Visual performance of a non-diffractive extended depth of focus intraocular lens

IF 1.1 4区 医学 Q3 OPHTHALMOLOGY
F.-X. Crahay
{"title":"Visual performance of a non-diffractive extended depth of focus intraocular lens","authors":"F.-X. Crahay","doi":"10.1016/j.jfo.2025.104651","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><div>This study aimed to evaluate the monocular and binocular visual performance of a new non-diffractive extended depth of focus (EDoF) intraocular lens (IOL).</div></div><div><h3>Methods</h3><div>In this prospective, non-randomized, single-arm, single-center, open-label study, the eyes of consecutive patients who received bilateral implantation of a non-diffractive EDoF IOL (Acrysof IQ Vivity, Alcon Laboratories) were evaluated. Monocular and binocular best corrected and uncorrected distance, intermediate, and near visual acuity (VA), as well as monocular defocus curve and contrast sensitivity, were assessed.</div></div><div><h3>Results</h3><div>Seventy-six eyes of 38 patients were included. The mean subjective spherical equivalent was −0.13. Binocular distance-corrected VA was −0.16 LogMAR at 6 meters, 0.12 LogMAR at 66 centimeters, and 0.35 LogMAR at 40 centimeters. Monocular distance-corrected VA was −0.10 LogMAR, 0.20 LogMAR, and 0.45 LogMAR at the same respective distances. Uncorrected VA at distance, intermediate, and near were −0.08 LogMAR, 0.14 LogMAR, and 0.33 LogMAR for binocular vision, and 0.00 LogMAR, 0.20 LogMAR, and 0.43 LogMAR for monocular vision. The monocular defocus curve was below 0.08 LogMAR up to −1.5 D defocus and below 0.25 LogMAR up to −2.0 D defocus. Average monocular contrast sensitivity peaked at 3.00 cycles/degree at 0.92 log units, decreasing at higher spatial frequencies. Most patients reported no visual disturbances post-implantation; 8% reported ghosting, and 5% reported glare.</div></div><div><h3>Conclusion</h3><div>This non-diffractive extended depth of focus intraocular lens provides good visual acuity and quality at far and intermediate distances, functional near vision, and high spectacle-independent outcomes with minimal visual disturbances.</div></div><div><h3>Objectif de l’étude</h3><div>L’étude visait à évaluer les performances visuelles monoculaires et binoculaires d’une nouvelle lentille intraoculaire (LIO) non diffractive à profondeur de champ étendue (EDoF).</div></div><div><h3>Méthodes</h3><div>Dans cette étude prospective, non randomisée, monocentrique et ouverte, les yeux de patients consécutifs ayant reçu une implantation bilatérale d’une LIO EDoF non diffractive (<em>Acrysof IQ Vivity, Alcon Laboratories</em>) ont été évalués. L’acuité visuelle (AV) monoculaire et binoculaire, corrigée et non corrigée, de loin, à distance intermédiaire et de près, ainsi que la courbe de défocalisation monoculaire et la sensibilité aux contrastes ont été évaluées.</div></div><div><h3>Résultats</h3><div>Soixante-seize yeux de 38 patients ont été inclus. L’équivalent sphérique subjectif moyen était de −0,13. L’AV corrigée de la distance binoculaire était de −0,16 LogMAR à 6 mètres, de 0,12 LogMAR à 66 centimètres et de 0,35 LogMAR à 40 centimètres. L’AV monoculaire corrigée de la distance était de −0,10 LogMAR, 0,20 LogMAR et 0,45 LogMAR aux mêmes distances. L’AV non corrigée à distance, à distance intermédiaire et de près était de −0,08 LogMAR, 0,14 LogMAR et 0,33 LogMAR pour la vision binoculaire et de 0,00 LogMAR, 0,20 LogMAR et 0,43 LogMAR pour la vision monoculaire. La courbe de défocalisation monoculaire était inférieure à 0,08 LogMAR jusqu’à −1,5 D de défocalisation et inférieure à 0,25 LogMAR jusqu’à −2,0 D de défocalisation. La sensibilité moyenne au contraste monoculaire a atteint son maximum à 3,00 cycles/degrés à 0,92 unités logarithmiques, diminuant à des fréquences spatiales plus élevées. La plupart des patients n’ont signalé aucun trouble visuel après l’implantation ; 8 % ont signalé des images fantômes et 5 % des éblouissements.</div></div><div><h3>Conclusion</h3><div>Cette lentille intraoculaire non diffractive à profondeur de champ étendue permet d’obtenir une bonne acuité visuelle et une bonne qualité visuelle de loin et à distance intermédiaire, une vision de près fonctionnelle et des résultats élevés en matière d’indépendance vis-à-vis des lunettes avec des troubles visuels minimes.</div></div>","PeriodicalId":14777,"journal":{"name":"Journal Francais D Ophtalmologie","volume":"48 9","pages":"Article 104651"},"PeriodicalIF":1.1000,"publicationDate":"2025-10-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal Francais D Ophtalmologie","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0181551225002335","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose

