Long Term Real World Performance of the Gore Excluder AAA Endoprosthesis Featuring the C3 Delivery System: Ten Year Results from the European C3 Module of the GREAT Registry
Moritz S. Bischoff, Eric Verhoeven, Athanasios Katsargyris, Dittmar Böckler
{"title":"Long Term Real World Performance of the Gore Excluder AAA Endoprosthesis Featuring the C3 Delivery System: Ten Year Results from the European C3 Module of the GREAT Registry","authors":"Moritz S. Bischoff, Eric Verhoeven, Athanasios Katsargyris, Dittmar Böckler","doi":"10.1016/j.ejvs.2025.11.044","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><div>This study aimed to report the ten year data for patients treated with the Gore Excluder AAA Endoprosthesis featuring the C3 Delivery System for infrarenal aortic disease, as recorded in the European C3 module of the Gore Global Registry for Endovascular Aortic Treatment.</div></div><div>This study aimed to report the ten year data for patients treated with the Gore Excluder AAA Endoprosthesis featuring the C3 Delivery System for infrarenal aortic disease, as recorded in the European C3 module of the Gore Global Registry for Endovascular Aortic Treatment.</div><div><h3>Methods</h3><div>The European C3 module included patients from 13 centres who underwent endovascular aortic aneurysm repair between August 2010 and December 2012. Outcomes were assessed using descriptive statistics and Kaplan–Meier survival analysis with 95% confidence intervals (CIs).</div></div><div>The European C3 module included patients from 13 centres who underwent endovascular aortic aneurysm repair between August 2010 and December 2012. Outcomes were assessed using descriptive statistics and Kaplan–Meier survival analysis with 95% confidence intervals (CIs).</div><div><h3>Results</h3><div>At module closure, 396 patients (87.1% men; mean age ± standard deviation 73.9 ± 7.8 years) were available for analysis, with a mean follow up duration ± standard deviation of 6.4 ± 3.2 years. Of these, 94.4% underwent elective endovascular aortic aneurysm repair for infrarenal aortic aneurysm, and 16.9% (67 of 396) were treated outside the instructions for use. At ten years, the all cause mortality rate was 46.0% with 182 deaths: five aortic related, 120 non-aortic related, and 57 of unknown cause. Freedom from aortic related death was 98.6% (95% CI 96.7 – 99.4%). Freedom from any re-intervention and device related re-intervention was 77.3% (95% CI 71.8 – 81.8%) and 80.6% (95% CI 75.3 – 84.9%), respectively. Freedom from type 1 endoleak was 95.4% (95% CI 92.3 – 97.5%), with type 2 endoleaks accounting for 82% (41 of 50) of all endoleaks. Structural complications were rare: two cases of migration (0.5%), one case of compression (0.3%), and no cases of device fractures. Nine patients (2.3%) required conversion, and secondary rupture occurred in three cases (0.8%).</div></div><div>At module closure, 396 patients (87.1% men; mean age ± standard deviation 73.9 ± 7.8 years) were available for analysis, with a mean follow up duration ± standard deviation of 6.4 ± 3.2 years. Of these, 94.4% underwent elective endovascular aortic aneurysm repair for infrarenal aortic aneurysm, and 16.9% (67 of 396) were treated outside the instructions for use. At ten years, the all cause mortality rate was 46.0% with 182 deaths: five aortic related, 120 non-aortic related, and 57 of unknown cause. Freedom from aortic related death was 98.6% (95% CI 96.7 – 99.4%). Freedom from any re-intervention and device related re-intervention was 77.3% (95% CI 71.8 – 81.8%) and 80.6% (95% CI 75.3 – 84.9%), respectively. Freedom from type 1 endoleak was 95.4% (95% CI 92.3 – 97.5%), with type 2 endoleaks accounting for 82% (41 of 50) of all endoleaks. Structural complications were rare: two cases of migration (0.5%), one case of compression (0.3%), and no cases of device fractures. Nine patients (2.3%) required conversion, and secondary rupture occurred in three cases (0.8%).</div><div><h3>Conclusion</h3><div>The Gore C3 Excluder shows long term safety and durability for infrarenal aortic disease, with low rates of aortic related death and major aortic events. These findings support its effectiveness in a real world setting. However, surveillance adherence remains challenging in daily practice.</div></div><div>The Gore C3 Excluder shows long term safety and durability for infrarenal aortic disease, with low rates of aortic related death and major aortic events. These findings support its effectiveness in a real world setting. However, surveillance adherence remains challenging in daily practice.</div>","PeriodicalId":55160,"journal":{"name":"European Journal of Vascular and Endovascular Surgery","volume":"72 4","pages":"Pages 613-621"},"PeriodicalIF":8.5000,"publicationDate":"2026-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Journal of Vascular and Endovascular Surgery","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1078588425011876","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/11/28 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0
Abstract
Objective
This study aimed to report the ten year data for patients treated with the Gore Excluder AAA Endoprosthesis featuring the C3 Delivery System for infrarenal aortic disease, as recorded in the European C3 module of the Gore Global Registry for Endovascular Aortic Treatment.
