Comparison of Pulse Oximetry and Blood Pressure Measurements Between Sensor Locations in Children During Dental Procedures.

IF 1.7
Pediatric dentistry Pub Date : 2025-09-15
Brianne N Cipich, Juan F Yepes, Stephen Wilson, Allison C Scully, LaQuia A Walker Vinson, James E Jones, George Eckert
{"title":"Comparison of Pulse Oximetry and Blood Pressure Measurements Between Sensor Locations in Children During Dental Procedures.","authors":"Brianne N Cipich, Juan F Yepes, Stephen Wilson, Allison C Scully, LaQuia A Walker Vinson, James E Jones, George Eckert","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p><b>Purpose:</b> To compare oxygen saturation measurements when pulse oximeter sensors were simultaneously placed on the index finger and the second toe, and compare blood pressure recordings with non-invasive blood pressure (NIBP) cuffs simultaneously placed on the upper arm and the lower calf in children receiving general anesthesia. <b>Methods:</b> A total of 50 children, aged three to nine years, defined as ASA Class I or II receiving comprehensive dental treatment in the operating room with general anesthesia, were included. Pulse oximeter sensors were simultaneously placed on each subject's index finger and ipsilateral second toe. Non-invasive blood pressure cuffs were simultaneously placed on the upper arm and ipsilateral lower calf above the ankle. The same anesthesia regimen was used on all patients. Following induction of general anesthesia, measurements were simultaneously recorded every five minutes during the dental procedure. <b>Results:</b> Index finger oxygen saturation measurements were significantly lower than the second toe (P<0.001). Arm systolic blood pressure measurements were significantly lower than calf measurements (P<0.001). Arm diastolic blood pressure measurements were significantly lower than calf measurements (P<0.001). <b>Conclusions:</b> Discrepancy of vital sign measurements may be notable depending on the site from which measurements are obtained. If blood pressure is taken on the calf of a child in the Trendelenburg position, the reading will likely produce a higher systolic and diastolic measurement and should be interpreted with caution. A pulse oximeter placed on the first finger or second toe of healthy children can be regarded as interchangeable in a clinical setting.</p>","PeriodicalId":101357,"journal":{"name":"Pediatric dentistry","volume":"47 5","pages":"322-327"},"PeriodicalIF":1.7000,"publicationDate":"2025-09-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pediatric dentistry","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To compare oxygen saturation measurements when pulse oximeter sensors were simultaneously placed on the index finger and the second toe, and compare blood pressure recordings with non-invasive blood pressure (NIBP) cuffs simultaneously placed on the upper arm and the lower calf in children receiving general anesthesia. Methods: A total of 50 children, aged three to nine years, defined as ASA Class I or II receiving comprehensive dental treatment in the operating room with general anesthesia, were included. Pulse oximeter sensors were simultaneously placed on each subject's index finger and ipsilateral second toe. Non-invasive blood pressure cuffs were simultaneously placed on the upper arm and ipsilateral lower calf above the ankle. The same anesthesia regimen was used on all patients. Following induction of general anesthesia, measurements were simultaneously recorded every five minutes during the dental procedure. Results: Index finger oxygen saturation measurements were significantly lower than the second toe (P<0.001). Arm systolic blood pressure measurements were significantly lower than calf measurements (P<0.001). Arm diastolic blood pressure measurements were significantly lower than calf measurements (P<0.001). Conclusions: Discrepancy of vital sign measurements may be notable depending on the site from which measurements are obtained. If blood pressure is taken on the calf of a child in the Trendelenburg position, the reading will likely produce a higher systolic and diastolic measurement and should be interpreted with caution. A pulse oximeter placed on the first finger or second toe of healthy children can be regarded as interchangeable in a clinical setting.

儿童牙科手术中传感器位置间脉搏氧饱和度和血压测量值的比较。
目的:比较脉搏血氧计传感器同时放置在食指和第二趾时的血氧饱和度测量结果,并比较在全麻儿童上臂和小腿下部同时放置无创血压(NIBP)袖带的血压记录。方法:选取ASA I类或II类患儿50例,年龄3 ~ 9岁,在手术室内接受全麻牙科综合治疗。脉搏血氧计传感器同时放置在每个受试者的食指和同侧第二个脚趾上。无创血压袖带同时置于上臂和同侧小腿脚踝上方。所有患者均采用相同的麻醉方案。在全身麻醉诱导后,在牙科手术过程中每五分钟同时记录一次测量结果。结果:食指血氧饱和度测量值明显低于第二趾(p结论:生命体征测量值的差异可能因测量位置的不同而显著。如果在Trendelenburg体位的孩子的小腿上测量血压,读数可能会产生更高的收缩压和舒张压,应该谨慎解释。将脉搏血氧计放在健康儿童的第一根手指或第二根脚趾上,在临床环境中可被视为可互换的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书