远端和中端附近,差异显著(p < 0.05)。结论:这些结果表明,在非负重条件下,使用手持式3D扫描系统获得的人体足部线性人体测量和几何测量数据的数字化具有良好到优异的可靠性。此外,与足底表面曲率相关的曲线指标显示出高度一致性,支持其用于量化足底几何形状,并为定制足部矫形器实践中几何特征在剂量-反应建模中的作用提供见解。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
2.30
自引率
13.30%
发文量
208
审稿时长
6-12 weeks
期刊介绍: Prosthetics and Orthotics International is an international, multidisciplinary journal for all professionals who have an interest in the medical, clinical, rehabilitation, technical, educational and research aspects of prosthetics, orthotics and rehabilitation engineering, as well as their related topics.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书