Use of NaCl 9‰ as a cheap diluent for cortisol assay in lateralized samples from adrenal veins catheterisation.

IF 0.4
Aurélie Sieuw, Julie Brossaud, Jean-Benoît Corcuff, Cindy Lauro, Agnès Georges
{"title":"Use of NaCl 9‰ as a cheap diluent for cortisol assay in lateralized samples from adrenal veins catheterisation.","authors":"Aurélie Sieuw, Julie Brossaud, Jean-Benoît Corcuff, Cindy Lauro, Agnès Georges","doi":"10.1684/abc.2025.1993","DOIUrl":null,"url":null,"abstract":"<p><p>Among the tools necessary for the diagnosis of primary hyperaldosteronism, catheterization of the adrenal veins represents the major examination to determine the presence of lateralized aldosterone secretion. To ensure that the catheter is well positioned at the level of the adrenal veins, a cortisol dosage is carried out in parallel with the aldosterone dosage. At the Hormonology and Tumor Markers laboratory of the Bordeaux University Hospital, this cortisol assay is carried out on an Abbott Architect i2000, making it possible to extend the calibration range up to > 3,300 nmol/L. Beyond that, the supplier recommends carrying out a manual dilution using calibrator A in which the cortisol concentration is equal to 0 nmol/L. The downside is that this calibrator cannot be supplied alone. It is only available in a common box with the 5 other calibrators necessary to carry out the cortisol calibration range. To overcome this expensive strategy, we studied the use of another diluent: 0.9% NaCl. Samples from 11 CVS were diluted with 0.9% NaCl and a comparison of cortisol results was performed (n = 128). Passing-Bablok regression of cortisol concentrations did not show significant deviation from linearity. The interpretation of CVS selectivity was not impacted by the change of diluent nor the interpretation of secretion lateralization. This change of diluent therefore did not modify the appropriate medical decision: medicinal treatment in the case of the identification of a bilateral secretion or proposal for surgical intervention in the case of the identification of a lateralized secretion.</p>","PeriodicalId":93870,"journal":{"name":"Annales de biologie clinique","volume":"83 5","pages":"0"},"PeriodicalIF":0.4000,"publicationDate":"2025-08-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annales de biologie clinique","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1684/abc.2025.1993","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Among the tools necessary for the diagnosis of primary hyperaldosteronism, catheterization of the adrenal veins represents the major examination to determine the presence of lateralized aldosterone secretion. To ensure that the catheter is well positioned at the level of the adrenal veins, a cortisol dosage is carried out in parallel with the aldosterone dosage. At the Hormonology and Tumor Markers laboratory of the Bordeaux University Hospital, this cortisol assay is carried out on an Abbott Architect i2000, making it possible to extend the calibration range up to > 3,300 nmol/L. Beyond that, the supplier recommends carrying out a manual dilution using calibrator A in which the cortisol concentration is equal to 0 nmol/L. The downside is that this calibrator cannot be supplied alone. It is only available in a common box with the 5 other calibrators necessary to carry out the cortisol calibration range. To overcome this expensive strategy, we studied the use of another diluent: 0.9% NaCl. Samples from 11 CVS were diluted with 0.9% NaCl and a comparison of cortisol results was performed (n = 128). Passing-Bablok regression of cortisol concentrations did not show significant deviation from linearity. The interpretation of CVS selectivity was not impacted by the change of diluent nor the interpretation of secretion lateralization. This change of diluent therefore did not modify the appropriate medical decision: medicinal treatment in the case of the identification of a bilateral secretion or proposal for surgical intervention in the case of the identification of a lateralized secretion.

用氯化钠9‰作为廉价稀释剂对肾上腺静脉置管的侧化样品进行皮质醇测定。
在诊断原发性高醛固酮症的必要工具中,肾上腺静脉插管是确定侧化醛固酮分泌存在的主要检查。为了确保导管在肾上腺静脉的水平位置良好,皮质醇剂量与醛固酮剂量并行进行。在波尔多大学医院的激素和肿瘤标志物实验室,这项皮质醇测定是在雅培建筑师i2000上进行的,这使得将校准范围扩展到> 3300 nmol/L成为可能。除此之外,供应商建议使用校准器a进行手动稀释,其中皮质醇浓度等于0 nmol/L。缺点是这个校准器不能单独提供。它只与执行皮质醇校准范围所需的其他5个校准器放在一个普通盒中。为了克服这种昂贵的策略,我们研究了使用另一种稀释剂:0.9% NaCl。11份CVS样品用0.9% NaCl稀释,比较皮质醇结果(n = 128)。皮质醇浓度的pass - bablok回归没有显示出明显的线性偏差。CVS选择性的解释不受稀释剂和分泌物侧化的影响。因此,这种稀释剂的改变并没有改变适当的医疗决定:在发现双侧分泌物的情况下进行药物治疗,或者在发现一侧分泌物的情况下建议进行手术干预。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信