{"title":"Comparison of Refractive Outcomes Between Small Lenticule Extraction Using a 500 KHz and a 2 MHz Femtosecond Laser.","authors":"Jung Wan Kim, Youngsub Eom, Dong Hyun Kim, Ji Won Jeong, Seh Kwang Park, Jong Suk Song","doi":"10.3390/diagnostics15192450","DOIUrl":null,"url":null,"abstract":"<p><p><b>Background/Objectives</b>: The aim of this study was to compare the efficacy of correcting myopia and myopic astigmatism between small incision lenticule extraction (SMILE) surgery performed using VisuMax 500 and that using VisuMax 800. <b>Methods</b>: Patients who underwent myopia correction using either VisuMax 500 (493 eyes of 249 patients) or VisuMax 800 (169 eyes of 85 patients) employing the nomogram of VisuMax 500 were enrolled in this retrospective case-control study. At 2 months after the operation, the percentage of eyes achieving a postoperative uncorrected distance visual acuity (UDVA) of 20/20 and residual refractive error were compared between the two groups. Additionally, the percentage of eyes with refractive astigmatism angle of error within ±15° and the mean absolute angle of error were analyzed. <b>Results</b>: In the VisuMax 500 and 800 groups, 99% and 97% of eyes achieved a postoperative UDVA of 20/20, respectively. Postoperative residual astigmatism was similar between the two groups, but residual myopia was significantly lower in the VisuMax 800 group (-0.09 ± 0.50 cylinder diopters [CD]) compared to the VisuMax 500 group (-0.21 ± 0.49 CD; <i>p</i>-value = 0.005). Additionally, 84% of eyes in the VisuMax 800 group achieved astigmatism correction within ±15° of the intended meridian, significantly outperforming the 71% in the VisuMax 500 group (<i>p</i>-value = 0.002), with a significantly smaller mean absolute angle of error (8.3 ± 12.2° and 14.1 ± 20.1°; <i>p</i>-value < 0.001). <b>Conclusions</b>: Both VisuMax 500 and VisuMax 800 effectively corrected myopia. However, in terms of the accuracy of astigmatism axis correction, VisuMax 800 demonstrated superior precision compared to VisuMax 500. This study may be limited by perceived bias associated with evaluating two platforms from the same manufacturer.</p>","PeriodicalId":11225,"journal":{"name":"Diagnostics","volume":"15 19","pages":""},"PeriodicalIF":3.3000,"publicationDate":"2025-09-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12523262/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Diagnostics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3390/diagnostics15192450","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0

Abstract

Background/Objectives: The aim of this study was to compare the efficacy of correcting myopia and myopic astigmatism between small incision lenticule extraction (SMILE) surgery performed using VisuMax 500 and that using VisuMax 800. Methods: Patients who underwent myopia correction using either VisuMax 500 (493 eyes of 249 patients) or VisuMax 800 (169 eyes of 85 patients) employing the nomogram of VisuMax 500 were enrolled in this retrospective case-control study. At 2 months after the operation, the percentage of eyes achieving a postoperative uncorrected distance visual acuity (UDVA) of 20/20 and residual refractive error were compared between the two groups. Additionally, the percentage of eyes with refractive astigmatism angle of error within ±15° and the mean absolute angle of error were analyzed. Results: In the VisuMax 500 and 800 groups, 99% and 97% of eyes achieved a postoperative UDVA of 20/20, respectively. Postoperative residual astigmatism was similar between the two groups, but residual myopia was significantly lower in the VisuMax 800 group (-0.09 ± 0.50 cylinder diopters [CD]) compared to the VisuMax 500 group (-0.21 ± 0.49 CD; p-value = 0.005). Additionally, 84% of eyes in the VisuMax 800 group achieved astigmatism correction within ±15° of the intended meridian, significantly outperforming the 71% in the VisuMax 500 group (p-value = 0.002), with a significantly smaller mean absolute angle of error (8.3 ± 12.2° and 14.1 ± 20.1°; p-value < 0.001). Conclusions: Both VisuMax 500 and VisuMax 800 effectively corrected myopia. However, in terms of the accuracy of astigmatism axis correction, VisuMax 800 demonstrated superior precision compared to VisuMax 500. This study may be limited by perceived bias associated with evaluating two platforms from the same manufacturer.

Abstract Image

Abstract Image

Abstract Image

500 KHz和2 MHz飞秒激光小透镜提取的折射结果比较。
背景/目的:本研究的目的是比较VisuMax 500和VisuMax 800在小切口晶状体摘除(SMILE)手术中的矫正近视和近视眼散光的效果。方法:选择使用VisuMax 500(249例493眼)或VisuMax 800(85例169眼)进行近视矫正的患者,采用VisuMax 500的nomogram矫正方法,进行回顾性病例对照研究。术后2个月,比较两组患者术后未矫正距离视力(UDVA)为20/20及残余屈光不正的百分率。此外,还分析了屈光散光误差角在±15°以内的眼睛比例和平均绝对误差角。结果:在VisuMax 500和800组中,99%和97%的眼睛分别达到了20/20的术后UDVA。两组术后残留散光相似,但VisuMax 800组(-0.09±0.50圆柱体屈光度[CD])明显低于VisuMax 500组(-0.21±0.49 CD, p值= 0.005)。此外,VisuMax 800组中84%的眼睛在预定子午线±15°范围内实现了散光矫正,显著优于VisuMax 500组的71% (p值= 0.002),平均绝对误差角明显较小(8.3±12.2°和14.1±20.1°,p值< 0.001)。结论:VisuMax 500和VisuMax 800均能有效矫正近视。然而,在散光轴校正精度方面,VisuMax 800比VisuMax 500表现出更高的精度。本研究可能受到与评估来自同一制造商的两个平台相关的感知偏差的限制。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Diagnostics
Diagnostics Biochemistry, Genetics and Molecular Biology-Clinical Biochemistry
CiteScore
4.70
自引率
8.30%
发文量
2699
审稿时长
19.64 days
期刊介绍: Diagnostics (ISSN 2075-4418) is an international scholarly open access journal on medical diagnostics. It publishes original research articles, reviews, communications and short notes on the research and development of medical diagnostics. There is no restriction on the length of the papers. Our aim is to encourage scientists to publish their experimental and theoretical research in as much detail as possible. Full experimental and/or methodological details must be provided for research articles.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信