Comparison of oscillometry ratios between IOS and AOS in patients with asthma and COPD

IF 3.1 3区 医学 Q2 CARDIAC & CARDIOVASCULAR SYSTEMS
Philipp Suter, Robert Greig, Chris RuiWen Kuo, Rory Chan, Brian Lipworth
{"title":"Comparison of oscillometry ratios between IOS and AOS in patients with asthma and COPD","authors":"Philipp Suter,&nbsp;Robert Greig,&nbsp;Chris RuiWen Kuo,&nbsp;Rory Chan,&nbsp;Brian Lipworth","doi":"10.1016/j.rmed.2025.108402","DOIUrl":null,"url":null,"abstract":"<div><div>The forced oscillation technique (FOT) enables effort-independent assessment of small airway dysfunction (SAD) in obstructive lung diseases. We compared resistance and reactance ratios between two FOT modalities—impulse oscillometry (IOS) and airwave oscillometry (AOS), in patients with asthma and COPD. We retrospectively analysed paired pre- and post-bronchodilator IOS and AOS measurements from 82 patients (58 asthma, 24 COPD). Resistance (R5–R20/R5 for IOS; R5–R19/R5 for AOS) and reactance (X5/AX) ratios were compared using correlation and Bland-Altman analyses. Resistance ratios showed significant agreement between devices, with minimal mean differences. Reactance ratios were also significantly correlated but were higher with IOS than AOS. This difference increased at higher X5/AX ratios, likely due to device specific signal characteristics and calibration differences. Post-bronchodilator improvements were observed with both devices, with a greater change in X5/AX using IOS (p &lt; 0.05). In conclusion resistance ratios were comparable between IOS and AOS, supporting their clinical use across devices. In contrast, reactance ratios differed, highlighting the need for device-specific normative values and improved standardization. FOT-derived ratios may offer a practical alternative to absolute values for assessing SAD in asthma and COPD.</div></div>","PeriodicalId":21057,"journal":{"name":"Respiratory medicine","volume":"248 ","pages":"Article 108402"},"PeriodicalIF":3.1000,"publicationDate":"2025-10-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Respiratory medicine","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0954611125004652","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

The forced oscillation technique (FOT) enables effort-independent assessment of small airway dysfunction (SAD) in obstructive lung diseases. We compared resistance and reactance ratios between two FOT modalities—impulse oscillometry (IOS) and airwave oscillometry (AOS), in patients with asthma and COPD. We retrospectively analysed paired pre- and post-bronchodilator IOS and AOS measurements from 82 patients (58 asthma, 24 COPD). Resistance (R5–R20/R5 for IOS; R5–R19/R5 for AOS) and reactance (X5/AX) ratios were compared using correlation and Bland-Altman analyses. Resistance ratios showed significant agreement between devices, with minimal mean differences. Reactance ratios were also significantly correlated but were higher with IOS than AOS. This difference increased at higher X5/AX ratios, likely due to device specific signal characteristics and calibration differences. Post-bronchodilator improvements were observed with both devices, with a greater change in X5/AX using IOS (p < 0.05). In conclusion resistance ratios were comparable between IOS and AOS, supporting their clinical use across devices. In contrast, reactance ratios differed, highlighting the need for device-specific normative values and improved standardization. FOT-derived ratios may offer a practical alternative to absolute values for assessing SAD in asthma and COPD.
哮喘和慢性阻塞性肺病患者IOS和AOS的振荡比比较。
强迫振荡技术(FOT)使阻塞性肺疾病的小气道功能障碍(SAD)的评估无需费力。我们比较了哮喘和慢性阻塞性肺病患者的两种FOT模式-脉冲振荡法(IOS)和电波振荡法(AOS)的电阻和电抗比。我们回顾性分析了82例患者(58例哮喘,24例COPD)使用支气管扩张剂前后的配对IOS和AOS测量值。电阻(IOS为R5- r20 /R5; AOS为R5- r19 /R5)和电抗(X5/AX)比值采用相关分析和Bland-Altman分析进行比较。电阻比在器件之间显示出显著的一致性,平均差异极小。电抗比与IOS也有显著相关,但与IOS相关的电抗比高于AOS。这种差异在较高的X5/AX比率下增加,可能是由于设备特定的信号特性和校准差异。两种设备均观察到支气管扩张剂后的改善,使用IOS时X5/AX的变化更大(p
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Respiratory medicine
Respiratory medicine 医学-呼吸系统
CiteScore
7.50
自引率
0.00%
发文量
199
审稿时长
38 days
期刊介绍: Respiratory Medicine is an internationally-renowned journal devoted to the rapid publication of clinically-relevant respiratory medicine research. It combines cutting-edge original research with state-of-the-art reviews dealing with all aspects of respiratory diseases and therapeutic interventions. Topics include adult and paediatric medicine, epidemiology, immunology and cell biology, physiology, occupational disorders, and the role of allergens and pollutants. Respiratory Medicine is increasingly the journal of choice for publication of phased trial work, commenting on effectiveness, dosage and methods of action.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信