Preventing microbial contamination during radiological imaging: Experimental evaluation of a multiuse contrast media infusion system incorporating sequential one-way valves within a dual-safety design.

IF 3.8 3区 医学 Q2 INFECTIOUS DISEASES
Matheus Roland van Rens, Kevin Hugill, L Natalia Galvan, Benjamin Tanner
{"title":"Preventing microbial contamination during radiological imaging: Experimental evaluation of a multiuse contrast media infusion system incorporating sequential one-way valves within a dual-safety design.","authors":"Matheus Roland van Rens, Kevin Hugill, L Natalia Galvan, Benjamin Tanner","doi":"10.1016/j.ajic.2025.06.006","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Health care-associated infections are a significant cause of patient morbidity, mortality, and health care expenditure. In diagnostic imaging, multiuse contrast media infusion systems are increasingly common; however, their use raises concerns regarding cross-contamination risk.</p><p><strong>Methods: </strong>A controlled experimental model was used to assess the ability of a multiuse infusion system to prevent microbial contamination during simulated clinical conditions. Escherichia coli and MS2 bacteriophage were selected to model bacterial and viral contamination risks, respectively. Inocula were introduced at key connection points during two 12-hour experimental sessions. The primary outcome was the presence or absence of microbial growth in retrieved fluid samples. Control testing was conducted to validate sterility, microbial viability, and experimental integrity.</p><p><strong>Results: </strong>Positive and negative control testing performed as expected. No microbial growth was detected in any of the 51 experimental fluid samples. This corresponds to a 95% confidence upper bound of 5.8% for undetected contamination events.</p><p><strong>Conclusions: </strong>The system's design, including sequential dual 1-way valves effectively prevented bacterial and viral contamination under laboratory conditions. These findings support its potential for safe multiuse in radiological settings.</p>","PeriodicalId":7621,"journal":{"name":"American journal of infection control","volume":" ","pages":""},"PeriodicalIF":3.8000,"publicationDate":"2025-06-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"American journal of infection control","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.ajic.2025.06.006","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"INFECTIOUS DISEASES","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Health care-associated infections are a significant cause of patient morbidity, mortality, and health care expenditure. In diagnostic imaging, multiuse contrast media infusion systems are increasingly common; however, their use raises concerns regarding cross-contamination risk.

Methods: A controlled experimental model was used to assess the ability of a multiuse infusion system to prevent microbial contamination during simulated clinical conditions. Escherichia coli and MS2 bacteriophage were selected to model bacterial and viral contamination risks, respectively. Inocula were introduced at key connection points during two 12-hour experimental sessions. The primary outcome was the presence or absence of microbial growth in retrieved fluid samples. Control testing was conducted to validate sterility, microbial viability, and experimental integrity.

Results: Positive and negative control testing performed as expected. No microbial growth was detected in any of the 51 experimental fluid samples. This corresponds to a 95% confidence upper bound of 5.8% for undetected contamination events.

Conclusions: The system's design, including sequential dual 1-way valves effectively prevented bacterial and viral contamination under laboratory conditions. These findings support its potential for safe multiuse in radiological settings.

预防放射成像中的微生物污染:双安全设计中包含顺序单向阀的多用途造影剂输注系统的实验评估。
背景:医疗保健相关感染(HAIs)是患者发病率、死亡率和医疗保健支出的重要原因。在诊断成像中,多用途造影剂输注系统越来越普遍;然而,它们的使用引起了人们对交叉污染风险的担忧。方法:采用对照实验模型评估多用途输液系统在模拟临床条件下预防微生物污染的能力。选择大肠杆菌和MS2噬菌体分别模拟细菌和病毒污染风险。在两个12小时的实验期间,在关键连接点接种疫苗。主要结果是在回收的液体样品中是否存在微生物生长。进行对照试验以验证无菌性、微生物活力和实验完整性。结果:阳性和阴性对照试验均按预期进行。51个实验液体样本中均未检测到微生物生长。这对应于未检测到的污染事件的95%置信上限为5.8%。结论:该系统的设计,包括顺序双单向阀,有效地防止了实验室条件下的细菌和病毒污染。这些发现支持其在放射环境中安全多用途的潜力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
7.40
自引率
4.10%
发文量
479
审稿时长
24 days
期刊介绍: AJIC covers key topics and issues in infection control and epidemiology. Infection control professionals, including physicians, nurses, and epidemiologists, rely on AJIC for peer-reviewed articles covering clinical topics as well as original research. As the official publication of the Association for Professionals in Infection Control and Epidemiology (APIC)
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信