Effect of stent-graft length and compliance on aortic hemodynamics in a bench-top physiological flow circuit

IF 4.1 2区 医学 Q2 ENGINEERING, BIOMEDICAL
Ramin Shahbad, Elizabeth Zermeno, Sayed Ahmadreza Razian, Kaspars Maleckis, Majid Jadidi, Anastasia Desyatova
{"title":"Effect of stent-graft length and compliance on aortic hemodynamics in a bench-top physiological flow circuit","authors":"Ramin Shahbad,&nbsp;Elizabeth Zermeno,&nbsp;Sayed Ahmadreza Razian,&nbsp;Kaspars Maleckis,&nbsp;Majid Jadidi,&nbsp;Anastasia Desyatova","doi":"10.1016/j.jmbbm.2025.107269","DOIUrl":null,"url":null,"abstract":"<div><div>The aortic elasticity plays a vital role in buffering pulsatile blood flow, propelling blood to distal organs and the heart, and reducing cardiac workload. Aortic repair with a stent-graft can reduce this elasticity and hinder the aorta's ability to effectively perform its function. Conventional stent-grafts are associated with increased arterial stiffness, elevated pulse wave velocity (PWV), and adverse hemodynamic changes. This is largely driven by stiffness mismatch between the stent-graft and the native aortic wall, which alters mechanical compliance and hemodynamic response. This study evaluates a novel compliant nanofiber stent-graft (NF-SG) developed to closely mimic native aortic mechanics. Using a bench-top physiological flow circuit, we assessed the hemodynamic impacts of stent-graft stiffness and length on arterial parameters, including PWV, pulse pressure (PP), and distensibility in vitro, and compared these effects with conventional stent-grafts. Stent-graft stiffness significantly affected PWV, PP, and distensibility. Conventional stent-grafts showed 14 %–52 % increase in PWV depending on stent-graft length (p &lt; 0.001), 5 %–32 % increase in PP, and 82 % reduction in mid-graft distensibility. In contrast, NF-SGs maintained PWV and PP near baseline levels with marginal effect of the stent-graft length. Distensibility in the mid-graft was reduced by 13 %–20 %, depending on the stent-graft length. The NF-SG's superior compliance and reduced hemodynamic perturbation were attributed to its mechanically optimized fabric and skeleton design. These findings underscore the clinical potential of the compliant stent-grafts to significantly mitigate long-term cardiovascular complications and preserve aortic functionality post-intervention.</div></div>","PeriodicalId":380,"journal":{"name":"Journal of the Mechanical Behavior of Biomedical Materials","volume":"174 ","pages":"Article 107269"},"PeriodicalIF":4.1000,"publicationDate":"2026-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of the Mechanical Behavior of Biomedical Materials","FirstCategoryId":"5","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1751616125003856","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/11/11 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

The aortic elasticity plays a vital role in buffering pulsatile blood flow, propelling blood to distal organs and the heart, and reducing cardiac workload. Aortic repair with a stent-graft can reduce this elasticity and hinder the aorta's ability to effectively perform its function. Conventional stent-grafts are associated with increased arterial stiffness, elevated pulse wave velocity (PWV), and adverse hemodynamic changes. This is largely driven by stiffness mismatch between the stent-graft and the native aortic wall, which alters mechanical compliance and hemodynamic response. This study evaluates a novel compliant nanofiber stent-graft (NF-SG) developed to closely mimic native aortic mechanics. Using a bench-top physiological flow circuit, we assessed the hemodynamic impacts of stent-graft stiffness and length on arterial parameters, including PWV, pulse pressure (PP), and distensibility in vitro, and compared these effects with conventional stent-grafts. Stent-graft stiffness significantly affected PWV, PP, and distensibility. Conventional stent-grafts showed 14 %–52 % increase in PWV depending on stent-graft length (p < 0.001), 5 %–32 % increase in PP, and 82 % reduction in mid-graft distensibility. In contrast, NF-SGs maintained PWV and PP near baseline levels with marginal effect of the stent-graft length. Distensibility in the mid-graft was reduced by 13 %–20 %, depending on the stent-graft length. The NF-SG's superior compliance and reduced hemodynamic perturbation were attributed to its mechanically optimized fabric and skeleton design. These findings underscore the clinical potential of the compliant stent-grafts to significantly mitigate long-term cardiovascular complications and preserve aortic functionality post-intervention.
支架长度和顺应性对台式生理血流回路主动脉血流动力学的影响
主动脉弹性在缓冲搏动血流、推动血液流向远端器官和心脏、减少心脏负荷等方面起着至关重要的作用。主动脉支架移植修复会降低这种弹性,阻碍主动脉有效发挥其功能。常规支架移植物与动脉僵硬度增加、脉波速度(PWV)升高和不利的血流动力学改变有关。这在很大程度上是由支架移植物与原生主动脉壁之间的刚度不匹配引起的,这改变了机械顺应性和血流动力学反应。这项研究评估了一种新型柔顺纳米纤维支架移植物(NF-SG)的发展,以密切模仿天然主动脉力学。使用台式生理血流回路,我们评估了支架移植刚度和长度对动脉参数的血流动力学影响,包括PWV、脉压(PP)和体外扩张性,并将这些影响与传统支架移植进行了比较。支架移植物刚度显著影响PWV、PP和膨胀性。传统支架移植显示,根据支架长度不同,PWV增加14% - 52% (p < 0.001), PP增加5% - 32%,移植物中期膨胀率降低82%。相比之下,NF-SGs维持PWV和PP接近基线水平,支架长度影响甚微。根据支架长度的不同,中间移植物的扩张率降低了13% - 20%。NF-SG优越的顺应性和减少的血流动力学扰动归功于其机械优化的织物和骨架设计。这些发现强调了顺应性支架移植的临床潜力,可以显著减轻长期心血管并发症,并在干预后保持主动脉功能。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of the Mechanical Behavior of Biomedical Materials
Journal of the Mechanical Behavior of Biomedical Materials 工程技术-材料科学:生物材料
CiteScore
7.20
自引率
7.70%
发文量
505
审稿时长
46 days
期刊介绍: The Journal of the Mechanical Behavior of Biomedical Materials is concerned with the mechanical deformation, damage and failure under applied forces, of biological material (at the tissue, cellular and molecular levels) and of biomaterials, i.e. those materials which are designed to mimic or replace biological materials. The primary focus of the journal is the synthesis of materials science, biology, and medical and dental science. Reports of fundamental scientific investigations are welcome, as are articles concerned with the practical application of materials in medical devices. Both experimental and theoretical work is of interest; theoretical papers will normally include comparison of predictions with experimental data, though we recognize that this may not always be appropriate. The journal also publishes technical notes concerned with emerging experimental or theoretical techniques, letters to the editor and, by invitation, review articles and papers describing existing techniques for the benefit of an interdisciplinary readership.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书