A Novel Technique to Extract Implanted Leads Using Simple Stylets and Reused Rotational Sheaths in Patients with a Cardiac Implantable Electronic Device.
Serkan Çay, Görkem Kuş, Meryem Kara, Elif Hande Özcan Çetin, Özcan Özeke, Idriz Merovci, Ahmet Korkmaz, Fırat Özcan, Serkan Topaloğlu
{"title":"A Novel Technique to Extract Implanted Leads Using Simple Stylets and Reused Rotational Sheaths in Patients with a Cardiac Implantable Electronic Device.","authors":"Serkan Çay, Görkem Kuş, Meryem Kara, Elif Hande Özcan Çetin, Özcan Özeke, Idriz Merovci, Ahmet Korkmaz, Fırat Özcan, Serkan Topaloğlu","doi":"10.5543/tkda.2025.64865","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Transvenous lead extraction (TLE) is used in various clinical scenarios, such as device-related infections. Mechanically powered sheaths are one of the most commonly used tools for TLE procedures. We evaluated the procedural and clinical outcomes of a novel extraction technique for chronically implanted leads in the treatment of device-related infections.</p><p><strong>Method: </strong>The novel extraction technique utilizing standard implantation stylets, snares, reused rotational sheaths, catheters, and wires was evaluated for procedural success and clinical outcomes.</p><p><strong>Results: </strong>A total of 12 consecutive patients with device-related infections underwent the novel TLE procedure. Complete procedural success was achieved in all patients, with a minor complication rate of 8% (one patient). No major complications or procedure-related mortality were observed. During a median follow-up period of 435 days, one patient died due to a multidrug-resistant systemic infection, one due to end-stage heart failure, and one underwent valve surgery for concomitant valve endocarditis. No cases of reinfection were reported in the study population. Additionally, this novel technique was approximately 85% less costly than the conventional standard technique using locking stylets and unused rotational sheaths.</p><p><strong>Conclusion: </strong>In situations where unused extraction tools are unavailable or limited by reimbursement constraints, this novel TLE technique offers an effective and safe alternative.</p>","PeriodicalId":94261,"journal":{"name":"Turk Kardiyoloji Dernegi arsivi : Turk Kardiyoloji Derneginin yayin organidir","volume":" ","pages":""},"PeriodicalIF":0.6000,"publicationDate":"2025-09-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Turk Kardiyoloji Dernegi arsivi : Turk Kardiyoloji Derneginin yayin organidir","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5543/tkda.2025.64865","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: Transvenous lead extraction (TLE) is used in various clinical scenarios, such as device-related infections. Mechanically powered sheaths are one of the most commonly used tools for TLE procedures. We evaluated the procedural and clinical outcomes of a novel extraction technique for chronically implanted leads in the treatment of device-related infections.
Method: The novel extraction technique utilizing standard implantation stylets, snares, reused rotational sheaths, catheters, and wires was evaluated for procedural success and clinical outcomes.
Results: A total of 12 consecutive patients with device-related infections underwent the novel TLE procedure. Complete procedural success was achieved in all patients, with a minor complication rate of 8% (one patient). No major complications or procedure-related mortality were observed. During a median follow-up period of 435 days, one patient died due to a multidrug-resistant systemic infection, one due to end-stage heart failure, and one underwent valve surgery for concomitant valve endocarditis. No cases of reinfection were reported in the study population. Additionally, this novel technique was approximately 85% less costly than the conventional standard technique using locking stylets and unused rotational sheaths.
Conclusion: In situations where unused extraction tools are unavailable or limited by reimbursement constraints, this novel TLE technique offers an effective and safe alternative.