Feasibility study of a multi-lesion cyberknife radiotherapy plan verification method using a 2D array with pre-set roll angles.

IF 3.6 Q4 Medicine
Precision Radiation Oncology Pub Date : 2025-06-28 eCollection Date: 2025-09-01 DOI:10.1002/pro6.70022
Tengxiang Li, Jinhu Chen, Ruimin He, Qingtao Qiu, Quan Tang, Yong Yin
{"title":"Feasibility study of a multi-lesion cyberknife radiotherapy plan verification method using a 2D array with pre-set roll angles.","authors":"Tengxiang Li, Jinhu Chen, Ruimin He, Qingtao Qiu, Quan Tang, Yong Yin","doi":"10.1002/pro6.70022","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>When validating intracranial multi-lesion CyberKnife M6/S7 plans with SRSmapcheck, setting the array to a fixed 0° measures only one target dose distribution, leaving the other lesions unmeasured. Moreover, the CyberKnife treatment planning system does not support roll verification tools, and testing confirms that X-sight fiducial marker guidance is incompatible with free array roll. A novel method and workflow are required to validate multi-lesion plans with random positions.</p><p><strong>Methods: </strong>A geometric model was established based on the relationship between SRSmapcheck and the tumor location. For two tumors spaced 77 mm apart (each 20 mm in diameter, or one 40 mm apart and the other infinitesimally small), the corresponding array roll angle interval was approximately 15.05°. The SRSmapcheck and StereoPHAN computed tomography (CT) images were acquired at 15° intervals, starting at 0°, and preprocessed into phantom plans for verification. A total of 101 intracranial multi-lesion plans were verified using the fixed 0° and pre-set roll angle methods to optimize the dose distribution, particularly in high-dose and rapidly varying areas. A two-sample test compared the results of the 0° versus pre-set roll angle verification and assessed the performance under different criteria to determine suitable criteria for pre-set roll angle verification.</p><p><strong>Results: </strong>The equivalent diameter of the 296 tumors ranged from 5 to 45 mm (average: 21.86 mm). Each plan had an average of 2.97 lesions. Using the pre-set roll angle method, 2.34 targets were assessed on average (89.83% of lesions had diameters ranging from 10 to 40 mm), compared to 1.32 targets on average in 0° plans. Statistically significant differences occurred at 2 mm/1% and 2 mm/2% in the γ analysis, showing that plan pass rates were stable between the fixed 0° and pre-set roll angle methods. Relaxing either the distance to agreement or dose deviation significantly increased the pass rates during pre-set roll angle verification, whereas cross-transforming criteria had minimal impact. For pre-set roll angle methods, it is recommended to use 1 mm/1% (action limit: 86.0% ± 13.3%) and 1 mm/2% (action limits: 91.6% ± 7.9%) for γ analysis.</p><p><strong>Conclusion: </strong>SRSmapcheck with the pre-set roll angle method can verify intracranial multi-lesion CyberKnife plans by measuring multiple targets in a single validation and comparing the 1 mm/1% and 1 mm/2% γ analysis criteria.</p>","PeriodicalId":32406,"journal":{"name":"Precision Radiation Oncology","volume":"9 3","pages":"167-176"},"PeriodicalIF":3.6000,"publicationDate":"2025-06-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12559907/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Precision Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1002/pro6.70022","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/9/1 0:00:00","PubModel":"eCollection","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Background: When validating intracranial multi-lesion CyberKnife M6/S7 plans with SRSmapcheck, setting the array to a fixed 0° measures only one target dose distribution, leaving the other lesions unmeasured. Moreover, the CyberKnife treatment planning system does not support roll verification tools, and testing confirms that X-sight fiducial marker guidance is incompatible with free array roll. A novel method and workflow are required to validate multi-lesion plans with random positions.

