Experimental approach for optimizing dose regimen of 68Ga-DOTATATE PET/CT for neuroendocrine tumor (NET) imaging in current high sensitivity scanners: Phantom and Patient Study.
{"title":"Experimental approach for optimizing dose regimen of 68Ga-DOTATATE PET/CT for neuroendocrine tumor (NET) imaging in current high sensitivity scanners: Phantom and Patient Study.","authors":"Abbas Monsef, Fatemeh Saboktakin, Fatemeh Sadeghi, Mohammadreza Elhaie, Mohsen Beheshti, Peyman Sheikhzadeh","doi":"10.1055/a-2710-7001","DOIUrl":null,"url":null,"abstract":"<p><p>This study aimed to determine the optimized scan time and injected activity regimen for clinical <sup>68</sup>Ga DOTATATE PET/CT in neuroendocrine tumor imaging through an experimental approach without using machine learning techniques.A NEMA PET body phantom was used with Ga-<sup>68</sup> to simulate a 9:1 sphere-to-background ratio. PET data were acquired on a high-sensitivity scanner at various scan times (15-300 s/bed). For each scan time, coefficient of variation (COV) and contrast-to-noise ratio (CNR) were calculated. The minimum scan time (T<sub>min</sub>) needed to meet the Rose Criterion (CNR > 5) for the smallest sphere was identified. This T<sub>min</sub> was then applied to patient scans with neuroendocrine tumors (originally acquired at 120 s/bed) to evaluate image quality and determine an optimized activity regimen for clinical <sup>68</sup>Ga-DOTATATE PET imaging.Phantom experiments showed that a COV<sub>max</sub> of 20% is the highest acceptable noise level for detecting the smallest lesions, corresponding to a minimum scan time of about 1 minute per bed position. Patient image analysis confirmed that all tumors visible at routine scan times were still detectable at this minimum duration. This supports the use of a lower activity regimen (~1 MBq/kg), which can reduce patient radiation exposure compared to the standard 1.85 MBq/kg protocol.This work demonstrated that scan time and activity for <sup>68</sup>Ga-DOTATATE NET imaging can be significantly minimized without compromising image interpretation and quantification.</p>","PeriodicalId":94161,"journal":{"name":"Nuklearmedizin. Nuclear medicine","volume":"64 6","pages":"341-351"},"PeriodicalIF":1.2000,"publicationDate":"2025-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nuklearmedizin. Nuclear medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1055/a-2710-7001","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/11/26 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
This study aimed to determine the optimized scan time and injected activity regimen for clinical 68Ga DOTATATE PET/CT in neuroendocrine tumor imaging through an experimental approach without using machine learning techniques.A NEMA PET body phantom was used with Ga-68 to simulate a 9:1 sphere-to-background ratio. PET data were acquired on a high-sensitivity scanner at various scan times (15-300 s/bed). For each scan time, coefficient of variation (COV) and contrast-to-noise ratio (CNR) were calculated. The minimum scan time (Tmin) needed to meet the Rose Criterion (CNR > 5) for the smallest sphere was identified. This Tmin was then applied to patient scans with neuroendocrine tumors (originally acquired at 120 s/bed) to evaluate image quality and determine an optimized activity regimen for clinical 68Ga-DOTATATE PET imaging.Phantom experiments showed that a COVmax of 20% is the highest acceptable noise level for detecting the smallest lesions, corresponding to a minimum scan time of about 1 minute per bed position. Patient image analysis confirmed that all tumors visible at routine scan times were still detectable at this minimum duration. This supports the use of a lower activity regimen (~1 MBq/kg), which can reduce patient radiation exposure compared to the standard 1.85 MBq/kg protocol.This work demonstrated that scan time and activity for 68Ga-DOTATATE NET imaging can be significantly minimized without compromising image interpretation and quantification.