DurAVR - a novel biomimetic balloon-expandable transcatheter valve for TAVI, valve-in-valve TAVI and redo-TAVI.

IF 9.5 1区 医学 Q1 CARDIAC & CARDIOVASCULAR SYSTEMS
David Meier, Julien Delarive, Althea Lai, Rebecca T Hahn, João L Cavalcante, Anita W Asgar, Azeem Latib, Pankaj Garg, Susheel Kodali, Vinayak N Bapat, Magnus Settergren, Janarthanan Sathananthan, Christopher U Meduri, Stephanie L Sellers
{"title":"DurAVR - a novel biomimetic balloon-expandable transcatheter valve for TAVI, valve-in-valve TAVI and redo-TAVI.","authors":"David Meier, Julien Delarive, Althea Lai, Rebecca T Hahn, João L Cavalcante, Anita W Asgar, Azeem Latib, Pankaj Garg, Susheel Kodali, Vinayak N Bapat, Magnus Settergren, Janarthanan Sathananthan, Christopher U Meduri, Stephanie L Sellers","doi":"10.4244/EIJ-D-24-01000","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The DurAVR transcatheter heart valve (THV) is a novel biomimetic balloon-expandable valve with promising early clinical results.</p><p><strong>Aims: </strong>We aimed to assess the hydrodynamic performance of the DurAVR THV in native, valve-in-valve (ViV), and redo-transcatheter aortic valve implantation (TAVI) procedures against commercially available THVs on the bench.</p><p><strong>Methods: </strong>The hydrodynamic function of the DurAVR THV was assessed by simulating native valve deployments at 0 mm, 3 mm, and 6 mm depths, compared to SAPIEN 3 (S3), Evolut PRO, Navitor, and ACURATE neo2 (ACn2) valves. For ViV simulations, THVs were implanted in 21 mm and 23 mm Magna Ease, Mosaic, and Hancock bioprostheses. For redo-TAVI simulations, the DurAVR THV was assessed within S3, Evolut PRO, Navitor, and ACn2 valves.</p><p><strong>Results: </strong>For native TAVI simulations, the DurAVR THV demonstrated superior or comparable hydrodynamic performance, independent of implant depth, with an effective orifice area (EOA) ≥3 cm<sup>2</sup> and a mean gradient (MG) <6 mmHg. The DurAVR THV had nil to mild pinwheeling (0-2%) at all depths, while the S3 and Evolut PRO showed moderate pinwheeling at 6 mm depth. For ViV simulations, the DurAVR THV exhibited larger EOAs and lower MGs than the comparator THVs and showed no more than mild pinwheeling in all ViV configurations. For redo-TAVI simulations, the DurAVR THV exhibited larger EOAs and lower MGs in each simulation compared to all other THVs tested, with no more than mild pinwheeling observed in all configurations except when implanted within the Evolut PRO.</p><p><strong>Conclusions: </strong>In this bench study, the DurAVR THV demonstrated excellent hydrodynamic performance in native, ViV, and redo-TAVI simulations. Future large-scale studies are needed to confirm these findings in clinical application and further characterise the valve's short- and long-term performance.</p>","PeriodicalId":54378,"journal":{"name":"Eurointervention","volume":"21 18","pages":"e1090-e1101"},"PeriodicalIF":9.5000,"publicationDate":"2025-09-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12438964/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Eurointervention","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.4244/EIJ-D-24-01000","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The DurAVR transcatheter heart valve (THV) is a novel biomimetic balloon-expandable valve with promising early clinical results.

Aims: We aimed to assess the hydrodynamic performance of the DurAVR THV in native, valve-in-valve (ViV), and redo-transcatheter aortic valve implantation (TAVI) procedures against commercially available THVs on the bench.

Methods: The hydrodynamic function of the DurAVR THV was assessed by simulating native valve deployments at 0 mm, 3 mm, and 6 mm depths, compared to SAPIEN 3 (S3), Evolut PRO, Navitor, and ACURATE neo2 (ACn2) valves. For ViV simulations, THVs were implanted in 21 mm and 23 mm Magna Ease, Mosaic, and Hancock bioprostheses. For redo-TAVI simulations, the DurAVR THV was assessed within S3, Evolut PRO, Navitor, and ACn2 valves.

Results: For native TAVI simulations, the DurAVR THV demonstrated superior or comparable hydrodynamic performance, independent of implant depth, with an effective orifice area (EOA) ≥3 cm2 and a mean gradient (MG) <6 mmHg. The DurAVR THV had nil to mild pinwheeling (0-2%) at all depths, while the S3 and Evolut PRO showed moderate pinwheeling at 6 mm depth. For ViV simulations, the DurAVR THV exhibited larger EOAs and lower MGs than the comparator THVs and showed no more than mild pinwheeling in all ViV configurations. For redo-TAVI simulations, the DurAVR THV exhibited larger EOAs and lower MGs in each simulation compared to all other THVs tested, with no more than mild pinwheeling observed in all configurations except when implanted within the Evolut PRO.

Conclusions: In this bench study, the DurAVR THV demonstrated excellent hydrodynamic performance in native, ViV, and redo-TAVI simulations. Future large-scale studies are needed to confirm these findings in clinical application and further characterise the valve's short- and long-term performance.

DurAVR -一种新型仿生球囊可膨胀经导管瓣膜,用于TAVI、阀中TAVI和再TAVI。
背景:DurAVR经导管心脏瓣膜(THV)是一种新型的仿生球囊可膨胀瓣膜,具有良好的早期临床效果。目的:我们的目的是评估DurAVR THV在原生、瓣中瓣(ViV)和经导管主动脉瓣植入(TAVI)手术中的流体动力学性能,并与市售的THV进行比较。方法:与SAPIEN 3 (S3)、Evolut PRO、Navitor和ACn2 (ACn2)阀门相比,通过模拟0 mm、3 mm和6 mm深度的天然阀门部署来评估DurAVR THV的流体动力学功能。为了进行ViV模拟,将thv植入21 mm和23 mm的Magna Ease、Mosaic和Hancock生物假体中。对于redo-TAVI模拟,在S3、Evolut PRO、Navitor和ACn2阀中对DurAVR THV进行了评估。结果:在原生TAVI模拟中,DurAVR THV表现出优越或相当的水动力性能,与种植体深度无关,有效孔面积(EOA)≥3 cm2,平均梯度(MG)。结论:在本实验中,DurAVR THV在原生、ViV和redo-TAVI模拟中表现出优异的水动力性能。未来的大规模研究需要在临床应用中证实这些发现,并进一步表征瓣膜的短期和长期性能。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Eurointervention
Eurointervention CARDIAC & CARDIOVASCULAR SYSTEMS-
CiteScore
10.30
自引率
4.80%
发文量
380
审稿时长
3-8 weeks
期刊介绍: EuroIntervention Journal is an international, English language, peer-reviewed journal whose aim is to create a community of high quality research and education in the field of percutaneous and surgical cardiovascular interventions.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信