This study aimed to report the ten year data for patients treated with the Gore Excluder AAA Endoprosthesis featuring the C3 Delivery System for infrarenal aortic disease, as recorded in the European C3 module of the Gore Global Registry for Endovascular Aortic Treatment.
Methods
The European C3 module included patients from 13 centres who underwent endovascular aortic aneurysm repair between August 2010 and December 2012. Outcomes were assessed using descriptive statistics and Kaplan–Meier survival analysis with 95% confidence intervals (CIs).
The European C3 module included patients from 13 centres who underwent endovascular aortic aneurysm repair between August 2010 and December 2012. Outcomes were assessed using descriptive statistics and Kaplan–Meier survival analysis with 95% confidence intervals (CIs).
Results
At module closure, 396 patients (87.1% men; mean age ± standard deviation 73.9 ± 7.8 years) were available for analysis, with a mean follow up duration ± standard deviation of 6.4 ± 3.2 years. Of these, 94.4% underwent elective endovascular aortic aneurysm repair for infrarenal aortic aneurysm, and 16.9% (67 of 396) were treated outside the instructions for use. At ten years, the all cause mortality rate was 46.0% with 182 deaths: five aortic related, 120 non-aortic related, and 57 of unknown cause. Freedom from aortic related death was 98.6% (95% CI 96.7 – 99.4%). Freedom from any re-intervention and device related re-intervention was 77.3% (95% CI 71.8 – 81.8%) and 80.6% (95% CI 75.3 – 84.9%), respectively. Freedom from type 1 endoleak was 95.4% (95% CI 92.3 – 97.5%), with type 2 endoleaks accounting for 82% (41 of 50) of all endoleaks. Structural complications were rare: two cases of migration (0.5%), one case of compression (0.3%), and no cases of device fractures. Nine patients (2.3%) required conversion, and secondary rupture occurred in three cases (0.8%).
At module closure, 396 patients (87.1% men; mean age ± standard deviation 73.9 ± 7.8 years) were available for analysis, with a mean follow up duration ± standard deviation of 6.4 ± 3.2 years. Of these, 94.4% underwent elective endovascular aortic aneurysm repair for infrarenal aortic aneurysm, and 16.9% (67 of 396) were treated outside the instructions for use. At ten years, the all cause mortality rate was 46.0% with 182 deaths: five aortic related, 120 non-aortic related, and 57 of unknown cause. Freedom from aortic related death was 98.6% (95% CI 96.7 – 99.4%). Freedom from any re-intervention and device related re-intervention was 77.3% (95% CI 71.8 – 81.8%) and 80.6% (95% CI 75.3 – 84.9%), respectively. Freedom from type 1 endoleak was 95.4% (95% CI 92.3 – 97.5%), with type 2 endoleaks accounting for 82% (41 of 50) of all endoleaks. Structural complications were rare: two cases of migration (0.5%), one case of compression (0.3%), and no cases of device fractures. Nine patients (2.3%) required conversion, and secondary rupture occurred in three cases (0.8%).
Conclusion
The Gore C3 Excluder shows long term safety and durability for infrarenal aortic disease, with low rates of aortic related death and major aortic events. These findings support its effectiveness in a real world setting. However, surveillance adherence remains challenging in daily practice.
The Gore C3 Excluder shows long term safety and durability for infrarenal aortic disease, with low rates of aortic related death and major aortic events. These findings support its effectiveness in a real world setting. However, surveillance adherence remains challenging in daily practice.
期刊介绍:
The European Journal of Vascular and Endovascular Surgery is aimed primarily at vascular surgeons dealing with patients with arterial, venous and lymphatic diseases. Contributions are included on the diagnosis, investigation and management of these vascular disorders. Papers that consider the technical aspects of vascular surgery are encouraged, and the journal includes invited state-of-the-art articles.
Reflecting the increasing importance of endovascular techniques in the management of vascular diseases and the value of closer collaboration between the vascular surgeon and the vascular radiologist, the journal has now extended its scope to encompass the growing number of contributions from this exciting field. Articles describing endovascular method and their critical evaluation are included, as well as reports on the emerging technology associated with this field.