{"title":"[Feasibility study on the clinical translation of a remote jaundice monitoring system for home-based screening of neonatal hyperbilirubinemia].","authors":"Xiao-Fan Sun, Yi Zheng, Ai-Ling Su, Shu-Ping Han, Xiao-Yue Dong","doi":"10.7499/j.issn.1008-8830.2503110","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>To evaluate the clinical utility and translational potential of a remote jaundice monitoring system for home-based screening of neonatal hyperbilirubinemia.</p><p><strong>Methods: </strong>A prospective self-controlled study was conducted, enrolling 538 newborns with gestational age ≥35 weeks, birth weight ≥2 000 g, and postnatal age ≤14 days at the Women's Hospital of Nanjing Medical University from March to October 2023. Four screening protocols with different predictive indicators were developed based on the Chinese Neonatal Transcutaneous Hourly Bilirubin Nomogram. The effectiveness of the system was evaluated, and the feasibility of using the remote jaundice monitoring system in actual home settings was analyzed.</p><p><strong>Results: </strong>A total of 538 paired transcutaneous bilirubin (TcB) and total serum bilirubin (TSB) measurements showed a strong correlation (<i>r</i>=0.85, <i>P</i><0.001), with 95.0% (511/538) of samples within the 95% limits of agreement. Using TcB ≥ the 95th percentile as the predictive indicator, the system achieved 100% sensitivity, 46.2% specificity, and an area under the receiver operating characteristic curve of 0.731 (95%<i>CI</i>: 0.682-0.780). This approach could reduce unnecessary hospital visits by 41.4% (221/538).</p><p><strong>Conclusions: </strong>The system integrates the QBH-801 transcutaneous bilirubinometer, intelligent early warning, and remote guidance services, establishing a closed-loop \"hospital-to-home\" management model. It demonstrates high safety and feasibility, with significant clinical translational value.</p>","PeriodicalId":39792,"journal":{"name":"中国当代儿科杂志","volume":"27 9","pages":"1057-1061"},"PeriodicalIF":0.0000,"publicationDate":"2025-07-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12447935/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国当代儿科杂志","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.7499/j.issn.1008-8830.2503110","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: To evaluate the clinical utility and translational potential of a remote jaundice monitoring system for home-based screening of neonatal hyperbilirubinemia.
Methods: A prospective self-controlled study was conducted, enrolling 538 newborns with gestational age ≥35 weeks, birth weight ≥2 000 g, and postnatal age ≤14 days at the Women's Hospital of Nanjing Medical University from March to October 2023. Four screening protocols with different predictive indicators were developed based on the Chinese Neonatal Transcutaneous Hourly Bilirubin Nomogram. The effectiveness of the system was evaluated, and the feasibility of using the remote jaundice monitoring system in actual home settings was analyzed.
Results: A total of 538 paired transcutaneous bilirubin (TcB) and total serum bilirubin (TSB) measurements showed a strong correlation (r=0.85, P<0.001), with 95.0% (511/538) of samples within the 95% limits of agreement. Using TcB ≥ the 95th percentile as the predictive indicator, the system achieved 100% sensitivity, 46.2% specificity, and an area under the receiver operating characteristic curve of 0.731 (95%CI: 0.682-0.780). This approach could reduce unnecessary hospital visits by 41.4% (221/538).
Conclusions: The system integrates the QBH-801 transcutaneous bilirubinometer, intelligent early warning, and remote guidance services, establishing a closed-loop "hospital-to-home" management model. It demonstrates high safety and feasibility, with significant clinical translational value.
中国当代儿科杂志Medicine-Pediatrics, Perinatology and Child Health
CiteScore
1.50
自引率
0.00%
发文量
5006
期刊介绍:
The Chinese Journal of Contemporary Pediatrics (CJCP) is a peer-reviewed open access periodical in the field of pediatrics that is sponsored by the Central South University/Xiangya Hospital of Central South University and under the auspices of the Ministry of Education of China. It is cited as a source in the scientific and technological papers of Chinese journals, the Chinese Science Citation Database (CSCD), and is one of the core Chinese periodicals in the Peking University Library. CJCP has been indexed by MEDLINE/PubMed/PMC of the American National Library, American Chemical Abstracts (CA), Holland Medical Abstracts (EM), Western Pacific Region Index Medicus (WPRIM), Scopus and EBSCO. It is a monthly periodical published on the 15th of every month, and is distributed both at home and overseas. The Chinese series publication number is CN 43-1301/R;ISSN 1008-8830. The tenet of CJCP is to “reflect the latest advances and be open to the world”. The periodical reports the most recent advances in the contemporary pediatric field. The majority of the readership is pediatric doctors and researchers.