Methods: A geometric model was established based on the relationship between SRSmapcheck and the tumor location. For two tumors spaced 77 mm apart (each 20 mm in diameter, or one 40 mm apart and the other infinitesimally small), the corresponding array roll angle interval was approximately 15.05°. The SRSmapcheck and StereoPHAN computed tomography (CT) images were acquired at 15° intervals, starting at 0°, and preprocessed into phantom plans for verification. A total of 101 intracranial multi-lesion plans were verified using the fixed 0° and pre-set roll angle methods to optimize the dose distribution, particularly in high-dose and rapidly varying areas. A two-sample test compared the results of the 0° versus pre-set roll angle verification and assessed the performance under different criteria to determine suitable criteria for pre-set roll angle verification.

Results: The equivalent diameter of the 296 tumors ranged from 5 to 45 mm (average: 21.86 mm). Each plan had an average of 2.97 lesions. Using the pre-set roll angle method, 2.34 targets were assessed on average (89.83% of lesions had diameters ranging from 10 to 40 mm), compared to 1.32 targets on average in 0° plans. Statistically significant differences occurred at 2 mm/1% and 2 mm/2% in the γ analysis, showing that plan pass rates were stable between the fixed 0° and pre-set roll angle methods. Relaxing either the distance to agreement or dose deviation significantly increased the pass rates during pre-set roll angle verification, whereas cross-transforming criteria had minimal impact. For pre-set roll angle methods, it is recommended to use 1 mm/1% (action limit: 86.0% ± 13.3%) and 1 mm/2% (action limits: 91.6% ± 7.9%) for γ analysis.

Conclusion: SRSmapcheck with the pre-set roll angle method can verify intracranial multi-lesion CyberKnife plans by measuring multiple targets in a single validation and comparing the 1 mm/1% and 1 mm/2% γ analysis criteria.

Abstract Image

Abstract Image

Abstract Image

一种多病变射波刀放疗计划验证方法的可行性研究。
背景:在使用SRSmapcheck验证颅内多病变射波刀M6/S7计划时,将阵列设置为固定的0°仅测量一个目标剂量分布,而不测量其他病变。此外,射波刀治疗计划系统不支持滚转验证工具,测试证实X-sight基准标记制导与自由阵列滚转不兼容。需要一种新的方法和工作流程来验证具有随机位置的多病变计划。方法:根据SRSmapcheck与肿瘤位置的关系建立几何模型。对于两个间距为77 mm的肿瘤(每个直径为20 mm,或一个间距为40 mm,另一个为无穷小),相应的阵列滚转角间隔约为15.05°。从0°开始,以15°间隔获取SRSmapcheck和StereoPHAN计算机断层扫描(CT)图像,并将其预处理成模拟图进行验证。采用固定0°和预设滚转角的方法对101个颅内多病变平面进行验证,以优化剂量分布,特别是在高剂量和快速变化的区域。通过两个样本的测试,比较了0°和预设滚转角验证的结果,并评估了不同标准下的性能,以确定适合的预设滚转角验证标准。结果:296例肿瘤的等效直径为5 ~ 45mm,平均21.86 mm。每个方案平均有2.97个病灶。使用预先设定的滚动角度方法,平均评估2.34个目标(89.83%的病灶直径在10至40 mm之间),而0°计划平均评估1.32个目标。在2 mm/1%和2 mm/2%的γ分析中,有统计学意义的差异,表明在固定的0°和预先设置的滚转角方法之间,计划通过率是稳定的。在预先设定的滚转角验证过程中,放宽达到一致的距离或剂量偏差显著提高了通过率,而交叉转换标准的影响最小。对于预先设定的滚转角方法,建议使用1 mm/1%(作用限值:86.0%±13.3%)和1 mm/2%(作用限值:91.6%±7.9%)进行γ分析。结论:SRSmapcheck采用预先设定滚角方法,通过在一次验证中测量多个靶点,并比较1 mm/1%和1 mm/2% γ分析标准,可以验证颅内多病变射波刀方案。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Precision Radiation Oncology
Precision Radiation Oncology Medicine-Oncology
CiteScore
1.20
自引率
0.00%
发文量
32
审稿时长
13 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书