Measurement Uncertainty Evaluation and Calibration of Electrical Safety Analyzer for Biomedical Equipment

IF 1.4 4区 工程技术 Q4 INSTRUMENTS & INSTRUMENTATION
MAPAN Pub Date : 2025-09-23 DOI:10.1007/s12647-025-00850-w
Sudesh Yadav, A. Hemavathi, Vinod Kumar Tanwar,  Vishesh, Ved Varun Agrawal, Gajjala Sumana,  Rajesh
{"title":"Measurement Uncertainty Evaluation and Calibration of Electrical Safety Analyzer for Biomedical Equipment","authors":"Sudesh Yadav,&nbsp;A. Hemavathi,&nbsp;Vinod Kumar Tanwar,&nbsp; Vishesh,&nbsp;Ved Varun Agrawal,&nbsp;Gajjala Sumana,&nbsp; Rajesh","doi":"10.1007/s12647-025-00850-w","DOIUrl":null,"url":null,"abstract":"<div><p>Calibration of electrical safety analyzers (ESA) are critical for ensuring the safety of medical electrical equipment (MEE) such as syringe pumps, defibrillators, and electrosurgical units, as per international and national standards. Accurate calibration traceable to the international system of units (SI) is essential to validate the safety parameters measured by ESAs. This paper presents the estimation of the measurement uncertainty for the calibration of ESAs for the electrical parameters essential in performing the electrical safety testing on the medical electrical equipment. The uncertainty of the relevant parameters, (i) Resistance (1 Ω) in protective earth resistance test mode, (ii) AC voltage (1–250 V) in point-to-point test mode (iii) DC (10 µA-6.5 mA) &amp; AC current (1000 µA -6.5 mA) in leakage current test mode and related sub modes, (iv) Insulation resistance voltage (250 V &amp; 500 V) and (v) Insulation resistance (0.5–90 MΩ) in insulation resistance test modes and associated sub modes is evaluated and the expanded uncertainty evaluated is 0.001 Ω, 0.1 V, 0.1 µA to 0.01 mA, 1 µA to 0.01 mA, 0.01 V and 0.1 MΩ- 0.4 MΩ respectively. The electrical safety testing has been conducted on key biomedical equipment such as neonatal intensive care incubator, defibrillator and infusion pump and their electrical safety parameters measured using calibrated ESA have been found to be in concurrence with the international electrotechnical commission (IEC) 60601–1.</p></div>","PeriodicalId":689,"journal":{"name":"MAPAN","volume":"40 4","pages":"919 - 928"},"PeriodicalIF":1.4000,"publicationDate":"2025-09-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"MAPAN","FirstCategoryId":"5","ListUrlMain":"https://link.springer.com/article/10.1007/s12647-025-00850-w","RegionNum":4,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0

Abstract

Calibration of electrical safety analyzers (ESA) are critical for ensuring the safety of medical electrical equipment (MEE) such as syringe pumps, defibrillators, and electrosurgical units, as per international and national standards. Accurate calibration traceable to the international system of units (SI) is essential to validate the safety parameters measured by ESAs. This paper presents the estimation of the measurement uncertainty for the calibration of ESAs for the electrical parameters essential in performing the electrical safety testing on the medical electrical equipment. The uncertainty of the relevant parameters, (i) Resistance (1 Ω) in protective earth resistance test mode, (ii) AC voltage (1–250 V) in point-to-point test mode (iii) DC (10 µA-6.5 mA) & AC current (1000 µA -6.5 mA) in leakage current test mode and related sub modes, (iv) Insulation resistance voltage (250 V & 500 V) and (v) Insulation resistance (0.5–90 MΩ) in insulation resistance test modes and associated sub modes is evaluated and the expanded uncertainty evaluated is 0.001 Ω, 0.1 V, 0.1 µA to 0.01 mA, 1 µA to 0.01 mA, 0.01 V and 0.1 MΩ- 0.4 MΩ respectively. The electrical safety testing has been conducted on key biomedical equipment such as neonatal intensive care incubator, defibrillator and infusion pump and their electrical safety parameters measured using calibrated ESA have been found to be in concurrence with the international electrotechnical commission (IEC) 60601–1.

生物医学设备电气安全分析仪测量不确定度评定与校准
根据国际和国家标准,对电气安全分析仪(ESA)进行校准对于确保注射器泵、除颤器和电外科手术装置等医疗电气设备(MEE)的安全至关重要。可溯源至国际单位制(SI)的精确校准对于验证欧空局测量的安全参数至关重要。本文介绍了对医用电气设备进行电气安全试验所必需的电气参数进行欧空局校准时测量不确定度的估计。相关参数的不确定性,(我)电阻(1Ω)在保护接地电阻测试模式,(ii)交流电压(1 - 250 V)在点对点测试模式(3)直流(10µ- 6.5 mA),交流电流(1000µ-6.5 mA)泄漏电流测试模式和相关的子模式,(iv)绝缘电阻电压(250 V, 500 V)和(V)绝缘电阻(0.5 -90Ω)绝缘电阻测试评估模式和相关的子模式和扩展不确定度评估0.001Ω,分别为0.1 V, 0.1µA至0.01 mA, 1µA至0.01 mA, 0.01 V和0.1 MΩ- 0.4 MΩ。对新生儿重症监护培养箱、除颤器、输液泵等关键生物医学设备进行了电气安全测试,使用校准欧空局测量的电气安全参数符合国际电工委员会(IEC) 60601-1标准。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
MAPAN
MAPAN 工程技术-物理:应用
CiteScore
2.30
自引率
20.00%
发文量
91
审稿时长
3 months
期刊介绍: MAPAN-Journal Metrology Society of India is a quarterly publication. It is exclusively devoted to Metrology (Scientific, Industrial or Legal). It has been fulfilling an important need of Metrologists and particularly of quality practitioners by publishing exclusive articles on scientific, industrial and legal metrology. The journal publishes research communication or technical articles of current interest in measurement science; original work, tutorial or survey papers in any metrology related area; reviews and analytical studies in metrology; case studies on reliability, uncertainty in measurements; and reports and results of intercomparison and proficiency testing.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书