Yuval Avidan MD , Amir Aker MD , Razi Khoury MD , Sameha Zahra MD , Nissan Ben Dov MD , Jorge E. Schliamser MD , Asaf Danon MD, MSc
{"title":"From Wrist to Precision: Enhanced Atrial Flutter Detection with Modified Smartwatch Single-Lead Electrocardiogram Placement","authors":"Yuval Avidan MD , Amir Aker MD , Razi Khoury MD , Sameha Zahra MD , Nissan Ben Dov MD , Jorge E. Schliamser MD , Asaf Danon MD, MSc","doi":"10.1016/j.cjco.2025.05.015","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Smartwatches, such as the Apple Watch (AW), are well-established tools for detecting atrial fibrillation (AF). We hypothesize that atrial flutter (AFL) is frequently misdiagnosed using traditional single-lead electrocardiogram configurations and that modified device positioning could substantially improve diagnostic accuracy.</div></div><div><h3>Methods</h3><div>Standard smartwatch lead-I (AW-I) recordings were obtained from 75 patients, including 25 with AFL, 25 with AF, and 25 with sinus rhythm. Additionally, modified lead-II (AW-II) recordings were collected for all AFL cases, resulting in a total of 100 tracings. Twenty blinded physicians from 4 different specialties independently analyzed all recordings.</div></div><div><h3>Results</h3><div>Physicians’ ability to detect AFL using the AW-I lead was poor, with only 11.6% of cases correctly identified (<em>P</em> = 0.362). AFL was most often misdiagnosed as AF (55.6%), undetermined (21%), or sinus rhythm (11.8%). Diagnostic accuracy improved significantly with the AW-II lead, exceeding 80% among electrophysiologists, cardiologists, and primary care physicians (<em>P</em> = 0.001). Variable atrioventricular conduction was associated strongly with correct diagnosis exclusively through the AW-II lead (odds ratio 1.85, 95% confidence interval 1.14-3.0, <em>P</em> = 0.012).</div></div><div><h3>Conclusions</h3><div>The standard lead-I configuration used by smartwatches is prone to misclassifying AF as AFL, particularly in the setting of variable conduction. A simple modification to lead-II positioning significantly enhances diagnostic accuracy. This adjustment may be especially valuable during post–pulmonary vein isolation surveillance and in broader clinical scenarios in which precise rhythm identification can influence therapeutic decision-making.</div></div>","PeriodicalId":36924,"journal":{"name":"CJC Open","volume":"7 9","pages":"Pages 1149-1156"},"PeriodicalIF":2.5000,"publicationDate":"2025-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"CJC Open","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2589790X25003968","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Smartwatches, such as the Apple Watch (AW), are well-established tools for detecting atrial fibrillation (AF). We hypothesize that atrial flutter (AFL) is frequently misdiagnosed using traditional single-lead electrocardiogram configurations and that modified device positioning could substantially improve diagnostic accuracy.
Methods
Standard smartwatch lead-I (AW-I) recordings were obtained from 75 patients, including 25 with AFL, 25 with AF, and 25 with sinus rhythm. Additionally, modified lead-II (AW-II) recordings were collected for all AFL cases, resulting in a total of 100 tracings. Twenty blinded physicians from 4 different specialties independently analyzed all recordings.
Results
Physicians’ ability to detect AFL using the AW-I lead was poor, with only 11.6% of cases correctly identified (P = 0.362). AFL was most often misdiagnosed as AF (55.6%), undetermined (21%), or sinus rhythm (11.8%). Diagnostic accuracy improved significantly with the AW-II lead, exceeding 80% among electrophysiologists, cardiologists, and primary care physicians (P = 0.001). Variable atrioventricular conduction was associated strongly with correct diagnosis exclusively through the AW-II lead (odds ratio 1.85, 95% confidence interval 1.14-3.0, P = 0.012).
Conclusions
The standard lead-I configuration used by smartwatches is prone to misclassifying AF as AFL, particularly in the setting of variable conduction. A simple modification to lead-II positioning significantly enhances diagnostic accuracy. This adjustment may be especially valuable during post–pulmonary vein isolation surveillance and in broader clinical scenarios in which precise rhythm identification can influence therapeutic decision-making.