K. Mitev , B. Sabot , V. Todorov , S. Georgiev , S. Pierre , S. Röttger , B. Krastev , I. Dimitrova
{"title":"On the techniques for primary calibration of electronic radon detectors","authors":"K. Mitev , B. Sabot , V. Todorov , S. Georgiev , S. Pierre , S. Röttger , B. Krastev , I. Dimitrova","doi":"10.1016/j.apradiso.2025.112209","DOIUrl":null,"url":null,"abstract":"<div><div>Successful calibration of RadonEye +2 electronic radon detectors was performed at typical indoor activity concentrations using the facilities at LNHB and PTB. The calibration uncertainties using primary radon activity standards were: below 1.5 % at 300 Bq/m<sup>3</sup>, below 1.7 % at 130 Bq/m<sup>3</sup> and below 2.5 % at 55 Bq/m<sup>3</sup> (<em>k</em> = 1). When using the secondary standard AlphaGUARD, the uncertainty at 55 Bq/m<sup>3</sup> was below 3.5 %. Maintaining stable activity concentrations proved crucial and appears to be the only feasible approach for calibrations below 100 Bq/m<sup>3</sup>. While calibration under exponentially decaying radon activity concentration remains useful for evaluating the devices’ linearity across a broad range, it proved unsuitable for calibration of user-grade monitors at low activities due to the high statistical variation in their signal.</div><div>The linearity of RadonEye +2 was demonstrated in the range 50 Bq/m<sup>3</sup> - 300 Bq/m<sup>3</sup> and they will be utilized for the sensor networks developed within the RadonNET project. Dynamic background correction, applicable to non-spectrometric detectors, was applied based on the monitor's exposure history. Furthermore, it was observed that the pulse-processing algorithm of RadonEyes +2 distorts the Poisson distribution of the signal, thereby increasing its variation. Potentially, lower measurement uncertainty could be achieved with electronic radon detectors that report the registered pulses and allow access to their processing algorithms.</div></div>","PeriodicalId":8096,"journal":{"name":"Applied Radiation and Isotopes","volume":"226 ","pages":"Article 112209"},"PeriodicalIF":1.8000,"publicationDate":"2025-09-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Applied Radiation and Isotopes","FirstCategoryId":"5","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0969804325005548","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CHEMISTRY, INORGANIC & NUCLEAR","Score":null,"Total":0}
引用次数: 0
Abstract
Successful calibration of RadonEye +2 electronic radon detectors was performed at typical indoor activity concentrations using the facilities at LNHB and PTB. The calibration uncertainties using primary radon activity standards were: below 1.5 % at 300 Bq/m3, below 1.7 % at 130 Bq/m3 and below 2.5 % at 55 Bq/m3 (k = 1). When using the secondary standard AlphaGUARD, the uncertainty at 55 Bq/m3 was below 3.5 %. Maintaining stable activity concentrations proved crucial and appears to be the only feasible approach for calibrations below 100 Bq/m3. While calibration under exponentially decaying radon activity concentration remains useful for evaluating the devices’ linearity across a broad range, it proved unsuitable for calibration of user-grade monitors at low activities due to the high statistical variation in their signal.
The linearity of RadonEye +2 was demonstrated in the range 50 Bq/m3 - 300 Bq/m3 and they will be utilized for the sensor networks developed within the RadonNET project. Dynamic background correction, applicable to non-spectrometric detectors, was applied based on the monitor's exposure history. Furthermore, it was observed that the pulse-processing algorithm of RadonEyes +2 distorts the Poisson distribution of the signal, thereby increasing its variation. Potentially, lower measurement uncertainty could be achieved with electronic radon detectors that report the registered pulses and allow access to their processing algorithms.
期刊介绍:
Applied Radiation and Isotopes provides a high quality medium for the publication of substantial, original and scientific and technological papers on the development and peaceful application of nuclear, radiation and radionuclide techniques in chemistry, physics, biochemistry, biology, medicine, security, engineering and in the earth, planetary and environmental sciences, all including dosimetry. Nuclear techniques are defined in the broadest sense and both experimental and theoretical papers are welcome. They include the development and use of α- and β-particles, X-rays and γ-rays, neutrons and other nuclear particles and radiations from all sources, including radionuclides, synchrotron sources, cyclotrons and reactors and from the natural environment.
The journal aims to publish papers with significance to an international audience, containing substantial novelty and scientific impact. The Editors reserve the rights to reject, with or without external review, papers that do not meet these criteria.
Papers dealing with radiation processing, i.e., where radiation is used to bring about a biological, chemical or physical change in a material, should be directed to our sister journal Radiation Physics and Chemistry.