{"title":"Reliability of liver stiffness measurements using shear wave elastography: impact of skin-to-capsule distance and subcutaneous tissue structure.","authors":"Keisuke Osakabe, Keiko Sugimoto, Hiroji Takai, Yusuke Sano, Keisuke Maeda, Koji Suzuki, Naohiro Ichino","doi":"10.1007/s10396-025-01582-9","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Shear wave velocity (Vs) measurements can be unstable in individuals with obesity or with a skin-capsule distance (SCD) of ≥ 20 mm. However, instability of Vs has also been observed in cases where SCD is not markedly increased; therefore, the present study aimed to investigate the factors contributing to reduced VsN (net amount of effective shear wave velocity), an index that reflects the reliability of individual Vs measurements.</p><p><strong>Methods: </strong>In total, 317 individuals who underwent community-based health screening were included in this study. We investigated the relationship between VsN and physical findings, laboratory data, SCD, and structure of the SCD (single vs. multilayered) to identify factors associated with decreased VsN.</p><p><strong>Results: </strong>Multiple regression analysis revealed that body fat percentage (β = - 0.135, p = 0.011), SCD (β = - 0.347, p < 0.0001), and the structure of the SC region (β = - 0.295, p < 0.0001) were independently associated with decreased VsN. In cases where the SCD exceeded 20 mm, the VsN within the SCD was significantly lower in the multilayer structural type [52 (28-75)] than in the single-layer structural type [95 (77-99)] (p = 0.004). Furthermore, even when the SCD was 20 mm or less, the VsN within the SCD was significantly lower in the multilayer structure type [80 (68-89)] than in the single-layer structure type [99 (92-100)] (p < 0.0001).</p><p><strong>Conclusion: </strong>The results show that, in addition to SCD thickness, its internal structure reduces the VsN and contributes to Vs measurement instability.</p>","PeriodicalId":50130,"journal":{"name":"Journal of Medical Ultrasonics","volume":" ","pages":""},"PeriodicalIF":2.1000,"publicationDate":"2025-10-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Ultrasonics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10396-025-01582-9","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: Shear wave velocity (Vs) measurements can be unstable in individuals with obesity or with a skin-capsule distance (SCD) of ≥ 20 mm. However, instability of Vs has also been observed in cases where SCD is not markedly increased; therefore, the present study aimed to investigate the factors contributing to reduced VsN (net amount of effective shear wave velocity), an index that reflects the reliability of individual Vs measurements.
Methods: In total, 317 individuals who underwent community-based health screening were included in this study. We investigated the relationship between VsN and physical findings, laboratory data, SCD, and structure of the SCD (single vs. multilayered) to identify factors associated with decreased VsN.
Results: Multiple regression analysis revealed that body fat percentage (β = - 0.135, p = 0.011), SCD (β = - 0.347, p < 0.0001), and the structure of the SC region (β = - 0.295, p < 0.0001) were independently associated with decreased VsN. In cases where the SCD exceeded 20 mm, the VsN within the SCD was significantly lower in the multilayer structural type [52 (28-75)] than in the single-layer structural type [95 (77-99)] (p = 0.004). Furthermore, even when the SCD was 20 mm or less, the VsN within the SCD was significantly lower in the multilayer structure type [80 (68-89)] than in the single-layer structure type [99 (92-100)] (p < 0.0001).
Conclusion: The results show that, in addition to SCD thickness, its internal structure reduces the VsN and contributes to Vs measurement instability.
期刊介绍:
The Journal of Medical Ultrasonics is the official journal of the Japan Society of Ultrasonics in Medicine. The main purpose of the journal is to provide forum for the publication of papers documenting recent advances and new developments in the entire field of ultrasound in medicine and biology, encompassing both the medical and the engineering aspects of the science.The journal welcomes original articles, review articles, images, and letters to the editor.The journal also provides state-of-the-art information such as announcements from the boards and the committees of the society.