This study aimed to evaluate the monocular and binocular visual performance of a new non-diffractive extended depth of focus (EDoF) intraocular lens (IOL).

Methods

In this prospective, non-randomized, single-arm, single-center, open-label study, the eyes of consecutive patients who received bilateral implantation of a non-diffractive EDoF IOL (Acrysof IQ Vivity, Alcon Laboratories) were evaluated. Monocular and binocular best corrected and uncorrected distance, intermediate, and near visual acuity (VA), as well as monocular defocus curve and contrast sensitivity, were assessed.

Results

Seventy-six eyes of 38 patients were included. The mean subjective spherical equivalent was −0.13. Binocular distance-corrected VA was −0.16 LogMAR at 6 meters, 0.12 LogMAR at 66 centimeters, and 0.35 LogMAR at 40 centimeters. Monocular distance-corrected VA was −0.10 LogMAR, 0.20 LogMAR, and 0.45 LogMAR at the same respective distances. Uncorrected VA at distance, intermediate, and near were −0.08 LogMAR, 0.14 LogMAR, and 0.33 LogMAR for binocular vision, and 0.00 LogMAR, 0.20 LogMAR, and 0.43 LogMAR for monocular vision. The monocular defocus curve was below 0.08 LogMAR up to −1.5 D defocus and below 0.25 LogMAR up to −2.0 D defocus. Average monocular contrast sensitivity peaked at 3.00 cycles/degree at 0.92 log units, decreasing at higher spatial frequencies. Most patients reported no visual disturbances post-implantation; 8% reported ghosting, and 5% reported glare.

Conclusion

This non-diffractive extended depth of focus intraocular lens provides good visual acuity and quality at far and intermediate distances, functional near vision, and high spectacle-independent outcomes with minimal visual disturbances.

Objectif de l’étude

L’étude visait à évaluer les performances visuelles monoculaires et binoculaires d’une nouvelle lentille intraoculaire (LIO) non diffractive à profondeur de champ étendue (EDoF).

Méthodes

Dans cette étude prospective, non randomisée, monocentrique et ouverte, les yeux de patients consécutifs ayant reçu une implantation bilatérale d’une LIO EDoF non diffractive (Acrysof IQ Vivity, Alcon Laboratories) ont été évalués. L’acuité visuelle (AV) monoculaire et binoculaire, corrigée et non corrigée, de loin, à distance intermédiaire et de près, ainsi que la courbe de défocalisation monoculaire et la sensibilité aux contrastes ont été évaluées.

Résultats

Soixante-seize yeux de 38 patients ont été inclus. L’équivalent sphérique subjectif moyen était de −0,13. L’AV corrigée de la distance binoculaire était de −0,16 LogMAR à 6 mètres, de 0,12 LogMAR à 66 centimètres et de 0,35 LogMAR à 40 centimètres. L’AV monoculaire corrigée de la distance était de −0,10 LogMAR, 0,20 LogMAR et 0,45 LogMAR aux mêmes distances. L’AV non corrigée à distance, à distance intermédiaire et de près était de −0,08 LogMAR, 0,14 LogMAR et 0,33 LogMAR pour la vision binoculaire et de 0,00 LogMAR, 0,20 LogMAR et 0,43 LogMAR pour la vision monoculaire. La courbe de défocalisation monoculaire était inférieure à 0,08 LogMAR jusqu’à −1,5 D de défocalisation et inférieure à 0,25 LogMAR jusqu’à −2,0 D de défocalisation. La sensibilité moyenne au contraste monoculaire a atteint son maximum à 3,00 cycles/degrés à 0,92 unités logarithmiques, diminuant à des fréquences spatiales plus élevées. La plupart des patients n’ont signalé aucun trouble visuel après l’implantation ; 8 % ont signalé des images fantômes et 5 % des éblouissements.

