Equipment Failure during Urological Surgery: Implications and Lessons to be Learned.

IF 0.9 4区 医学 Q4 UROLOGY & NEPHROLOGY
Evangelos N Symeonidis, Asterios Symeonidis, Aris Kaltsas, Georgios Tsampoukas, Bartosz Dybowski, Mauro Ragonese, Francesco Greco, Uros Bumbasirevic, Athanasios Bekos, Andreas Andreou
{"title":"Equipment Failure during Urological Surgery: Implications and Lessons to be Learned.","authors":"Evangelos N Symeonidis, Asterios Symeonidis, Aris Kaltsas, Georgios Tsampoukas, Bartosz Dybowski, Mauro Ragonese, Francesco Greco, Uros Bumbasirevic, Athanasios Bekos, Andreas Andreou","doi":"10.56434/j.arch.esp.urol.20257807.121","DOIUrl":null,"url":null,"abstract":"<p><p>The literature on the exact incidence of equipment failure during urological surgery is rather heterogeneous. Although failure rates are unacceptably high in other surgical disciplines, more compelling evidence is needed in urology. The present study provides case examples to illustrate several instances of urological instrument malfunction encountered in daily surgical practice, from the field of endourology to the newer robotic systems. Five cases of endourological device failures involving a resectoscope, a semi-rigid ureteroscope and alligator forceps and one case involving monopolar scissors of the Hugo RAS robotic system are presented, without any resultant complications to patients. Urologists and surgical personnel need to become acquainted with various equipment utilised in their field. Initiatives aimed at improvement of handling, inspection and prompt replacement of defective devices, rigorous adherence to manufacturers' guidelines, implementation of evaluative checklists for quality control and the adoption of incident-specific reporting forms are commendable actions toward enhancing operational efficiency. In the context of robotic surgery, a team of technical experts possessing sufficient knowledge of the equipment should offer continuous assistance in accordance with troubleshooting protocols. By implementing a standardised problem-solving strategy, they could diagnose and rectify emerging errors promptly and efficiently. Overall, failure in the theatre demands collective understanding and prompt management. Urologists need to communicate such events and collaborate to improve overall surgical efficiency.</p>","PeriodicalId":48852,"journal":{"name":"Archivos Espanoles De Urologia","volume":"78 7","pages":"917-923"},"PeriodicalIF":0.9000,"publicationDate":"2025-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Archivos Espanoles De Urologia","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.56434/j.arch.esp.urol.20257807.121","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"UROLOGY & NEPHROLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

The literature on the exact incidence of equipment failure during urological surgery is rather heterogeneous. Although failure rates are unacceptably high in other surgical disciplines, more compelling evidence is needed in urology. The present study provides case examples to illustrate several instances of urological instrument malfunction encountered in daily surgical practice, from the field of endourology to the newer robotic systems. Five cases of endourological device failures involving a resectoscope, a semi-rigid ureteroscope and alligator forceps and one case involving monopolar scissors of the Hugo RAS robotic system are presented, without any resultant complications to patients. Urologists and surgical personnel need to become acquainted with various equipment utilised in their field. Initiatives aimed at improvement of handling, inspection and prompt replacement of defective devices, rigorous adherence to manufacturers' guidelines, implementation of evaluative checklists for quality control and the adoption of incident-specific reporting forms are commendable actions toward enhancing operational efficiency. In the context of robotic surgery, a team of technical experts possessing sufficient knowledge of the equipment should offer continuous assistance in accordance with troubleshooting protocols. By implementing a standardised problem-solving strategy, they could diagnose and rectify emerging errors promptly and efficiently. Overall, failure in the theatre demands collective understanding and prompt management. Urologists need to communicate such events and collaborate to improve overall surgical efficiency.

泌尿外科手术中的设备故障:启示和教训。
关于泌尿外科手术中设备故障的确切发生率的文献是相当不一致的。尽管失败率在其他外科学科中高得令人无法接受,但在泌尿外科中需要更多令人信服的证据。本研究提供了一些案例来说明在日常手术实践中遇到的泌尿外科仪器故障的几个实例,从泌尿外科领域到较新的机器人系统。本文报道了5例输尿管镜、半刚性输尿管镜、鳄鱼钳等输尿管镜手术失败的病例,以及1例使用Hugo RAS机器人系统的单极子剪刀手术失败的病例,均未发生并发症。泌尿科医生和外科人员需要熟悉在他们的领域使用的各种设备。改善处理、检查和迅速更换有缺陷设备的措施、严格遵守制造商的指引、实施质量控制的评估清单,以及采用针对事故的报告表格,这些措施都是提高运作效率的值得赞扬的行动。在机器人手术的背景下,一个拥有足够设备知识的技术专家团队应该根据故障排除协议提供持续的帮助。通过实施标准化的问题解决策略,他们可以迅速有效地诊断和纠正新出现的错误。总的来说,剧院的失败需要集体的理解和及时的管理。泌尿科医生需要对这些事件进行沟通和合作,以提高整体手术效率。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Archivos Espanoles De Urologia
Archivos Espanoles De Urologia UROLOGY & NEPHROLOGY-
CiteScore
0.90
自引率
0.00%
发文量
111
期刊介绍: Archivos Españoles de Urología published since 1944, is an international peer review, susbscription Journal on Urology with original and review articles on different subjets in Urology: oncology, endourology, laparoscopic, andrology, lithiasis, pediatrics , urodynamics,... Case Report are also admitted.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信