Clinical Assessment of a Virtual Reality Perimeter Versus the Humphrey Field Analyzer: Comparative Reliability, Usability, and Prospective Applications.
Marco Zeppieri, Caterina Gagliano, Francesco Cappellani, Federico Visalli, Fabiana D'Esposito, Alessandro Avitabile, Roberta Amato, Alessandra Cuna, Francesco Pellegrini
{"title":"Clinical Assessment of a Virtual Reality Perimeter Versus the Humphrey Field Analyzer: Comparative Reliability, Usability, and Prospective Applications.","authors":"Marco Zeppieri, Caterina Gagliano, Francesco Cappellani, Federico Visalli, Fabiana D'Esposito, Alessandro Avitabile, Roberta Amato, Alessandra Cuna, Francesco Pellegrini","doi":"10.3390/vision9040086","DOIUrl":null,"url":null,"abstract":"<p><p><i>Background:</i> This study compared the performance of a Head-mounted Virtual Reality Perimeter (HVRP) with the Humphrey Field Analyzer (HFA), the standard in automated perimetry. The HFA is the established standard for automated perimetry but is constrained by lengthy testing, bulky equipment, and limited patient comfort. Comparative data on newer head-mounted virtual reality perimeters are limited, leaving uncertainty about their clinical reliability and potential advantages. <i>Aim:</i> The aim was to evaluate parameters such as visual field outcomes, portability, patient comfort, eye tracking, and usability. <i>Methods:</i> Participants underwent testing with both devices, assessing metrics like mean deviation (MD), pattern standard deviation (PSD), and duration. <i>Results:</i> The HVRP demonstrated small but statistically significant differences in MD and PSD compared to the HFA, while maintaining a consistent trend across participants. MD values were slightly more negative for HFA than HVRP (average difference -0.60 dB, <i>p</i> = 0.0006), while pattern standard deviation was marginally higher with HFA (average difference 0.38 dB, <i>p</i> = 0.00018). Although statistically significant, these differences were small in magnitude and do not undermine the clinical utility or reproducibility of the device. Notably, HVRP showed markedly shorter testing times with HVRP (7.15 vs. 18.11 min, mean difference 10.96 min, <i>p</i> < 0.0001). Its lightweight, portable design allowed for bedside and home testing, enhancing accessibility for pediatric, geriatric, and mobility-impaired patients. Participants reported greater comfort due to the headset design, which eliminated the need for chin rests. The device also offers potential for AI integration and remote data analysis. <i>Conclusions:</i> The HVRP proved to be a reliable, user-friendly alternative to traditional perimetry. Its advantages in comfort, portability, and test efficiency support its use in both clinical settings and remote screening programs for visual field assessment. Its portability and user-friendly design support broader use in clinical practice and expand possibilities for bedside assessment, home monitoring, and remote screening, particularly in populations with limited access to conventional perimetry.</p>","PeriodicalId":36586,"journal":{"name":"Vision (Switzerland)","volume":"9 4","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2025-10-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12550896/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Vision (Switzerland)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3390/vision9040086","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Background: This study compared the performance of a Head-mounted Virtual Reality Perimeter (HVRP) with the Humphrey Field Analyzer (HFA), the standard in automated perimetry. The HFA is the established standard for automated perimetry but is constrained by lengthy testing, bulky equipment, and limited patient comfort. Comparative data on newer head-mounted virtual reality perimeters are limited, leaving uncertainty about their clinical reliability and potential advantages. Aim: The aim was to evaluate parameters such as visual field outcomes, portability, patient comfort, eye tracking, and usability. Methods: Participants underwent testing with both devices, assessing metrics like mean deviation (MD), pattern standard deviation (PSD), and duration. Results: The HVRP demonstrated small but statistically significant differences in MD and PSD compared to the HFA, while maintaining a consistent trend across participants. MD values were slightly more negative for HFA than HVRP (average difference -0.60 dB, p = 0.0006), while pattern standard deviation was marginally higher with HFA (average difference 0.38 dB, p = 0.00018). Although statistically significant, these differences were small in magnitude and do not undermine the clinical utility or reproducibility of the device. Notably, HVRP showed markedly shorter testing times with HVRP (7.15 vs. 18.11 min, mean difference 10.96 min, p < 0.0001). Its lightweight, portable design allowed for bedside and home testing, enhancing accessibility for pediatric, geriatric, and mobility-impaired patients. Participants reported greater comfort due to the headset design, which eliminated the need for chin rests. The device also offers potential for AI integration and remote data analysis. Conclusions: The HVRP proved to be a reliable, user-friendly alternative to traditional perimetry. Its advantages in comfort, portability, and test efficiency support its use in both clinical settings and remote screening programs for visual field assessment. Its portability and user-friendly design support broader use in clinical practice and expand possibilities for bedside assessment, home monitoring, and remote screening, particularly in populations with limited access to conventional perimetry.
背景:本研究比较了头戴式虚拟现实周界(HVRP)与自动周界测量标准Humphrey Field Analyzer (HFA)的性能。HFA是自动视野检查的既定标准,但受到测试时间长、设备笨重和患者舒适度有限的限制。新的头戴式虚拟现实周边的比较数据有限,留下了不确定的临床可靠性和潜在的优势。目的:目的是评估诸如视野结果、便携性、患者舒适度、眼动追踪和可用性等参数。方法:参与者接受两种设备的测试,评估指标如平均偏差(MD)、模式标准差(PSD)和持续时间。结果:与HFA相比,HVRP在MD和PSD方面表现出较小但具有统计学意义的差异,同时在参与者中保持一致的趋势。HFA的MD值略高于HVRP(平均差值-0.60 dB, p = 0.0006),而HFA的模式标准差略高于HVRP(平均差值0.38 dB, p = 0.00018)。虽然具有统计学意义,但这些差异幅度很小,不会影响该装置的临床实用性或可重复性。值得注意的是,HVRP与HVRP的检测时间明显缩短(7.15 vs. 18.11 min,平均差10.96 min, p < 0.0001)。其轻巧、便携的设计允许床边和家庭测试,提高儿童、老年人和行动不便患者的可及性。参与者报告说,由于耳机的设计更舒适,不需要下巴托。该设备还提供了人工智能集成和远程数据分析的潜力。结论:HVRP被证明是一个可靠的,用户友好的替代传统的视野。它在舒适性、便携性和测试效率方面的优势支持它在临床环境和远程视野评估筛查项目中的使用。它的便携性和用户友好的设计支持在临床实践中更广泛的应用,并扩大了床边评估、家庭监测和远程筛查的可能性,特别是在无法获得常规视野检查的人群中。