Conclusion

Cette lentille intraoculaire non diffractive à profondeur de champ étendue permet d’obtenir une bonne acuité visuelle et une bonne qualité visuelle de loin et à distance intermédiaire, une vision de près fonctionnelle et des résultats élevés en matière d’indépendance vis-à-vis des lunettes avec des troubles visuels minimes.

Abstract Image

非衍射扩展聚焦深度人工晶状体的视觉表现
目的评价新型无衍射扩展焦深人工晶状体(IOL)的单眼和双眼视觉性能。方法在这项前瞻性、非随机、单臂、单中心、开放标签的研究中,对连续接受双侧无衍射EDoF人工晶体(Acrysof IQ Vivity,爱尔康实验室)植入的患者的眼睛进行评估。评估单眼和双眼最佳矫正和未矫正距离、中间和近视敏度(VA)以及单眼离焦曲线和对比敏感度。结果纳入38例患者76只眼。平均主观球当量为−0.13。双目距离校正后的VA在6米处为- 0.16 LogMAR,在66厘米处为0.12 LogMAR,在40厘米处为0.35 LogMAR。在相同距离下,单眼距离校正后的VA分别为- 0.10 LogMAR、0.20 LogMAR和0.45 LogMAR。双眼视力的未校正VA分别为- 0.08、0.14和0.33 LogMAR,单眼视力的未校正VA分别为0.00、0.20和0.43 LogMAR。单眼离焦曲线在- 1.5 D离焦范围内低于0.08 LogMAR,在- 2.0 D离焦范围内低于0.25 LogMAR。平均单眼对比灵敏度在0.92对数单位时达到3.00周期/度的峰值,在更高的空间频率下下降。大多数患者报告植入后无视力障碍;8%的人报告了鬼影,5%的人报告了眩光。结论该无衍射扩展焦深度人工晶状体具有良好的中远视力和质量,具有良好的近视力功能,具有高度的眼镜不依赖性,视觉障碍最小。目的:研究 ······················3种类型的 ()和 () - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -单目镜和双筒望远镜、矫正型和非矫正型、矫正型和非矫正型、矫正型和非矫正型、矫正型和非矫正型、矫正型和非矫正型、矫正型和矫正型、矫正型和矫正型、矫正型和矫正型。包括38例患者在内的所有病例。L 'equivalent spherique主观的平均是de−0,13岁。L 'AV校正了远距离双筒望远镜的 远程通讯通讯系统- 0,16 - LogMAR 6厘米,- 0,12 - LogMAR 66厘米,- 0,35 - LogMAR 40厘米。L 'AV monoculaire修正了- 0,10 LogMAR, 0,20 LogMAR和0,45 LogMAR aux mêmes距离。L 'AV非corrigee距离,距离intermediaire et de总统是de−0,08年LogMAR, 0, 14 LogMAR et 0, 33 LogMAR pour la愿景binoculaire 0, 00 LogMAR, 0, 20 LogMAR et 0, 43 LogMAR pour la monoculaire愿景。La courbe de dd录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带录影带1 .最大数为3 000个周期/度,单位/对数,最小数为3 000个周期/度,单位/对数,最小数为3 000个周期/度,单位/空间/次,单位/空间/次,单位/空间/次。La plupart des患者没有信号,没有视觉障碍;8%的人不使用信号处理图像fantômes, 5%的人不使用电子信号处理图像。结论黄豆眼内无衍射视差、无衍射视差、无衍射视差、无衍射视差、无衍射视差、无衍射视差、无衍射视差、无衍射视差、无衍射视差、无衍射视差、无衍射视差、无衍射视差、无衍射视差、无衍射视差、无衍射视差。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
1.10
自引率
8.30%
发文量
317
审稿时长
49 days
期刊介绍: The Journal français d''ophtalmologie, official publication of the French Society of Ophthalmology, serves the French Speaking Community by publishing excellent research articles, communications of the French Society of Ophthalmology, in-depth reviews, position papers, letters received by the editor and a rich image bank in each issue. The scientific quality is guaranteed through unbiased peer-review, and the journal is member of the Committee of Publication Ethics (COPE). The editors strongly discourage editorial misconduct and in particular if duplicative text from published sources is identified without proper citation, the submission will not be considered for peer review and returned to the authors or immediately rejected.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信