B-058 Evaluating the analytical performance of the OsmoPRO® MAX Automated Osmometer

IF 6.3 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY
Lisa Salvucci
{"title":"B-058 Evaluating the analytical performance of the OsmoPRO® MAX Automated Osmometer","authors":"Lisa Salvucci","doi":"10.1093/clinchem/hvaf086.456","DOIUrl":null,"url":null,"abstract":"Background Accurate and timely serum, plasma, and urine osmolality measurements are crucial for diagnosing and treating conditions such as hyponatremia, toxic alcohol ingestion, and diabetes insipidus. Most clinical osmometers rely on manual methods which require pipetting and testing one sample, control, or calibration standard at a time. These manual steps are time-consuming and increase the risk of human error. A new osmometer, the OsmoPRO® MAX, saves time and reduces these risks by automating freezing-point depression measurements (Advanced Instruments, LLC). The purpose of this study was to evaluate the analytical performance of the OsmoPRO MAX in measuring the osmolality of serum, plasma, and urine. Methods OsmoPRO MAX analytical performance was evaluated at Boston Medical Center in Boston, Massachusetts using four criteria: (1) Linearity was assessed with an allowable systematic error of <2 mOsm/kg H2O and an allowable total error of <6.0 mOsm/kg H2O over a range of 0-2000 mOsm/kg H2O using an Osmolality Linearity Set. (2) Simple Accuracy was evaluated over a range of 240 to 800 mOsm/kg H2O using Clinitrol™ 290 with an acceptable range of +/- 4 mOsm/kg H2O as well as Protinol™ 240, 280, and 320, each with an acceptable range of +/- 7 mOsm/kg H2O, and Renol™ 300 and 800, both with an acceptable range of +/- 10 mOsm/kg H2O. (3) Simple Precision was measured using Clinitrol 290 with an acceptable within-run standard deviation (SD) of <2.0 mOsm/kg H2O as well as Protinol 240, 280, and 320, each with an acceptable within-run SD of <3 mOsm/kg H2O, and Renol 300 and 800, with an acceptable within-run SD of <3 mOsm/kg H2O and <4.7 mOsm/kg H2O, respectively. (4) Alternative Method Comparison was conducted against the OsmoPRO® Multi-Sample Micro-Osmometer on 80 patient samples analyzed by Passing-Bablok regression analysis. Results OsmoPRO MAX passed all criteria. Linearity testing produced a negligible observable error of 0.1 mOsm/kg H2O, accurate within the allowable systematic error and all results were accurate within the total allowable error. Simple Accuracy testing produced replicates all within their specified target range. Simple Precision testing of all replicates for each standard tested fell within their respective within-run SD. Alternative Quantitative Method Comparison demonstrated strong agreement between the two methods with a correlation coefficient of 0.9998 and a bias of -2.1 (-0.6%). Conclusion This study demonstrates that the OsmoPRO MAX Automated Osmometer meets all acceptance criteria for analytical performance including linearity, simple accuracy, simple precision, and alternative quantitative method comparison against the OsmoPRO Multi-Sample Micro-Osmometer.","PeriodicalId":10690,"journal":{"name":"Clinical chemistry","volume":"109 1","pages":""},"PeriodicalIF":6.3000,"publicationDate":"2025-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical chemistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/clinchem/hvaf086.456","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background Accurate and timely serum, plasma, and urine osmolality measurements are crucial for diagnosing and treating conditions such as hyponatremia, toxic alcohol ingestion, and diabetes insipidus. Most clinical osmometers rely on manual methods which require pipetting and testing one sample, control, or calibration standard at a time. These manual steps are time-consuming and increase the risk of human error. A new osmometer, the OsmoPRO® MAX, saves time and reduces these risks by automating freezing-point depression measurements (Advanced Instruments, LLC). The purpose of this study was to evaluate the analytical performance of the OsmoPRO MAX in measuring the osmolality of serum, plasma, and urine. Methods OsmoPRO MAX analytical performance was evaluated at Boston Medical Center in Boston, Massachusetts using four criteria: (1) Linearity was assessed with an allowable systematic error of <2 mOsm/kg H2O and an allowable total error of <6.0 mOsm/kg H2O over a range of 0-2000 mOsm/kg H2O using an Osmolality Linearity Set. (2) Simple Accuracy was evaluated over a range of 240 to 800 mOsm/kg H2O using Clinitrol™ 290 with an acceptable range of +/- 4 mOsm/kg H2O as well as Protinol™ 240, 280, and 320, each with an acceptable range of +/- 7 mOsm/kg H2O, and Renol™ 300 and 800, both with an acceptable range of +/- 10 mOsm/kg H2O. (3) Simple Precision was measured using Clinitrol 290 with an acceptable within-run standard deviation (SD) of <2.0 mOsm/kg H2O as well as Protinol 240, 280, and 320, each with an acceptable within-run SD of <3 mOsm/kg H2O, and Renol 300 and 800, with an acceptable within-run SD of <3 mOsm/kg H2O and <4.7 mOsm/kg H2O, respectively. (4) Alternative Method Comparison was conducted against the OsmoPRO® Multi-Sample Micro-Osmometer on 80 patient samples analyzed by Passing-Bablok regression analysis. Results OsmoPRO MAX passed all criteria. Linearity testing produced a negligible observable error of 0.1 mOsm/kg H2O, accurate within the allowable systematic error and all results were accurate within the total allowable error. Simple Accuracy testing produced replicates all within their specified target range. Simple Precision testing of all replicates for each standard tested fell within their respective within-run SD. Alternative Quantitative Method Comparison demonstrated strong agreement between the two methods with a correlation coefficient of 0.9998 and a bias of -2.1 (-0.6%). Conclusion This study demonstrates that the OsmoPRO MAX Automated Osmometer meets all acceptance criteria for analytical performance including linearity, simple accuracy, simple precision, and alternative quantitative method comparison against the OsmoPRO Multi-Sample Micro-Osmometer.
