E B Podgorsak, A Olivier, M Pla, J Hazel, A de Lotbinière, B Pike
{"title":"Physical aspects of dynamic stereotactic radiosurgery.","authors":"E B Podgorsak, A Olivier, M Pla, J Hazel, A de Lotbinière, B Pike","doi":"10.1159/000100723","DOIUrl":null,"url":null,"abstract":"<p><p>Dynamic stereotactic radiosurgery is a radiosurgical technique based on a medium-energy isocentric linear accelerator and a stereotactic frame. The technique uses concurrent and continuous rotations of both the gantry (300 degrees, from 30 to 330 degrees) and the couch (150 degrees, from 75 to -75 degrees). It gives a uniform dose (+/- 5%) within the target volume and dose fall-offs outside the target volume comparable to those obtained from presently known radiosurgical techniques.</p>","PeriodicalId":75525,"journal":{"name":"Applied neurophysiology","volume":"50 1-6","pages":"263-8"},"PeriodicalIF":0.0000,"publicationDate":"1987-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1159/000100723","citationCount":"57","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Applied neurophysiology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1159/000100723","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 57
Abstract
Dynamic stereotactic radiosurgery is a radiosurgical technique based on a medium-energy isocentric linear accelerator and a stereotactic frame. The technique uses concurrent and continuous rotations of both the gantry (300 degrees, from 30 to 330 degrees) and the couch (150 degrees, from 75 to -75 degrees). It gives a uniform dose (+/- 5%) within the target volume and dose fall-offs outside the target volume comparable to those obtained from presently known radiosurgical techniques.