COR-KNOT-induced aortic root injury during minimally invasive aortic valve replacement of a bicuspid aortic valve: a case report.

IF 0.1
Takahiro Ishigaki, Kazuma Okamoto, Satoshi Asada, Genichi Sakaguchi
{"title":"COR-KNOT-induced aortic root injury during minimally invasive aortic valve replacement of a bicuspid aortic valve: a case report.","authors":"Takahiro Ishigaki, Kazuma Okamoto, Satoshi Asada, Genichi Sakaguchi","doi":"10.1186/s44215-025-00217-2","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The COR-KNOT device is widely used in minimally invasive cardiac surgery, and its efficacy in reducing operative time is well established. However, it has been associated with complications, such as valve leaflet perforation and prosthetic valve damage.</p><p><strong>Case presentation: </strong>We report a rare case of aortic root injury caused by COR-KNOT usage. A 72-year-old man with severe aortic stenosis due to a bicuspid aortic valve and distorted aortic root anatomy underwent minimally invasive aortic valve replacement using the COR-KNOT. After aortotomy closure, vigorous bleeding from the aortic root was observed. Punctate injuries of the aortic root, corresponding to the COR-KNOT clip sites, were identified following re-aortic cross-clamping. The procedure was converted to median sternotomy, the aortic root injury was repaired with an autologous pericardial patch, and the bioprosthetic valve was re-implanted using hand-tied sutures. The patient recovered uneventfully.</p><p><strong>Conclusions: </strong>This case highlights the importance of careful device selection and clip positioning in anatomically challenging cases to avoid life-threatening complications.</p>","PeriodicalId":520286,"journal":{"name":"General Thoracic and Cardiovascular Surgery Cases","volume":"4 1","pages":"33"},"PeriodicalIF":0.1000,"publicationDate":"2025-07-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC12291292/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"General Thoracic and Cardiovascular Surgery Cases","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s44215-025-00217-2","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The COR-KNOT device is widely used in minimally invasive cardiac surgery, and its efficacy in reducing operative time is well established. However, it has been associated with complications, such as valve leaflet perforation and prosthetic valve damage.

Case presentation: We report a rare case of aortic root injury caused by COR-KNOT usage. A 72-year-old man with severe aortic stenosis due to a bicuspid aortic valve and distorted aortic root anatomy underwent minimally invasive aortic valve replacement using the COR-KNOT. After aortotomy closure, vigorous bleeding from the aortic root was observed. Punctate injuries of the aortic root, corresponding to the COR-KNOT clip sites, were identified following re-aortic cross-clamping. The procedure was converted to median sternotomy, the aortic root injury was repaired with an autologous pericardial patch, and the bioprosthetic valve was re-implanted using hand-tied sutures. The patient recovered uneventfully.

Conclusions: This case highlights the importance of careful device selection and clip positioning in anatomically challenging cases to avoid life-threatening complications.

微创二尖瓣主动脉瓣置换术中cor - knot诱导的主动脉根部损伤1例报告。
背景:COR-KNOT装置广泛应用于微创心脏手术,其缩短手术时间的效果已得到证实。然而,它与并发症有关,如瓣膜小叶穿孔和人工瓣膜损伤。病例介绍:我们报告一例罕见的使用co - knot导致主动脉根部损伤的病例。一名72岁的男性,因主动脉瓣双尖瓣狭窄和主动脉根部解剖扭曲而严重主动脉瓣狭窄,采用COR-KNOT微创主动脉瓣置换术。主动脉切开术关闭后,观察到主动脉根部大量出血。主动脉根部的点状损伤,与COR-KNOT夹夹部位相对应,在再次主动脉交叉夹夹后被发现。手术转为胸骨正中切口,主动脉根部损伤用自体心包补片修复,生物人工瓣膜用手扎缝合线重新植入。病人平静地康复了。结论:本病例强调了在解剖困难的病例中谨慎选择装置和夹定位的重要性,以避免危及生命的并发症。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信