Impact of dry heat sterilization on the mechanical and functional performance of 3D printed medical devices for image-guided intervention

Q3 Medicine
Mary Chase Sheehan , Sarah Kim , Sunandita Sarker , Govindarajan Srimathveeravalli
{"title":"Impact of dry heat sterilization on the mechanical and functional performance of 3D printed medical devices for image-guided intervention","authors":"Mary Chase Sheehan ,&nbsp;Sarah Kim ,&nbsp;Sunandita Sarker ,&nbsp;Govindarajan Srimathveeravalli","doi":"10.1016/j.stlm.2025.100216","DOIUrl":null,"url":null,"abstract":"<div><div>Polymer additive manufacturing (AM) is a powerful method for medical device prototyping, producing low-cost medical devices for resource limited settings and patient-specific customization. Plastic medical devices created with AM must undergo sterilization prior to in vivo experiments or use with patients. While prior studies have verified the feasibility and safety of sterilization of such devices, the impact on mechanical performance has not been studied. Temperatures during autoclave sterilization, a commonly used and widely available method, can match or exceed the melt temperature of Nylon-12 used for selective laser sintering. Here we tested the impact of single or multiple cycles of autoclave sterilization on the mechanical and functional performance of plastic components used with interventional radiology related medical devices (catheter hub, ablation probe handle, and implantable port). We found that up to 2 cycles of sterilization did not tangibly impact the manufacturability, function or surface finish of these devices. However, we found that more than one cycle of sterilization can compromise the mechanical strength of the devices, with geometric-linked variations in the level of change in stiffness (ranging from 12 – 23 %). In conclusion, autoclave sterilization is safe for single use AM medical devices, where repetition of the sterilization for device reuse can compromise mechanical performance.</div></div>","PeriodicalId":72210,"journal":{"name":"Annals of 3D printed medicine","volume":"19 ","pages":"Article 100216"},"PeriodicalIF":0.0000,"publicationDate":"2025-07-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of 3D printed medicine","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2666964125000311","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Polymer additive manufacturing (AM) is a powerful method for medical device prototyping, producing low-cost medical devices for resource limited settings and patient-specific customization. Plastic medical devices created with AM must undergo sterilization prior to in vivo experiments or use with patients. While prior studies have verified the feasibility and safety of sterilization of such devices, the impact on mechanical performance has not been studied. Temperatures during autoclave sterilization, a commonly used and widely available method, can match or exceed the melt temperature of Nylon-12 used for selective laser sintering. Here we tested the impact of single or multiple cycles of autoclave sterilization on the mechanical and functional performance of plastic components used with interventional radiology related medical devices (catheter hub, ablation probe handle, and implantable port). We found that up to 2 cycles of sterilization did not tangibly impact the manufacturability, function or surface finish of these devices. However, we found that more than one cycle of sterilization can compromise the mechanical strength of the devices, with geometric-linked variations in the level of change in stiffness (ranging from 12 – 23 %). In conclusion, autoclave sterilization is safe for single use AM medical devices, where repetition of the sterilization for device reuse can compromise mechanical performance.
干热灭菌对用于图像引导干预的3D打印医疗设备机械和功能性能的影响
聚合物增材制造(AM)是一种强大的医疗设备原型制造方法,可以为资源有限的环境和特定患者定制生产低成本的医疗设备。用AM制造的塑料医疗器械必须在体内实验或与患者一起使用之前进行灭菌。虽然先前的研究已经验证了此类装置灭菌的可行性和安全性,但尚未研究其对机械性能的影响。高压灭菌器灭菌是一种常用和广泛可用的方法,其温度可以匹配或超过用于选择性激光烧结的尼龙-12的熔体温度。在这里,我们测试了单次或多次高压灭菌器灭菌对用于介入放射学相关医疗设备(导管毂、消融探头手柄和植入端口)的塑料部件的机械和功能性能的影响。我们发现长达2次的灭菌对这些设备的可制造性、功能或表面光洁度没有明显的影响。然而,我们发现,超过一个灭菌周期会损害设备的机械强度,在刚度变化水平的几何相关变化(范围从12 - 23%)。总之,高压灭菌器灭菌对于一次性使用的AM医疗设备是安全的,其中重复灭菌以重复使用设备可能会损害机械性能。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Annals of 3D printed medicine
Annals of 3D printed medicine Medicine and Dentistry (General), Materials Science (General)
CiteScore
4.70
自引率
0.00%
发文量
0
审稿时长
131 days
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信