Mechanical assessment of a titanium cervical spine corpectomy cage assembled with 3D-printed patient-specific endplate-conformed contact surfaces and a traditionally manufactured expandable mechanism.

IF 3.1 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Shih-Chieh Shen, Shao-Fu Huang, Wei-Hsiang Sun, Chun-Li Lin
{"title":"Mechanical assessment of a titanium cervical spine corpectomy cage assembled with 3D-printed patient-specific endplate-conformed contact surfaces and a traditionally manufactured expandable mechanism.","authors":"Shih-Chieh Shen, Shao-Fu Huang, Wei-Hsiang Sun, Chun-Li Lin","doi":"10.1186/s41205-025-00299-2","DOIUrl":null,"url":null,"abstract":"<p><p>This study developed a new titanium cervical spine expandable corpectomy cage (ECC) that consisted of two superior/inferior patient-specific endplate-conformed contact surfaces (PECSs) fabricated through 3D printing and a standardized centrally expandable mechanism (CEM) manufactured using traditional machining. Fatigue testing was conducted to evaluate whether the components of ECC could be assembled using different manufacturing methods to meet the functional testing requirements in compliance with FDA regulations. The titanium 3D-printed superior/inferior PECSs were designed based on the CT images/CAD system to assembly with a CNC milling/precision wire cutting cuboid CEM (14 mm square cross-section) expansion driven by a set of sliders interlocked via key and keyway mechanisms for the ECC of two-segmental vertebral bodies. The ECC underwent FDA-compliant static/dynamic compression, shear, torsion, subsidence tests and finite element (FE) analysis for understanding the stress differences on endplates between PECS and flat-shaped endplate contact surface (FECS). Stiffness/yield strength were found to be 2127 ± 146 N/mm/8547 ± 1213 N for the compression test, 1158 ± 139 N/mm/ 2561 ± 114 N for the shear test, and 1.22 ± 0.32 Nm/deg/16.5 ± 0.58 Nm for the torque test. Some failure samples showed that fixation screws were fractured/loosened under the shear test. The endure limits were 1600 N, 350 N and 1.0 N*m under compression, shear and torsion cyclic load tests, respectively. The results of the static axial compression and static/dynamic compression-shear testing did not meet the acceptance criteria of ISO 23,089. The PECS with higher stiffness value showed that better subsidence resistance was achieved than FECS and was consistent with that the maximum stress value and distribution for the FECS were more harmful than those for the PECS models. This study demonstrated that mechanical testing is essential when assembling 3D-printed components with CNC-manufactured parts in multi-component medical implants, as mismatches in interface behavior may result in mechanical failure. The results of FE analysis and subsidence tests indicated that PECS can decrease stress concentration between the ECC and endplates to present better performance for subsidence resistance.</p>","PeriodicalId":72036,"journal":{"name":"3D printing in medicine","volume":"11 1","pages":"50"},"PeriodicalIF":3.1000,"publicationDate":"2025-10-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12548260/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"3D printing in medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s41205-025-00299-2","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

This study developed a new titanium cervical spine expandable corpectomy cage (ECC) that consisted of two superior/inferior patient-specific endplate-conformed contact surfaces (PECSs) fabricated through 3D printing and a standardized centrally expandable mechanism (CEM) manufactured using traditional machining. Fatigue testing was conducted to evaluate whether the components of ECC could be assembled using different manufacturing methods to meet the functional testing requirements in compliance with FDA regulations. The titanium 3D-printed superior/inferior PECSs were designed based on the CT images/CAD system to assembly with a CNC milling/precision wire cutting cuboid CEM (14 mm square cross-section) expansion driven by a set of sliders interlocked via key and keyway mechanisms for the ECC of two-segmental vertebral bodies. The ECC underwent FDA-compliant static/dynamic compression, shear, torsion, subsidence tests and finite element (FE) analysis for understanding the stress differences on endplates between PECS and flat-shaped endplate contact surface (FECS). Stiffness/yield strength were found to be 2127 ± 146 N/mm/8547 ± 1213 N for the compression test, 1158 ± 139 N/mm/ 2561 ± 114 N for the shear test, and 1.22 ± 0.32 Nm/deg/16.5 ± 0.58 Nm for the torque test. Some failure samples showed that fixation screws were fractured/loosened under the shear test. The endure limits were 1600 N, 350 N and 1.0 N*m under compression, shear and torsion cyclic load tests, respectively. The results of the static axial compression and static/dynamic compression-shear testing did not meet the acceptance criteria of ISO 23,089. The PECS with higher stiffness value showed that better subsidence resistance was achieved than FECS and was consistent with that the maximum stress value and distribution for the FECS were more harmful than those for the PECS models. This study demonstrated that mechanical testing is essential when assembling 3D-printed components with CNC-manufactured parts in multi-component medical implants, as mismatches in interface behavior may result in mechanical failure. The results of FE analysis and subsidence tests indicated that PECS can decrease stress concentration between the ECC and endplates to present better performance for subsidence resistance.

由3d打印患者特异性终板符合接触面和传统制造的可扩展机构组装的钛颈椎椎体切除术笼的力学评估。
本研究开发了一种新型的钛合金颈椎可扩展椎体切除笼(ECC),该笼由两个通过3D打印制造的优/差患者特异性终板符合接触面(pics)和一个采用传统加工制造的标准化中央可扩展机构(CEM)组成。通过疲劳试验,评估采用不同的制造方法组装ECC组件是否能够满足FDA规定的功能测试要求。基于CT图像/CAD系统设计钛材3d打印的优质/优质pics,装配在一组通过键和键槽机构连锁的滑块驱动的数控铣削/精密线切割长方体CEM (14 mm平方横截面)扩展装置上,用于两节段椎体的ECC。ECC进行了符合fda标准的静/动态压缩、剪切、扭转、沉降测试和有限元(FE)分析,以了解PECS和平板端板接触面(FECS)之间端板的应力差异。压缩试验的刚度/屈服强度为2127±146 N/mm/8547±1213 N,剪切试验的刚度/屈服强度为1158±139 N/mm/ 2561±114 N,扭矩试验的刚度/屈服强度为1.22±0.32 Nm/度/16.5±0.58 Nm。部分破坏试样在剪切试验中出现固定螺钉断裂/松动现象。压缩、剪切和扭转循环荷载试验的耐受力极限分别为1600 N、350 N和1.0 N*m。静轴压和静/动压剪试验结果不符合ISO 23,089的验收标准。具有较高刚度值的PECS比FECS具有更好的抗沉降能力,这与FECS模型的最大应力值和最大应力分布比PECS模型的更有害是一致的。该研究表明,在多部件医疗植入物中,3d打印部件与cnc制造部件组装时,机械测试是必不可少的,因为界面行为的不匹配可能导致机械故障。有限元分析和沉降试验结果表明,PECS可以降低ECC与端板之间的应力集中,具有较好的抗沉降性能。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
审稿时长
5 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书