Extended m/z range of coaxial linear ion trap/time-of-flight mass spectrometry using dynamic electric field ejection

IF 1.8 3区 化学 Q3 PHYSICS, ATOMIC, MOLECULAR & CHEMICAL
Yixue Cao , Xuesong Zhang , Jilong Wang , Chenxin Wu , Lei Hua , Ping Chen , Haiyang Li
{"title":"Extended m/z range of coaxial linear ion trap/time-of-flight mass spectrometry using dynamic electric field ejection","authors":"Yixue Cao ,&nbsp;Xuesong Zhang ,&nbsp;Jilong Wang ,&nbsp;Chenxin Wu ,&nbsp;Lei Hua ,&nbsp;Ping Chen ,&nbsp;Haiyang Li","doi":"10.1016/j.ijms.2025.117517","DOIUrl":null,"url":null,"abstract":"<div><div>Linear ion trap/time-of-flight mass spectrometry (LIT/TOFMS) to perform ion accumulation in a separate LIT prior to mass analysis is a more generally accepted method to improve duty cycle. However, a long distance between the accumulation region and the TOF pusher led to time-of-flight dispersion of the injected ions with different <em>m/z</em>, thereby limiting the detectable mass range in a single scan. This paper studies the characteristic behaviors of ion ejection from a segmented LIT through SIMION simulation. A novel dynamic electric field ejection strategy is proposed to broaden the ejection mass range of LIT. By linearly scanning the voltages on the segmented LIT and the ion gate, the axial potential distribution is dynamically adjusted to gradually shift the ion accumulation region toward the TOF pusher. As a result, the ejection distance between the accumulation region and the TOF pusher was slowly shortened, reducing time dispersion among different <em>m/z</em> ions. The simulation results demonstrate that this method significantly improves mass range of the coaxial dual LIT/TOFMS, achieving near 100 % injection efficiency for <em>m/z</em> 130–2000 with a mass window ratio <em>(m/z)</em><sub><em>max</em></sub><em>/(m/z)</em><sub><em>min</em></sub> of 15. This technology provides a solution for broadening the mass range of LIT-based hybrid instruments.</div></div>","PeriodicalId":338,"journal":{"name":"International Journal of Mass Spectrometry","volume":"518 ","pages":"Article 117517"},"PeriodicalIF":1.8000,"publicationDate":"2025-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Mass Spectrometry","FirstCategoryId":"92","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1387380625001216","RegionNum":3,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/8/30 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"PHYSICS, ATOMIC, MOLECULAR & CHEMICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Linear ion trap/time-of-flight mass spectrometry (LIT/TOFMS) to perform ion accumulation in a separate LIT prior to mass analysis is a more generally accepted method to improve duty cycle. However, a long distance between the accumulation region and the TOF pusher led to time-of-flight dispersion of the injected ions with different m/z, thereby limiting the detectable mass range in a single scan. This paper studies the characteristic behaviors of ion ejection from a segmented LIT through SIMION simulation. A novel dynamic electric field ejection strategy is proposed to broaden the ejection mass range of LIT. By linearly scanning the voltages on the segmented LIT and the ion gate, the axial potential distribution is dynamically adjusted to gradually shift the ion accumulation region toward the TOF pusher. As a result, the ejection distance between the accumulation region and the TOF pusher was slowly shortened, reducing time dispersion among different m/z ions. The simulation results demonstrate that this method significantly improves mass range of the coaxial dual LIT/TOFMS, achieving near 100 % injection efficiency for m/z 130–2000 with a mass window ratio (m/z)max/(m/z)min of 15. This technology provides a solution for broadening the mass range of LIT-based hybrid instruments.

Abstract Image

使用动态电场喷射的同轴线性离子阱/飞行时间质谱扩展m/z范围
线性离子阱/飞行时间质谱(LIT/TOFMS)在质量分析之前在单独的LIT中进行离子积累是一种更普遍接受的改善占空比的方法。然而,积累区和TOF推子之间的距离较长,导致注入的离子具有不同的m/z的飞行时间色散,从而限制了单次扫描可检测的质量范围。本文通过SIMION仿真研究了分段式LIT中离子喷射的特征行为。提出了一种新的动态电场抛射策略,以扩大抛射质量范围。通过线性扫描分段式LIT和离子栅上的电压,动态调整轴向电位分布,使离子积累区逐渐向TOF推子移动。累积区与TOF推子之间的弹射距离逐渐缩短,减少了不同m/z离子之间的时间色散。仿真结果表明,该方法显著提高了同轴双LIT/TOFMS的质量范围,在质量窗比(m/z)max/(m/z)min为15的情况下,m/z 130 ~ 2000的喷射效率接近100%。该技术为扩大基于lit的混合仪器的质量范围提供了一种解决方案。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.60
自引率
5.60%
发文量
145
审稿时长
71 days
期刊介绍: The journal invites papers that advance the field of mass spectrometry by exploring fundamental aspects of ion processes using both the experimental and theoretical approaches, developing new instrumentation and experimental strategies for chemical analysis using mass spectrometry, developing new computational strategies for data interpretation and integration, reporting new applications of mass spectrometry and hyphenated techniques in biology, chemistry, geology, and physics. Papers, in which standard mass spectrometry techniques are used for analysis will not be considered. IJMS publishes full-length articles, short communications, reviews, and feature articles including young scientist features.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书