Characteristic of volume-controlled ventilation with small tide volume in hyperbaric oxygen chamber.

IF 0.6 4区 医学 Q4 MARINE & FRESHWATER BIOLOGY
Undersea and Hyperbaric Medicine Pub Date : 2025-03-01
Cong Wang, Lei Zhang, Qiuhong Yu, Yaling Liu, Ziqi Ren, Lianbi Xue
{"title":"Characteristic of volume-controlled ventilation with small tide volume in hyperbaric oxygen chamber.","authors":"Cong Wang, Lei Zhang, Qiuhong Yu, Yaling Liu, Ziqi Ren, Lianbi Xue","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the characteristics of tidal volume (VT) delivered by a Shangrila590 ventilator with preset tide volume (VTset) 50-300 mL in volume-controlled ventilation (VCV) during hyperbaric oxygen (HBO₂) therapy.</p><p><strong>Methods: </strong>Experiments were conducted in a multi-place HBO₂ chamber at 1.0, 2.0, and 2.8 atmospheres absolute (ATA abs) and were divided into three groups accordingly. The ventilator was connected to the test lung in the chamber. The VTset of the ventilator was adjusted before the experiments. At five VTset levels (50, 100, 150, 200, 300 mL), the VT and inspiratory peak pressure (Ppeak) detected by the ventilator and the test lung were recorded for 30 cycles (n=30). The measurements of the test lung were considered to be the true value, and the ventilator and test lung data were compared to evaluate the accuracy of the ventilator. Test lung compliance (CTL) was detected by the ventilator, and breathing circuit compliance (CBC) was calculated by measuring the pressure and volume of the breathing circuit. Ventilation data were compared among three groups to clarify the change during HBO₂.</p><p><strong>Results: </strong>At every VTset, the VT detected by the test lung was different from the ventilator at 1.0~2.8 atm abs (p<0.05), and the VT changed differently among the three groups (p<0.05). CTL and the CBC decreased with increasing ambient pressure (p<0.05). The Ppeak of VCV increased (p<0.05) with increasing ambient pressure.</p><p><strong>Conclusions: </strong>The Shangrila590 ventilator has limitations in small VTset (50- 300 mL) VCV during HBO₂ therapy. CBC and CTL levels change due to high gas density. High Ppeak may occur in VCV with stable VT, also caused by high gas density.</p>","PeriodicalId":49396,"journal":{"name":"Undersea and Hyperbaric Medicine","volume":"52 3","pages":"305-312"},"PeriodicalIF":0.6000,"publicationDate":"2025-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Undersea and Hyperbaric Medicine","FirstCategoryId":"3","ListUrlMain":"","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MARINE & FRESHWATER BIOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To evaluate the characteristics of tidal volume (VT) delivered by a Shangrila590 ventilator with preset tide volume (VTset) 50-300 mL in volume-controlled ventilation (VCV) during hyperbaric oxygen (HBO₂) therapy.

Methods: Experiments were conducted in a multi-place HBO₂ chamber at 1.0, 2.0, and 2.8 atmospheres absolute (ATA abs) and were divided into three groups accordingly. The ventilator was connected to the test lung in the chamber. The VTset of the ventilator was adjusted before the experiments. At five VTset levels (50, 100, 150, 200, 300 mL), the VT and inspiratory peak pressure (Ppeak) detected by the ventilator and the test lung were recorded for 30 cycles (n=30). The measurements of the test lung were considered to be the true value, and the ventilator and test lung data were compared to evaluate the accuracy of the ventilator. Test lung compliance (CTL) was detected by the ventilator, and breathing circuit compliance (CBC) was calculated by measuring the pressure and volume of the breathing circuit. Ventilation data were compared among three groups to clarify the change during HBO₂.

Results: At every VTset, the VT detected by the test lung was different from the ventilator at 1.0~2.8 atm abs (p<0.05), and the VT changed differently among the three groups (p<0.05). CTL and the CBC decreased with increasing ambient pressure (p<0.05). The Ppeak of VCV increased (p<0.05) with increasing ambient pressure.

Conclusions: The Shangrila590 ventilator has limitations in small VTset (50- 300 mL) VCV during HBO₂ therapy. CBC and CTL levels change due to high gas density. High Ppeak may occur in VCV with stable VT, also caused by high gas density.

高压氧舱小潮气量气量控制通风特性研究。
目的:评价容积控制通气(VCV)时,Shangrila590呼吸机潮汐量(VTset)为50 ~ 300 mL的潮汐量(VT)在高压氧(HBO₂)治疗过程中的特点。方法:在1、2、2.8个大气压(ataabs)下的多地点高压氧室内进行实验,并将实验分为3组。呼吸机与室内的试验肺相连。实验前对呼吸机的VTset进行了调整。在5个VTset水平(50、100、150、200、300 mL)下,记录呼吸机和试验肺检测的VT和吸气峰压(Ppeak),共30个周期(n=30)。测试肺的测量值被认为是真实值,并将呼吸机和测试肺数据进行比较,以评估呼吸机的准确性。呼吸机检测试验肺顺应性(CTL),通过测量呼吸回路的压力和容积计算呼吸回路顺应性(CBC)。比较三组患者的通气数据,以明确HBO 2治疗期间的变化。结果:每次VTset时,试验肺检测到的VT与呼吸机在1.0~2.8 atm abs时不同。结论:Shangrila590呼吸机在小VTset (50 ~ 300 mL)的VCV治疗中存在局限性。CBC和CTL水平因高气体密度而改变。在VT稳定的VCV中可能出现高峰值,这也是由高气体密度引起的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Undersea and Hyperbaric Medicine
Undersea and Hyperbaric Medicine 医学-海洋与淡水生物学
CiteScore
1.60
自引率
11.10%
发文量
37
审稿时长
>12 weeks
期刊介绍: Undersea and Hyperbaric Medicine Journal accepts manuscripts for publication that are related to the areas of diving research and physiology, hyperbaric medicine and oxygen therapy, submarine medicine, naval medicine and clinical research related to the above topics. To be considered for UHM scientific papers must deal with significant and new research in an area related to biological, physical and clinical phenomena related to the above environments.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书