B-058评估OsmoPRO®MAX自动渗透仪的分析性能
准确和及时的血清、血浆和尿液渗透压测量对于诊断和治疗低钠血症、中毒性酒精摄入和尿崩症等疾病至关重要。大多数临床渗透仪依赖于手动方法,需要移液和测试一个样品,控制,或校准标准一次。这些手动步骤非常耗时,并且增加了人为错误的风险。一种新的渗透计,OsmoPRO®MAX,通过自动测量冰点降低,节省了时间,降低了这些风险(Advanced Instruments, LLC)。本研究的目的是评价OsmoPRO MAX在测定血清、血浆和尿液渗透压方面的分析性能。方法采用4个标准评价OsmoPRO MAX在马萨诸塞州波士顿医疗中心的分析性能:(1)线性评价,允许系统误差为&;lt;使用渗透压线性集,在0-2000 mOsm/kg H2O范围内,允许总误差为&;lt;6.0 mOsm/kg H2O。(2)使用Clinitrol™290(可接受范围为+/- 4 mOsm/kg H2O)以及Protinol™240、280和320(可接受范围为+/- 7 mOsm/kg H2O)和Renol™300和800(可接受范围均为+/- 10 mOsm/kg H2O)在240至800 mOsm/kg H2O范围内进行简单精度评估。(3)使用Clinitrol 290和Protinol 240、280和320测定简单精密度(Simple Precision),其可接受的运行标准偏差(SD)为2.0 mOsm/kg H2O;3 mOsm/kg H2O,雷诺300和800,运行内可接受的SD为&;lt;3 mOsm/kg H2O和4.7 mOsm/kg H2O。(4)替代方法采用passingbablok回归分析80例患者样本,与OsmoPRO®多样本微渗仪进行对比。结果OsmoPRO MAX符合各项标准。线性测试产生的可忽略的观测误差为0.1 mOsm/kg H2O,准确在允许的系统误差范围内,所有结果都准确在总允许误差范围内。简单的准确度测试产生的副本都在指定的目标范围内。简单精密度测试的所有重复的每个标准测试落在各自的运行SD内。两种方法的对比结果一致,相关系数为0.9998,偏差为-2.1(-0.6%)。本研究表明OsmoPRO MAX自动渗透仪符合分析性能的所有验收标准,包括线性度、简单准确度、简单精密度,以及与OsmoPRO多样品微渗透仪的替代定量方法比较。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Clinical chemistry
Clinical chemistry 医学-医学实验技术
CiteScore
11.30
自引率
4.30%
发文量
212
审稿时长
1.7 months
期刊介绍: Clinical Chemistry is a peer-reviewed scientific journal that is the premier publication for the science and practice of clinical laboratory medicine. It was established in 1955 and is associated with the Association for Diagnostics & Laboratory Medicine (ADLM). The journal focuses on laboratory diagnosis and management of patients, and has expanded to include other clinical laboratory disciplines such as genomics, hematology, microbiology, and toxicology. It also publishes articles relevant to clinical specialties including cardiology, endocrinology, gastroenterology, genetics, immunology, infectious diseases, maternal-fetal medicine, neurology, nutrition, oncology, and pediatrics. In addition to original research, editorials, and reviews, Clinical Chemistry features recurring sections such as clinical case studies, perspectives, podcasts, and Q&A articles. It has the highest impact factor among journals of clinical chemistry, laboratory medicine, pathology, analytical chemistry, transfusion medicine, and clinical microbiology. The journal is indexed in databases such as MEDLINE and Web of Science.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信