C Ubeda, E Vano, M D Perez, R Ramirez, P Miranda, P Azcurra, M Oliveira, H Schelin, A Yagui, H Khoury, V Asfora, M Freitas, R Medeiros, D Aguirre, N Riquelme, L Cardenas, W Mosquera, A M Aristizabal, F Arias, R Gutierrez, R De la Mora, T Rivera, M C Franco, C Britton, J Zapata, P Araujo, S Donoso
{"title":"Regional diagnostic reference levels in paediatric interventional cardiology in Latin America and the Caribbean countries.","authors":"C Ubeda, E Vano, M D Perez, R Ramirez, P Miranda, P Azcurra, M Oliveira, H Schelin, A Yagui, H Khoury, V Asfora, M Freitas, R Medeiros, D Aguirre, N Riquelme, L Cardenas, W Mosquera, A M Aristizabal, F Arias, R Gutierrez, R De la Mora, T Rivera, M C Franco, C Britton, J Zapata, P Araujo, S Donoso","doi":"10.1088/1361-6498/ae0650","DOIUrl":null,"url":null,"abstract":"<p><p>The aim of the present study was to present an updated set of regional diagnostic reference levels (DRLs) for paediatric interventional cardiology procedures in Latin America and the Caribbean countries using the quantity kerma area-product (<i>P</i><sub>ka</sub>) and<i>P</i><sub>ka</sub>per body weigth, classified by age and weight groups within the framework of OPRIPALC international programme. Data based on diagnostic and therapeutic procedures were collected between December 2020 and December 2023. Regional DRLs have been set as the 3rd quartile of<i>P</i><sub>ka</sub>collected in 23 hospitals from 11 countries in a final sample of 2906 procedures. DRLs have been set for four age groups and five weight groups. The proposed set of regional DRLs for different diagnostic and therapeutic procedures by age and weight groups for<i>P</i><sub>ka</sub>quantity, were 4.3-5.4 Gy cm<sup>2</sup>(<1 yr) to 20.6-34.8 Gy cm<sup>2</sup>(10 to <16 yrs) and 3.4-4.5 Gy cm<sup>2</sup>(<5 kg) to 26.9-63.9 Gy cm<sup>2</sup>(50-< 80 kg), respectively. The proposed set of regional DRLs for<i>P</i><sub>ka</sub>/weight values were 0.885-1.181 Gy cm<sup>2</sup>kg<sup>-1</sup>(<1 yr) to 0.510-0.838 Gy cm<sup>2</sup>kg<sup>-1</sup>(10 to <16 yrs) and 0.984-1.333 Gy cm<sup>2</sup>kg<sup>-1</sup>(<5 kg) to 0.448-1.196 Gy cm<sup>2</sup>kg<sup>-1</sup>(50-< 80 kg), respectively. The values proposed as DRLs in OPRIPALC can be adopted by other countries and centers performing paediatric interventional cardiology procedures to compare their own dose values. The OPRIPALC methodology for establishing DRLs, with its variations and limitations, can also be considered in other regions of the world to develop their own regional DRLs.</p>","PeriodicalId":50068,"journal":{"name":"Journal of Radiological Protection","volume":" ","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2025-09-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Radiological Protection","FirstCategoryId":"93","ListUrlMain":"https://doi.org/10.1088/1361-6498/ae0650","RegionNum":4,"RegionCategory":"环境科学与生态学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENVIRONMENTAL SCIENCES","Score":null,"Total":0}
引用次数: 0
Abstract
The aim of the present study was to present an updated set of regional diagnostic reference levels (DRLs) for paediatric interventional cardiology procedures in Latin America and the Caribbean countries using the quantity kerma area-product (Pka) andPkaper body weigth, classified by age and weight groups within the framework of OPRIPALC international programme. Data based on diagnostic and therapeutic procedures were collected between December 2020 and December 2023. Regional DRLs have been set as the 3rd quartile ofPkacollected in 23 hospitals from 11 countries in a final sample of 2906 procedures. DRLs have been set for four age groups and five weight groups. The proposed set of regional DRLs for different diagnostic and therapeutic procedures by age and weight groups forPkaquantity, were 4.3-5.4 Gy cm2(<1 yr) to 20.6-34.8 Gy cm2(10 to <16 yrs) and 3.4-4.5 Gy cm2(<5 kg) to 26.9-63.9 Gy cm2(50-< 80 kg), respectively. The proposed set of regional DRLs forPka/weight values were 0.885-1.181 Gy cm2kg-1(<1 yr) to 0.510-0.838 Gy cm2kg-1(10 to <16 yrs) and 0.984-1.333 Gy cm2kg-1(<5 kg) to 0.448-1.196 Gy cm2kg-1(50-< 80 kg), respectively. The values proposed as DRLs in OPRIPALC can be adopted by other countries and centers performing paediatric interventional cardiology procedures to compare their own dose values. The OPRIPALC methodology for establishing DRLs, with its variations and limitations, can also be considered in other regions of the world to develop their own regional DRLs.
期刊介绍:
Journal of Radiological Protection publishes articles on all aspects of radiological protection, including non-ionising as well as ionising radiations. Fields of interest range from research, development and theory to operational matters, education and training. The very wide spectrum of its topics includes: dosimetry, instrument development, specialized measuring techniques, epidemiology, biological effects (in vivo and in vitro) and risk and environmental impact assessments.
The journal encourages publication of data and code as well as results.