Ahmad Parvinian, Nathan R Huber, Rebecca Hibbert, Daniel A Adamo, Juna Musa, Andrea Ferrero, Brian T Welch, Christopher P Favazza
{"title":"Interventional High Monoenergetic Spectral CT for Percutaneous Thermal Ablation Procedures.","authors":"Ahmad Parvinian, Nathan R Huber, Rebecca Hibbert, Daniel A Adamo, Juna Musa, Andrea Ferrero, Brian T Welch, Christopher P Favazza","doi":"10.1097/RCT.0000000000001786","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>The purpose of this study was to evaluate the potential clinical value of spectral CT for image guidance during percutaneous thermal ablation, particularly investigating high monoenergetic (MonoE) for metal artifact reduction.</p><p><strong>Materials and methods: </strong>A phantom experiment was performed in which an ablation probe was scanned in \"continuous CT\" (CCT) mode on a spectral CT scanner. ROI measurements quantified metal artifacts for conventional and MonoE CCT images. Subsequently, a retrospective reader study was conducted with 10 patient cases. Three radiologists reviewed cases to select the optimal MonoE setting and perform a Likert score rating (+2: MonoE preferred, -2: conventional preferred).</p><p><strong>Results: </strong>Phantom measurements indicated a reduction in metal artifact for high MonoE (150 keV: -46 HU) relative to conventional CCT (-122 HU). The MonoE setting preference was 151±21 keV. The mean Likert score was 1.2±0.7. Radiologists indicated a perceived benefit of high MonoE CCT when probe placement was near a critical structure.</p><p><strong>Conclusions: </strong>This study indicates a clinical preference for high MonoE CCT compared with conventional CCT. Insofar as spectral CCT can be offered at a clinically feasible time frame, our results support the assessment of 150 keV image to reduce metal artifact while maintaining satisfactory soft tissue contrast.</p>","PeriodicalId":15402,"journal":{"name":"Journal of Computer Assisted Tomography","volume":" ","pages":""},"PeriodicalIF":1.3000,"publicationDate":"2025-09-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Computer Assisted Tomography","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/RCT.0000000000001786","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: The purpose of this study was to evaluate the potential clinical value of spectral CT for image guidance during percutaneous thermal ablation, particularly investigating high monoenergetic (MonoE) for metal artifact reduction.
Materials and methods: A phantom experiment was performed in which an ablation probe was scanned in "continuous CT" (CCT) mode on a spectral CT scanner. ROI measurements quantified metal artifacts for conventional and MonoE CCT images. Subsequently, a retrospective reader study was conducted with 10 patient cases. Three radiologists reviewed cases to select the optimal MonoE setting and perform a Likert score rating (+2: MonoE preferred, -2: conventional preferred).
Results: Phantom measurements indicated a reduction in metal artifact for high MonoE (150 keV: -46 HU) relative to conventional CCT (-122 HU). The MonoE setting preference was 151±21 keV. The mean Likert score was 1.2±0.7. Radiologists indicated a perceived benefit of high MonoE CCT when probe placement was near a critical structure.
Conclusions: This study indicates a clinical preference for high MonoE CCT compared with conventional CCT. Insofar as spectral CCT can be offered at a clinically feasible time frame, our results support the assessment of 150 keV image to reduce metal artifact while maintaining satisfactory soft tissue contrast.
期刊介绍:
The mission of Journal of Computer Assisted Tomography is to showcase the latest clinical and research developments in CT, MR, and closely related diagnostic techniques. We encourage submission of both original research and review articles that have immediate or promissory clinical applications. Topics of special interest include: 1) functional MR and CT of the brain and body; 2) advanced/innovative MRI techniques (diffusion, perfusion, rapid scanning); and 3) advanced/innovative CT techniques (perfusion, multi-energy, dose-reduction, and processing).