Ion-Selective Optical Sensors with Internal Reference: on the Quest for Prolonged Functionality and Calibration-Free Measurements

IF 8 1区 化学 Q1 CHEMISTRY, ANALYTICAL
Nikolai Yu. Tiuftiakov, Andrey V. Kalinichev, Ivan P. Gryazev, Maria A. Peshkova
{"title":"Ion-Selective Optical Sensors with Internal Reference: on the Quest for Prolonged Functionality and Calibration-Free Measurements","authors":"Nikolai Yu. Tiuftiakov, Andrey V. Kalinichev, Ivan P. Gryazev, Maria A. Peshkova","doi":"10.1016/j.snb.2025.138231","DOIUrl":null,"url":null,"abstract":"As with most chemical sensors, the proliferation of optodes in analytical routine is hampered by the need for periodic sensor calibration. This fundamental limitation arises from a number of chemical and technological factors (e.g., instability of the illumination source, changes in the composition of the sensor phase, etc.), leading to deviation of the optodes signal in parallel measurements and over time. The development of new approaches for calibration-free determination of analytes with optodes remains a pertinent challenge in the field of chemical sensors. The presented study is aimed at the development and validation of ion-selective optical sensor arrays that do not require calibration. These arrays incorporate an internal optical signal reference composed of polymeric optodes with suppressed exchange capacity. The applicability of such optodes as optical signal standards is demonstrated in a broad range of pH and background electrolyte concentrations: from pH 5 to 9 in solutions of cations and lipophilic anions with concentrations reaching 0.1<!-- --> <!-- -->M. This experimental observation is supported by theoretical analyses performed by numerical simulation. The study identifies the effect of coextraction and ion exchange on the range of the optical signal stability. Furthermore, it is shown both theoretically and experimentally that by modulating the acidity of the lipophilic indicator in the standard optode, the operational range of the internal reference can be adjusted to the desired pH or electrolyte concentrations. In addition, the validation results demonstrate the feasibility of determining Na<sup>+</sup> concentration in the sample using internal standards with fair accuracy.","PeriodicalId":425,"journal":{"name":"Sensors and Actuators B: Chemical","volume":"76 1","pages":""},"PeriodicalIF":8.0000,"publicationDate":"2025-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Sensors and Actuators B: Chemical","FirstCategoryId":"92","ListUrlMain":"https://doi.org/10.1016/j.snb.2025.138231","RegionNum":1,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0

Abstract

As with most chemical sensors, the proliferation of optodes in analytical routine is hampered by the need for periodic sensor calibration. This fundamental limitation arises from a number of chemical and technological factors (e.g., instability of the illumination source, changes in the composition of the sensor phase, etc.), leading to deviation of the optodes signal in parallel measurements and over time. The development of new approaches for calibration-free determination of analytes with optodes remains a pertinent challenge in the field of chemical sensors. The presented study is aimed at the development and validation of ion-selective optical sensor arrays that do not require calibration. These arrays incorporate an internal optical signal reference composed of polymeric optodes with suppressed exchange capacity. The applicability of such optodes as optical signal standards is demonstrated in a broad range of pH and background electrolyte concentrations: from pH 5 to 9 in solutions of cations and lipophilic anions with concentrations reaching 0.1 M. This experimental observation is supported by theoretical analyses performed by numerical simulation. The study identifies the effect of coextraction and ion exchange on the range of the optical signal stability. Furthermore, it is shown both theoretically and experimentally that by modulating the acidity of the lipophilic indicator in the standard optode, the operational range of the internal reference can be adjusted to the desired pH or electrolyte concentrations. In addition, the validation results demonstrate the feasibility of determining Na+ concentration in the sample using internal standards with fair accuracy.

Abstract Image

具有内部参考的离子选择光学传感器:对延长功能和无校准测量的追求
与大多数化学传感器一样,光电器件在分析常规中的扩散受到周期性传感器校准需求的阻碍。这种基本限制源于许多化学和技术因素(例如,照明源的不稳定性,传感器相位组成的变化等),导致光电二极管信号在并行测量和随时间变化的偏差。在化学传感器领域,开发新的无校准分析方法仍然是一个相关的挑战。本研究旨在开发和验证不需要校准的离子选择性光学传感器阵列。这些阵列包含一个内部光信号参考,由抑制交换容量的聚合物光电器件组成。这种光电器件作为光信号标准的适用性在广泛的pH和背景电解质浓度范围内得到了证明:从pH 5到9,在浓度达到0.1 m的阳离子和亲脂阴离子溶液中,这一实验观察得到了数值模拟理论分析的支持。研究了共萃取和离子交换对光信号稳定范围的影响。此外,理论和实验都表明,通过调节标准光电二极管中亲脂指示剂的酸度,内参的工作范围可以调整到所需的pH或电解质浓度。验证结果表明,采用内标法测定样品中Na+浓度是可行的,准确度较高。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Sensors and Actuators B: Chemical
Sensors and Actuators B: Chemical 工程技术-电化学
CiteScore
14.60
自引率
11.90%
发文量
1776
审稿时长
3.2 months
期刊介绍: Sensors & Actuators, B: Chemical is an international journal focused on the research and development of chemical transducers. It covers chemical sensors and biosensors, chemical actuators, and analytical microsystems. The journal is interdisciplinary, aiming to publish original works showcasing substantial advancements beyond the current state of the art in these fields, with practical applicability to solving meaningful analytical problems. Review articles are accepted by invitation from an Editor of the journal.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信