Electron Microscopic Irregularities Resulting from Post-manufacture Modifications to Intraocular Lenses.

IF 3.2 3区 医学 Q2 OPHTHALMOLOGY
Adam R Leone, Megan M Tuohy, Michael E Snyder
{"title":"Electron Microscopic Irregularities Resulting from Post-manufacture Modifications to Intraocular Lenses.","authors":"Adam R Leone, Megan M Tuohy, Michael E Snyder","doi":"10.1097/j.jcrs.0000000000001762","DOIUrl":null,"url":null,"abstract":"<p><p>Intraocular lenses undergo rigorous safety testing before market approval, though are sometimes modified by surgeons post-manufacture in practice. This includes techniques such as haptic curing, IOL \"punch,\" and trans-optic suture passage. We employed reported techniques to modify IOLs and imaged them with a scanning electron microscope. We found that the cured haptic was asymmetric and not as wide as expected for the degree of melt, thus not as robust to prevent possible haptic intrusion. The IOL \"punch\" created sharp edges, a radial matrix crack, irregular debris, and sharp shards of acrylic emanating from the edges. The mattress suture caused disruption at suture entry and exit points as well as surface undulation overlying the intra-optic suture pass. Post-manufacture modification of IOLs changes morphology and surface topography, which may affect IOL position, tolerance, optical function, and/or inflammatory potential. Surgeons should be aware of these possible consequences when considering post-manufacture modification of IOLs.</p>","PeriodicalId":15214,"journal":{"name":"Journal of cataract and refractive surgery","volume":" ","pages":""},"PeriodicalIF":3.2000,"publicationDate":"2025-08-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of cataract and refractive surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/j.jcrs.0000000000001762","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Intraocular lenses undergo rigorous safety testing before market approval, though are sometimes modified by surgeons post-manufacture in practice. This includes techniques such as haptic curing, IOL "punch," and trans-optic suture passage. We employed reported techniques to modify IOLs and imaged them with a scanning electron microscope. We found that the cured haptic was asymmetric and not as wide as expected for the degree of melt, thus not as robust to prevent possible haptic intrusion. The IOL "punch" created sharp edges, a radial matrix crack, irregular debris, and sharp shards of acrylic emanating from the edges. The mattress suture caused disruption at suture entry and exit points as well as surface undulation overlying the intra-optic suture pass. Post-manufacture modification of IOLs changes morphology and surface topography, which may affect IOL position, tolerance, optical function, and/or inflammatory potential. Surgeons should be aware of these possible consequences when considering post-manufacture modification of IOLs.

人工晶状体制造后修饰导致的电子显微镜不规则性。
人工晶状体在上市前要经过严格的安全测试,尽管有时外科医生会在生产后对其进行修改。这包括触觉固化、人工晶状体“打孔”和跨视界缝合通道等技术。我们采用已有报道的技术对iol进行修饰,并用扫描电子显微镜对其进行成像。我们发现,固化的触觉是不对称的,并且没有预期的融化程度那么宽,因此不那么坚固,以防止可能的触觉侵入。人工晶状体的“打孔”产生了尖锐的边缘、放射状基质裂缝、不规则碎片和从边缘发出的尖锐丙烯酸碎片。床垫缝线在缝线入口和出口点造成破坏,以及覆盖在视内缝线通道上的表面波动。人工晶状体制造后的修饰会改变晶状体的形态和表面形貌,从而影响晶状体的位置、耐受性、光学功能和/或炎症潜能。外科医生在考虑人工晶状体的术后修复时应该意识到这些可能的后果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.60
自引率
14.30%
发文量
259
审稿时长
8.5 weeks
期刊介绍: The Journal of Cataract & Refractive Surgery (JCRS), a preeminent peer-reviewed monthly ophthalmology publication, is the official journal of the American Society of Cataract and Refractive Surgery (ASCRS) and the European Society of Cataract and Refractive Surgeons (ESCRS). JCRS publishes high quality articles on all aspects of anterior segment surgery. In addition to original clinical studies, the journal features a consultation section, practical techniques, important cases, and reviews as well as basic science articles.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信