{"title":"Echocardiographic assessment of a novel blood-guiding accessory for left ventricular assist devices in swine.","authors":"Michelle Costa Galbas, Florian Meissner, Hendrik Straky, Johanna Reuter, Marius Schimmel, Martin Czerny, Wolfgang Bothe","doi":"10.1093/icvts/ivaf168","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>Left ventricular assist devices are part of the foundations of mechanical circulatory support. As such devices evolve in technology, so does their implantation techniques, becoming less invasive. We developed a novel blood-guiding accessory for less invasive implantation, redirecting the blood intracardially into the ascending aorta through a transaortic outflow.</p><p><strong>Methods: </strong>Ten healthy female swine (German Landrace, 104.4 ± 13.0 kg) underwent epicardial echocardiography before and after left ventricular assist device implantation. Cardiac geometry, aortic valve function and outflow positioning, were assessed before and after implantation.</p><p><strong>Results: </strong>Epicardial echocardiography revealed sufficient biventricular unloading after device implantation. The end-diastolic and end-systolic left ventricular diameters decreased by 16% and 20%, respectively. Retrograde unloading was displayed by smaller diameters in left and right atria at 22% and 25%, respectively. At baseline, one animal presented mild aortic regurgitation, unchanged under mechanical support, whereas one developed mild de novo insufficiency. The outflow graft was either well centered within the aortic valve (n = 3) or between two cusps.</p><p><strong>Conclusions: </strong>A left ventricular assist device with the accessory allowed sufficient ventricular unloading, preserving biventricular function. Although the outflow graft was well centered within the aortic valve in only 30% of the cases, no relevant aortic regurgitation was found in the acute setting. Chronic testing in larger samples is required to analyze results in long term.</p>","PeriodicalId":73406,"journal":{"name":"Interdisciplinary cardiovascular and thoracic surgery","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2025-10-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Interdisciplinary cardiovascular and thoracic surgery","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1093/icvts/ivaf168","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"0","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: Left ventricular assist devices are part of the foundations of mechanical circulatory support. As such devices evolve in technology, so does their implantation techniques, becoming less invasive. We developed a novel blood-guiding accessory for less invasive implantation, redirecting the blood intracardially into the ascending aorta through a transaortic outflow.
Methods: Ten healthy female swine (German Landrace, 104.4 ± 13.0 kg) underwent epicardial echocardiography before and after left ventricular assist device implantation. Cardiac geometry, aortic valve function and outflow positioning, were assessed before and after implantation.
Results: Epicardial echocardiography revealed sufficient biventricular unloading after device implantation. The end-diastolic and end-systolic left ventricular diameters decreased by 16% and 20%, respectively. Retrograde unloading was displayed by smaller diameters in left and right atria at 22% and 25%, respectively. At baseline, one animal presented mild aortic regurgitation, unchanged under mechanical support, whereas one developed mild de novo insufficiency. The outflow graft was either well centered within the aortic valve (n = 3) or between two cusps.
Conclusions: A left ventricular assist device with the accessory allowed sufficient ventricular unloading, preserving biventricular function. Although the outflow graft was well centered within the aortic valve in only 30% of the cases, no relevant aortic regurgitation was found in the acute setting. Chronic testing in larger samples is required to analyze results in long term.