{"title":"Feasibility and efficacy of a thermal ablation device utilizing a nitinol-based coil.","authors":"Fuyuki Nagano, Akitoshi Inoue, Shohei Chatani, Yugo Imai, Akinaga Sonoda, Norihisa Nitta, Ken-Ichi Mukaisho, Naotake Nakamura, Yoshiyuki Watanabe","doi":"10.1177/02841851251396571","DOIUrl":null,"url":null,"abstract":"<p><p>BackgroundA fibrous actuator, composed of a nitinol-based coil memory alloy, functions as a flexible coil under normal conditions. It contracts and generates heat when activated with an electric current. This property makes it a promising candidate for use as an interventional thermal device.PurposeTo evaluate the thermal effects of a nitinol coil ablation device on liver tissue and the intrahepatic portal vein.Material and MethodsAn in vitro experiment using bovine resected liver was performed. Ten Japanese white rabbits were divided equally into two groups (0.6-W and 1.1-W) based on electric power settings. Ablation was performed in three liver regions for each rabbit. Ablated areas were measured using fat-suppressed T1-weighted imaging following gadolinium administration and histological examination with hematoxylin and eosin staining. For the intrahepatic portal vein ablation, nine rabbits were assigned to the ablation group and eight rabbits served as controls. Portography was conducted before and after the procedure.ResultsIn vitro, temperature increases at ablation site and at 5-, 10-, and 15-mm distances and ellipsoid-shaped ablated area were proportional to electric power. In the in vivo experiment, the 1.1-W group exhibited significantly greater ablation effects compared to the 0.6-W group, as measured by magnetic resonance imaging (50.6 vs. 22.7 mm²; <i>P</i> <0.001) and histopathology (38.9 vs. 21.7 mm²; <i>P</i> = 0.001). Nearly complete occlusion (>80% stenosis) was more frequently observed in the ablation group (72.2% vs. 22.2%; <i>P</i> = 0.007).ConclusionThe thermal ablation device utilizing a nitinol coil demonstrated effective liver ablation. Furthermore, it shows potential for customization in endovascular ablation applications.</p>","PeriodicalId":7143,"journal":{"name":"Acta radiologica","volume":" ","pages":"178-187"},"PeriodicalIF":1.2000,"publicationDate":"2026-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta radiologica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/02841851251396571","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/12/1 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
BackgroundA fibrous actuator, composed of a nitinol-based coil memory alloy, functions as a flexible coil under normal conditions. It contracts and generates heat when activated with an electric current. This property makes it a promising candidate for use as an interventional thermal device.PurposeTo evaluate the thermal effects of a nitinol coil ablation device on liver tissue and the intrahepatic portal vein.Material and MethodsAn in vitro experiment using bovine resected liver was performed. Ten Japanese white rabbits were divided equally into two groups (0.6-W and 1.1-W) based on electric power settings. Ablation was performed in three liver regions for each rabbit. Ablated areas were measured using fat-suppressed T1-weighted imaging following gadolinium administration and histological examination with hematoxylin and eosin staining. For the intrahepatic portal vein ablation, nine rabbits were assigned to the ablation group and eight rabbits served as controls. Portography was conducted before and after the procedure.ResultsIn vitro, temperature increases at ablation site and at 5-, 10-, and 15-mm distances and ellipsoid-shaped ablated area were proportional to electric power. In the in vivo experiment, the 1.1-W group exhibited significantly greater ablation effects compared to the 0.6-W group, as measured by magnetic resonance imaging (50.6 vs. 22.7 mm²; P <0.001) and histopathology (38.9 vs. 21.7 mm²; P = 0.001). Nearly complete occlusion (>80% stenosis) was more frequently observed in the ablation group (72.2% vs. 22.2%; P = 0.007).ConclusionThe thermal ablation device utilizing a nitinol coil demonstrated effective liver ablation. Furthermore, it shows potential for customization in endovascular ablation applications.
由镍钛镍基线圈记忆合金组成的纤维致动器在正常情况下具有柔性线圈的功能。当电流激活时,它会收缩并产生热量。这一特性使其成为一种有希望用作介入性热器件的候选材料。目的探讨镍钛诺线圈消融器对肝组织及肝内门静脉的热影响。材料与方法以牛肝为材料进行体外实验。将10只日本大白兔按功率设置分为0.6 w和1.1 w两组。每只兔在三个肝脏区域进行消融术。消融区域采用钆注射后脂肪抑制t1加权成像测量,并用苏木精和伊红染色进行组织学检查。肝内门静脉消融9只兔作为消融组,8只兔作为对照组。手术前后分别进行门静脉摄影。结果体外实验中,烧蚀部位及5、10、15 mm处的温度升高与电功率成正比,烧蚀面积呈椭球状。在体内实验中,1.1 w组的消融效果明显优于0.6 w组(50.6 vs 22.7 mm²;P P = 0.001)。消融组几乎完全闭塞(bbb80 %狭窄)发生率更高(72.2% vs. 22.2%; P = 0.007)。结论采用镍钛诺线圈热消融装置对肝脏有较好的消融效果。此外,它显示了在血管内消融应用中定制化的潜力。
期刊介绍:
Acta Radiologica publishes articles on all aspects of radiology, from clinical radiology to experimental work. It is known for articles based on experimental work and contrast media research, giving priority to scientific original papers. The distinguished international editorial board also invite review articles, short communications and technical and instrumental notes.